Literature DB >> 10660164

Aortic valve replacement with Medtronic Freestyle bioprosthesis: 5-year results.

D B Doty1, A Cafferty, P Cartier, H A Huysmans, N D Kon, A H Krause, R C Millar, C F Sintek, S Westaby.   

Abstract

The Medtronic Freestyle aortic root bioprosthesis has been implanted in patients since August 1992. This study reviews clinical and echocardiographic results at midterm (5 years) after implantation. The Freestyle bioprosthesis was implanted in 1,100 patients in a 21-center Food and Drug Administration clinical trial from August 1992 to October 1998. The device was implanted (1) as a subcoronary valve replacement, (2) as a complete aortic root replacement (full-root), or (3) as a root inclusion. Patients were followed annually by clinical examination and echocardiography. There were 47 deaths early after operation (7.1%). There were 2,478 patient-years of follow-up during which there were 99 deaths or 4.0/pt.-yr. Before implantation, 73% of patients were in New York Heart Association functional class III or IV. After operation, 95% were in class I or II. Transvalvular gradient 4 years after operation was low (7.5+/-5.3 mm Hg) for all valve sizes (subcoronary implant). Small valves (19 and 21 mm) had mean gradients (10.2+/-3.0, 9.1+/-4.4 mm Hg). There was no or mild valve insufficiency in 98% of patients. Actuarial analysis at 5 years showed the rate for freedom from thromboembolism of 86% in subcoronary implant and 93% in full root replacement. Freedom from endocarditis was 98%. Freedom from reoperation for explant of the valve highest in patients having full root replacement (98%) and lower with root inclusion (94%). There were 20 bioprostheses explanted; 10 for endocarditis, 8 for technical reasons, and 2 for structural deterioration. The Medtronic Freestyle bioprosthesis (1) has excellent hemodynamic performance, (2) techniques for insertion that result in a competent valve, (3) low rates of thromboembolism and endocarditis, and (4) a rare rate of structural deterioration at 5 years.

Entities:  

Mesh:

Year:  1999        PMID: 10660164

Source DB:  PubMed          Journal:  Semin Thorac Cardiovasc Surg        ISSN: 1043-0679


  5 in total

1.  Stentless xenograft implantation in aortic position in a 7-year-old girl.

Authors:  Sadik Sharef; Ryan Coleman; Shannon Rivenes; Ivan Wilmot; Carolyn Altman; Ramiro Madden-Fuentes; Jorge Salazar
Journal:  Tex Heart Inst J       Date:  2010

2.  Early results of Bentall-type operations during the last 10 years: comparison of mechanical valves and stentless bioprostheses.

Authors:  Koki Nakamura; Tomohiro Asai; Mikiko Murakami; Yosuke Saitoh; Hiroki Yamaguchi
Journal:  Gen Thorac Cardiovasc Surg       Date:  2007-01

3.  New modification of modified bentall procedure (A single centre experience).

Authors:  Ghulam Hussain; Naseem Ahmad; Sohail Ahmad; Mirza Ahmad Raza Baig; Sara Zaheer; Aamir Furkan
Journal:  Pak J Med Sci       Date:  2015 Nov-Dec       Impact factor: 1.088

4.  Hospital Results of a Single Center Database for Stentless Xenograft Use in a Full Root Technique in Over 970 Patients.

Authors:  Jerry Easo; Alexander Weymann; Philipp Hölzl; Michael Horst; Harald Eichstaedt; Ahmed Mashhour; Konstantin Zhigalov; Marcin Szczechowicz; Rohit Philip Thomas; Anton Sabashnikov; Otto E Dapunt
Journal:  Sci Rep       Date:  2019-03-13       Impact factor: 4.379

5.  Effect of surgeon on transprosthetic gradients after aortic valve replacement with Freestyle stentless bioprosthesis and its consequences: a follow-up study in 587 patients.

Authors:  Alexander Albert; Ines Florath; Ulrich Rosendahl; Wael Hassanein; Eberhard V Hodenberg; Stefan Bauer; Ina Ennker; Jürgen Ennker
Journal:  J Cardiothorac Surg       Date:  2007-10-05       Impact factor: 1.637

  5 in total