| Literature DB >> 10642130 |
C M Brodsky1, F Chang, S C Vlay.
Abstract
To reassess the function of the implantable cardioverter defibrillator (ICD) many electrophysiology centers perform a second test after the initial test at implant. A prospective multicenter study evaluated the necessity and yield of routine postimplant defibrillator testing. The results of 843 postimplant defibrillator tests were collected from 31 centers. The 764 routine tests in which ventricular fibrillation was successfully induced were analyzed. Variables examined included patient age, presenting arrhythmia, underlying heart disease, left ventricular ejection fraction, defibrillator age, make and model of ICD, electrode system, defibrillation threshold, polarity, and waveform. The overall failure rate was 3.1% (24/764). Units tested later than 365 days after implant tended to have a higher failure rate than those tested within the first month or the next eleven months (6.5%, 3.0%, 2.3%, respectively, P = 0.374). The failure rate was higher in patients with left ventricular ejection fraction < 40% than those with higher ejection fractions (3.8% vs 2.0%, P = 0.167). These trends did not reach statistical significance. No other baseline characteristic was associated with higher failure rates. Routine testing of ICDs reveals an overall failure rate of 3.1%. While the rate was low, defibrillator failure places the patient at high risk for sudden cardiac death. As any failure in this population is associated with a high risk of sudden cardiac death, routine defibrillator testing may be justified.Entities:
Mesh:
Year: 1999 PMID: 10642130 DOI: 10.1111/j.1540-8159.1999.tb00409.x
Source DB: PubMed Journal: Pacing Clin Electrophysiol ISSN: 0147-8389 Impact factor: 1.976