A comparison between gravimetric and volumetric techniques of wear measurement of UHMWPE acetabular cups against zirconia and cobalt-chromium-molybdenum femoral heads in a hip simulator.

S L Smith1, A Unsworth.   

Abstract

Five cobalt-chromium-molybdenum (CoCrMo) and five zirconia femoral head components have been wear tested against 28 mm diameter ultra high molecular weight polyethylene (UHMWPE) acetabular cups for 5 million cycles in the Durham hip joint wear simulator using bovine serum as a lubricant. Wear measurements used gravimetric and volumetric techniques and no statistically significant difference was found between the measurement methods. The wear rates of the acetabular cups against both femoral heads are presented for both measurement methods. The UHMWPE acetabular cups showed a statistically significant higher linear wear rate for the first 2 million cycles than the lower linear wear rate from 2 million cycles to the end of the test, against both femoral head materials. Over the full duration of the wear test, the wear rates of acetabular cups articulating against zirconia femoral heads were lower than against CoCrMo femoral heads. The wear rates up to 2 million cycles and from 2 to 5 million cycles for both femoral head materials were consistent with other studies.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10635696     DOI: 10.1243/0954411991535086

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  2 in total

1.  Increased volumetric wear of polyethylene liners with more than 3 years of shelf-life time.

Authors:  T J S Puolakka; J T Keränen; K A Juhola; K J J Pajamäki; P J Halonen; J K Nevalainen; V Saikko; M U K Lehto; M Järvinen
Journal:  Int Orthop       Date:  2003-04-01       Impact factor: 3.075

2.  Biotribological study of large diameter ceramic-on-CFR-PEEK hip joint including fluid uptake, wear and frictional heating.

Authors:  Qian Qian Wang; Jun Jie Wu; Anthony Unsworth; Adam Briscoe; Marcus Jarman-Smith; Conor Lowry; David Simpson; Simon Collins
Journal:  J Mater Sci Mater Med       Date:  2012-03-28       Impact factor: 3.896

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.