Endotracheal cardiac output monitor.

A W Wallace1, A Salahieh, A Lawrence, K Spector, C Owens, D Alonso.   

Abstract

BACKGROUND: The endotracheal cardiac output monitor (ECOM) is a new device that uses an endotracheal tube with multiple electrodes to measure cardiac output (CO). It measures the changes in electrical impedance caused by pulsatile blood flow in the aorta. The system was tested for safety and efficacy in 10 swine.
METHODS: Swine (60-80 kg) were chronically instrumented with a transit time flow probe on the ascending aorta and vascular occluders on the vena cava and pulmonary artery. After a minimum recovery of 4 days, the animals were anesthetized and intubated with an ECOM endotracheal tube. CO measurements from the ECOM system were compared to transit time flow probe measurements using linear regression and Bland-Altman analysis. Three different inotropic states were studied: (1) baseline; (2) increased (dobutamine); and (3) decreased (esmolol). CO was changed at each inotropic state by impeding left ventricular filling with the vena cava or pulmonary artery occluders. CO values between 0 and 15 l/min were studied. Pigs were studied for 24 h consecutively.
RESULTS: There was no deterioration of the impedance signal with time and no tracheal injury from the ECOM electrodes. There is a linear relationship between the ECOM and transit time flow probe CO between 0 and 15 l/min (slope = 0.94; intercept = 0.15 l/min; R2= 0.77). The mean difference between the two measures (bias) is 0.15 l/min and the SD is 1.34 l/min. The limits of agreement are -2.53 to 2.82 l/min.
CONCLUSION: Endotracheal CO monitor is a promising technology that needs further evaluation in clinical trials.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10638915     DOI: 10.1097/00000542-200001000-00030

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  8 in total

Review 1.  Accuracy and repeatability of pediatric cardiac output measurement using Doppler: 20-year review of the literature.

Authors:  Michelle S Chew; Jan Poelaert
Journal:  Intensive Care Med       Date:  2003-09-04       Impact factor: 17.440

2.  Applications of minimally invasive cardiac output monitors.

Authors:  Jahan Porhomayon; Gino Zadeii; Samuel Congello; Nader D Nader
Journal:  Int J Emerg Med       Date:  2012-04-24

3.  Descending aortic flow contribution to intrathoracic impedance-development and preliminary testing of a dual impedance model.

Authors:  A Barry Baker; Chris N McLeod; Alastair J Roxburgh; Paul Bannister
Journal:  J Clin Monit Comput       Date:  2007-11-15       Impact factor: 2.502

4.  Endotracheal bioimpedance cardiography improves immediate postoperative outcome: a case-control study in off-pump coronary surgery.

Authors:  Thomas Leclercq; Marc Lilot; Thomas Schulz; Alexandre Meyer; Fadi Farhat; Jean-Luc Fellahi
Journal:  J Clin Monit Comput       Date:  2017-02-08       Impact factor: 2.502

Review 5.  [Meta-analyses on measurement precision of non-invasive hemodynamic monitoring technologies in adults].

Authors:  G Pestel; K Fukui; M Higashi; I Schmidtmann; C Werner
Journal:  Anaesthesist       Date:  2018-06       Impact factor: 1.041

6.  Early goal-directed therapy based on endotracheal bioimpedance cardiography: a prospective, randomized controlled study in coronary surgery.

Authors:  Jean-Luc Fellahi; David Brossier; Fabien Dechanet; Marc-Olivier Fischer; Vladimir Saplacan; Jean-Louis Gérard; Jean-Luc Hanouz
Journal:  J Clin Monit Comput       Date:  2014-11-09       Impact factor: 2.502

7.  Can endotracheal bioimpedance cardiography assess hemodynamic response to passive leg raising following cardiac surgery?

Authors:  Jean-Luc Fellahi; Marc-Olivier Fischer; Audrey Dalbera; Massimo Massetti; Jean-Louis Gérard; Jean-Luc Hanouz
Journal:  Ann Intensive Care       Date:  2012-07-11       Impact factor: 6.925

8.  Cardiac output assessed by invasive and minimally invasive techniques.

Authors:  Allison J Lee; Jennifer Hochman Cohn; J Sudharma Ranasinghe
Journal:  Anesthesiol Res Pract       Date:  2011-07-06
  8 in total