Literature DB >> 10650301

Sex and other predictors of intra-aortic balloon counterpulsation-related complications: prospective study of 1119 consecutive patients.

M Cohen1, M S Dawson, C Kopistansky, R McBride.   

Abstract

BACKGROUND AND
OBJECTIVE: Intra-aortic balloon counterpulsation (IABC) complication rates remain significant despite technical advances that have simplified and expanded its use. Previous reports implicated patient height, diabetes, or peripheral vascular disease as risk factors. However, these studies were small and not prospective. Therefore a prospective study at one high-volume center was conducted examining the complications associated with IABC and the role of sex and other risk factors in the current era.
METHODS: The study prospectively evaluated 1174 consecutive percutaneous IAB insertions in 1119 patients between 1993 and 1997. Major complications were defined as embolism or limb ischemia requiring surgery; bleeding requiring transfusion or surgery; systemic infection; balloon rupture; or death from one of these causes. Minor complications were defined as limb ischemia or pulse loss resolving without surgery or after IAB removal or bleeding not requiring transfusion or surgical intervention. All variables were analyzed with univariate and stepwise multivariate analysis.
RESULTS: Data were collected on 1119 patients (727 men and 392 women) with a mean age of 65 +/- 11 years. The prevalence of diabetes, hypertension, and peripheral vascular disease was 27%, 52%, and 8%, respectively. Complications occurred in 166 patients (15%) and a major complication occurred in 126 (11%) of the 1119 patients. Multivariate logistic regression analysis was done with demographic, clinical, and procedural variables in a cohort of 1106 patients. The analysis identified peripheral vascular disease (relative risk [RR] 4.1), female sex (RR 2.3), and body surface area (RR 0.26 per m(2)) as independent predictors of a major complication. In addition, cardiac index (RR 0.7) was also identified as an independent predictor of any or major complications in a subset of 915 patients. In 754 high-risk patients (women or patients with peripheral vascular disease, diabetes, cardiac index <2.2 L/min/m(2), or body surface area <1.8 m(2)), 114 major complications occurred (15%) compared with 8 (3%) among 278 non-high-risk patients (P <.0001).
CONCLUSIONS: The current complication rate associated with IABC remains significant. Advances in IAB technology need to focus on the high-risk subset of patients that includes women, smaller patients, and those with peripheral vascular disease.

Entities:  

Mesh:

Year:  2000        PMID: 10650301     DOI: 10.1067/mhj.2000.101489

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  11 in total

1.  Intra-aortic balloon pump counterpulsation: are we optimizing the management of cardiogenic shock?

Authors:  F Shahid; M Akhtar; F Khan; C A A Chahal
Journal:  J R Soc Med       Date:  2013-02       Impact factor: 5.344

2.  Efficacy of Elective Intra-aortic Balloon Pump Therapy for High-Risk Off-Pump Coronary Artery Bypass: A Prospective Comparative Study.

Authors:  Mizuho Imamaki; Kaoru Matsuura; Yuriko Niitsuma; Hitoshi Shimura; Masaru Miyazaki
Journal:  Ann Vasc Dis       Date:  2009-12-14

3.  Safety of intra-aortic balloon pump using glycoprotein IIb/IIIa antagonists.

Authors:  Avishag Laish-Farkash; Hanoch Hod; Shlomo Matetzky; Victor Guetta
Journal:  Clin Cardiol       Date:  2009-02       Impact factor: 2.882

Review 4.  Mechanical Circulatory Assist Devices: Available Modalities and Review of Literature.

Authors:  Mohammed Faluk; Syed Mustajab Hasan; Tianyu Jiang; Ramy Abdelmaseih; Jigar Patel
Journal:  Heart Views       Date:  2021-01-14

5.  Perioperative intra-aortic balloon counterpulsation in a patient with myocardium at risk undergoing urgent noncardiac surgery.

Authors:  Tim Nebelsiek; Florian Weis; Martin Angele; Florian Brettner
Journal:  Ann Card Anaesth       Date:  2015 Apr-Jun

Review 6.  Vascular complications following intra-aortic balloon pump implantation: an updated review.

Authors:  Monique M de Jong; Roberto Lorusso; Fatima Al Awami; Francesco Matteuci; Orlando Parise; Pieter Lozekoot; Massimo Bonacchi; Jos G Maessen; Daniel M Johnson; Sandro Gelsomino
Journal:  Perfusion       Date:  2017-08-17       Impact factor: 1.972

7.  Application of intra-aortic balloon pump in resection and anastomosis of trachea.

Authors:  Zargham Hossein Ahmadi; Amir Ali Rahnemai-Azar; Mohammad Behgam Shadmehr; Tahereh Parsa; Neda Behzadnia; Zahra Ansari Aval; Bahareh Mokri; Hamideh Gholamhoseini
Journal:  Tanaffos       Date:  2014

8.  Leg amputation: A rare complication of intra-aortic balloon pump.

Authors:  Alka Gupta; Vinod Bala Dhir; Michell Gulabani; Anupama Gill Sharma
Journal:  Anesth Essays Res       Date:  2016 Jan-Apr

9.  Safety and efficacy of the StarClose vascular closure system following 8-Fr sheath placement for intra-aortic balloon pump: a single-center analysis of 42 consecutive patients.

Authors:  Guang Chu; Wenyi Yang; Guobing Zhang; Zhi Zhang; Shaowen Liu; Baogui Sun; Fang Wang
Journal:  Med Princ Pract       Date:  2014-04-23       Impact factor: 1.927

10.  Resuscitative endovascular balloon occlusion of the aorta: rupture risk and implications for blind inflation.

Authors:  Philip J Wasicek; William A Teeter; Megan L Brenner; Melanie R Hoehn; Thomas M Scalea; Jonathan J Morrison
Journal:  Trauma Surg Acute Care Open       Date:  2018-01-24
View more