Literature DB >> 10615970

Radiological factors influencing femoral and acetabular failure in cemented Charnley total hip arthroplasties.

M A Ritter1, H Zhou, C M Keating, E M Keating, P M Faris, J B Meding, M E Berend.   

Abstract

We have made a retrospective review of 185 cemented Charnley total hip arthroplasties performed between 1970 and 1974 to determine the relationships between radiological variables and failure of the femoral and acetabular components. We measured the acetabular wear, the orientation of the cup, the thickness and consistency of acetabular and femoral cement mantles, radiolucency and femoral alignment. The mean follow-up was for 11.7 years. Femoral loosening was demonstrable radiologically in 15 hips (8.1%), ten (5.4%) of which were revised during the period of follow-up. Only when the first postoperative radiograph showed a thin cement mantle in Gruen zone 5 was there a significant association with failure of the femoral component. There were 12 loose acetabular components (6.5%), nine (4.8%) of which were revised. When the initial radiograph after operation showed radiolucency in DeLee and Charnley zone 1, the incidence of acetabular loosening was 28.21%. If such radiolucency was not present, the incidence of acetabular loosening was only 0.69%. Our findings emphasise the importance of careful cementing.

Entities:  

Mesh:

Year:  1999        PMID: 10615970     DOI: 10.1302/0301-620x.81b6.9634

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  14 in total

1.  Peri-acetabular radiolucent lines: inter- and intra-observer agreement on post-operative radiographs.

Authors:  D Kneif; M Downing; G P Ashcroft; P Gibson; D Knight; W Ledingham; J Hutchison
Journal:  Int Orthop       Date:  2005-04-02       Impact factor: 3.075

2.  Prediction of Charnley femoral stem aseptic loosening by early post-operative radiological features.

Authors:  M H A Malik; N Fisher; J Gray; B M Wroblewski; P R Kay
Journal:  Int Orthop       Date:  2005-08-05       Impact factor: 3.075

3.  Pulsed lavage reduces the incidence of radiolucent lines under the tibial tray of Oxford unicompartmental knee arthroplasty: pulsed lavage versus syringe lavage.

Authors:  Michael Clarius; Christian Hauck; Joern B Seeger; Andrew James; David W Murray; Peter R Aldinger
Journal:  Int Orthop       Date:  2009-02-14       Impact factor: 3.075

4.  The influence of a suction device on fixation of a cemented cup using RSA.

Authors:  A John Timperley; Sarah L Whitehouse; Patrick G Hourigan
Journal:  Clin Orthop Relat Res       Date:  2008-11-08       Impact factor: 4.176

5.  Comparison of flanged and unflanged acetabular cup design. An experimental study using ceramic and cadaveric acetabuli.

Authors:  Mette Ørskov; Saba Abdulghani; Ian McCarthy; Kjeld Søballe; Gunnar Flivik
Journal:  Acta Orthop       Date:  2010-10       Impact factor: 3.717

6.  Impaction grafting of the acetabulum with ceramic bone graft substitute: high survivorship in 43 patients with a mean follow-up period of 4 years.

Authors:  Michael R Whitehouse; Peter J Dacombe; Jason C J Webb; Ashley W Blom
Journal:  Acta Orthop       Date:  2013-08       Impact factor: 3.717

7.  Radiographic grading of femoral stem cementation in hip arthroplasty.

Authors:  Flávio Luís Garcia; Arthur Tomotaka Sugo; Celso Hermínio Ferraz Picado
Journal:  Acta Ortop Bras       Date:  2013-01       Impact factor: 0.513

8.  [Predictive factors of aseptic loosening of cemented total hip prostheses].

Authors:  Rateb Kochbati; Hedi Rbai; Marouene Jlailia; Hassen Makhlouf; Abderrazak Bouguira; Med Samir Daghfous
Journal:  Pan Afr Med J       Date:  2016-07-20

9.  Once-weekly oral medication with alendronate does not prevent migration of knee prostheses: A double-blind randomized RSA study.

Authors:  Ulrik Hansson; Sören Toksvig-Larsen; Leif Ryd; Per Aspenberg
Journal:  Acta Orthop       Date:  2009-02       Impact factor: 3.717

10.  Impaction grafting of the acetabulum with ceramic bone graft substitute mixed with femoral head allograft: high survivorship in 43 patients with a median follow-up of 7 years: a follow-up report.

Authors:  Michael R Whitehouse; Peter J Dacombe; Jason C J Webb; Ashley W Blom
Journal:  Acta Orthop       Date:  2013-04-17       Impact factor: 3.717

View more