B Nivbrant1, J Kärrholm, P Söderlund.
Abstract
40 patients with primary arthrosis were randomized to receive either a cemented SHP (Scientific Hip Prosthesis) or Lubinus SP2 prosthesis. At 2 years radiostereometric measurements showed increased proximal migration (0.4/0.2 mm; p = 0.02) and more proximal wear (0.3/0.1 mm, p = 0.01) of the SHP socket. The SHP stem also subsided (-O.6/-0.1 mm, p<0.001) and rotated more into retroversion (2.6/0.3 degrees ) than did the SP2 design. This subsidence mainly occurred inside the cement mantle in 17 of 18 cases (13 SHP, 4 SP2), where this type of motion could be measured. The changes in bone mineral density evaluated with DEXA and the clinical results did not differ between the 2 groups. The subsidence of the SHP stem is the most pronounced so far recorded with radiostereometry in stems without a completely polished surface. This subsidence and the rotational instability imply a substantial risk of abrasive wear and increased stresses in the cement mantle.Entities:
Mesh:
Year: 1999 PMID: 10665721 DOI: 10.3109/17453679908997844
Source DB: PubMed Journal: Acta Orthop Scand ISSN: 0001-6470