Assessment of blood lactate: practical evaluation of the Biosen 5030 lactate analyzer.

R C Davison1, D Coleman, J Balmer, M Nunn, S Theakston, M Burrows, S Bird.   

Abstract

PURPOSE: The aim of this study was to assess the validity and reliability of the Biosen 5030 lactate analyzer compared with a YSI 2300 lactate analyzer and a Kodak Ektachem DTII in a practical laboratory study context.
METHODS: To assess validity, 144 triplicate capillarized blood samples, across a range of values, were analyzed using the three analyzers. To assess reliability a further 665 samples were repeat analyzed. Temporal stability was determined by the reanalysis of resting and maximal exercise blood samples, after a period of storage ranging from 7 to 20 h, at room temperature. To measure inter- and intra-investigator reliability, 20 resting samples were taken from three different subjects by different investigators and a coefficient of variation was determined.
RESULTS: There were strong relationships between the Biosen, the YSI (r2 = 0.97), and the Kodak Ektachem (r2 = 0.91). An analysis of Biosen compared with YSI revealed a positive bias of 0.37 mmol x L(-1) (95% limits of agreement, -0.85 to 1.59 mmol x L(-1)). The test-retest reliability correlation was significant (r2 = 0.99, P < 0.05), but a paired t-test revealed a small (0.03 mmol x L(-1), P < 0.05) significant difference. The coefficient of variation from the three investigators across the 20 samples ranged from 1.3 to 3%. Blood lactate concentration in resting blood samples did significantly increase in value (0.2 mmol x L(-1), P < 0.05) after 7-h exposure to the air, whereas there was no change in maximal exercise blood lactate values after 20-h exposure to the air.
CONCLUSIONS: In a practical context, the Biosen 5030 lactate analyzer was comparable to the other analyzers giving fast reliable measures of blood lactate concentrations over the full range of values, which remained stable over extended periods at room temperature.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10647556     DOI: 10.1097/00005768-200001000-00036

Source DB:  PubMed          Journal:  Med Sci Sports Exerc        ISSN: 0195-9131            Impact factor:   5.411


  4 in total

1.  Familiarisation and reliability of sprint test indices during laboratory and field assessment.

Authors:  James G Hopker; Damian A Coleman; Jonathan D Wiles; Andrew Galbraith
Journal:  J Sports Sci Med       Date:  2009-12-01       Impact factor: 2.988

2.  Effects of mild-exercise training cessation in human skeletal muscle.

Authors:  Jonny St-Amand; Mayumi Yoshioka; Yuichiro Nishida; Takuro Tobina; Naoko Shono; Hiroaki Tanaka
Journal:  Eur J Appl Physiol       Date:  2011-06-17       Impact factor: 3.078

3.  Metabolic and respiratory adaptations during intense exercise following long-sprint training of short duration.

Authors:  Claire Thomas; Olivier Bernard; Carina Enea; Chadi Jalab; Christine Hanon
Journal:  Eur J Appl Physiol       Date:  2011-06-09       Impact factor: 3.078

4.  A Validation Study of a Noninvasive Lactate Threshold Device.

Authors:  Rebecca M McMorries; Dustin P Joubert; Eric J Jones; Mark D Faries
Journal:  Int J Exerc Sci       Date:  2019-01-01
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.