Literature DB >> 10638870

Bladder volume determination: portable 3-D versus stationary 2-D ultrasound device.

P Schnider1, P Birner, A Gendo, K Ratheiser, E Auff.   

Abstract

OBJECTIVE: To investigate how accurately a portable three-dimensional (3-D) scanner and a multipurpose two-dimensional (2-D) real-time scanner determined bladder volumes. STUDY
DESIGN: Prospective, controlled clinical trial, single-blind, crossover design. SETTING AND PARTICIPANTS: Twenty-three inpatients with permanent bladder catheters participated voluntarily in this study.
METHODS: The bladders of 20 patients were filled through an indwelling catheter with 60, 110, 160, 210, and 260 mL sterile normal saline. Volumes were measured twice with each device. Measurements were compared with the actual bladder volumes.
RESULTS: The 2-D device showed better reproducibility, particularly at lower bladder volumes. The 3-D scanner showed a significant difference between the two measurements at 160 mL (p<.05) and had poor reproducibility at 110, 210, and 260 mL. Both devices overestimated actual bladder volume at fillings of <160 mL and underestimated it at fillings of > or =160 mL. The range between the 25th and 75th percentiles was always larger for the 3-D scanner, except for the 210 mL reading.
CONCLUSION: Both devices showed sufficient accuracy for clinical practice. Ultrasound measurements of >110 mL should be followed by catheterization to detect potentially harmful bladder volumes.

Entities:  

Mesh:

Year:  2000        PMID: 10638870     DOI: 10.1016/s0003-9993(00)90215-6

Source DB:  PubMed          Journal:  Arch Phys Med Rehabil        ISSN: 0003-9993            Impact factor:   3.966


  8 in total

Review 1.  A comprehensive survey on non-invasive wearable bladder volume monitoring systems.

Authors:  Morteza Zakeri Nasrabadi; Hamideh Tabibi; Mahsa Salmani; Mahdieh Torkashvand; Eisa Zarepour
Journal:  Med Biol Eng Comput       Date:  2021-07-14       Impact factor: 2.602

2.  Validation of 3-dimensional mathematical computation of feline bladder volume using orthogonal ultrasonographic bladder images.

Authors:  Chien Hsien Kitty Yang; Shane W Bateman; Xiu Ting Yiew; Allan R Willms
Journal:  Can J Vet Res       Date:  2022-07       Impact factor: 0.897

Review 3.  Advances in Assistive Electronic Device Solutions for Urology.

Authors:  Kieran Holmes-Martin; Minghui Zhu; Shujun Xiao; Faezeh Arab Hassani
Journal:  Micromachines (Basel)       Date:  2022-03-30       Impact factor: 3.523

4.  Covert postpartum urinary retention: causes and consequences (PAREZ study).

Authors:  Pavel Dolezal; Michaela Ostatnikova; Barbora Balazovjechova; Petra Psenkova; Jozef Zahumensky
Journal:  Int Urogynecol J       Date:  2022-06-18       Impact factor: 1.932

Review 5.  Sonography of the bladder.

Authors:  Pietro Pavlica; Caterina Gaudiano; Libero Barozzi
Journal:  World J Urol       Date:  2004-07-28       Impact factor: 4.226

6.  Comparison of 2D and 3D ultrasound methods to measure serial bladder volumes during filling: Steps toward development of non-invasive ultrasound urodynamics.

Authors:  Anna S Nagle; Rachel J Bernardo; Jary Varghese; Laura R Carucci; Adam P Klausner; John E Speich
Journal:  Bladder (San Franc)       Date:  2018-01-04

7.  Development and evaluation of automated ultrasonographic detection of bladder diameter for estimation of bladder urine volume.

Authors:  Masaru Matsumoto; Takuya Tsutaoka; Koichi Yabunaka; Mayumi Handa; Mikako Yoshida; Gojiro Nakagami; Hiromi Sanada
Journal:  PLoS One       Date:  2019-09-05       Impact factor: 3.240

8.  Three-dimensional bladder ultrasound for estimation of urine volume in dogs compared with traditional 2-dimensional ultrasound methods.

Authors:  Allison Kendall; Erin Keenihan; Zachary T Kern; Crystal Lindaberry; Adam Birkenheuer; George E Moore; Shelly L Vaden
Journal:  J Vet Intern Med       Date:  2020-11-06       Impact factor: 3.333