Prospective comparison of plateletapheresis using four apheresis systems on the same donors.

E A Burgstaler1, A A Pineda, S C Bryant.   

Abstract

To address the demand for higher plateletapheresis efficiency while maintaining consistent leukoreduction, manufacturers of apheresis systems have introduced new equipment or modifications to existing equipment. Using the same 20 donors, we compared the Fresenius AS104 (AS104), Fenwal Amicus, and COBE Spectra Version 7 Leukoreduction System (Turbo) to the COBE Spectra Version 5 Leukoreduction System (V5-LRS) and each other in regard to platelet (plt) collection efficiencies, processing times, and leukoreduction consistency. Using current pre-procedure platelet counts, target endpoints were set at 6-6.5 x 10(11) plt or 3.3-4.0 x 10(11) plt in up to 100 minutes processing time. Median platelet yields for V5-LRS, AS104, Amicus, and Turbo were 3.98, 3.63, 5.03, and 4.99 x 10(11) plt respectively; median collection efficiencies were 53, 46, 73, and 56% respectively; median collection rates were.049,.039,.065, and. 060 x 10(11) plt/minute respectively; double product frequencies were 35, 10, 40, and 30% respectively; and median processing times were 87, 92, 77, and 79 minutes, respectively. Amicus had a significantly higher collection efficiency and higher incidence of double products than all other systems. While AS104 had a significantly lower collection efficiency and lowest double product frequency than all other systems, Amicus and Turbo had significantly lower processing times than V5-LRS and AS104. AS104 leukoreduction was inconsistent, but V5-LRS, Amicus, and Turbo were consistently leukoreduced ((99.8% had <5 x 10(6) WBC at 95% confidence interval). The best overall performance was for the Amicus with Turbo a close second. Copyright 1999 Wiley-Liss, Inc.

Mesh:

Year:  1999        PMID: 10611625     DOI: 10.1002/(sici)1098-1101(1999)14:4<163::aid-jca2>3.0.co;2-3

Source DB:  PubMed          Journal:  J Clin Apher        ISSN: 0733-2459            Impact factor:   2.821


  3 in total

1.  Comparison of Plateletpheresis on the Fenwal Amicus and Fresenius Com.Tec Cell Separators.

Authors:  Fevzi Altuntas; Ismail Sari; Ismail Kocyigit; Leylagul Kaynar; Sibel Hacioglu; Ahmet Ozturk; Mehmet Oztekin; Musa Solmaz; Bulent Eser; Mustafa Cetin; Ali Unal
Journal:  Transfus Med Hemother       Date:  2008-09-16       Impact factor: 3.747

2.  Comparison of plateletpheresis on three continuous flow cell separators.

Authors:  Anita Tendulkar; Sunil B Rajadhyaksha
Journal:  Asian J Transfus Sci       Date:  2009-07

3.  Plateletpheresis in the Era of Automation: Optimizing Donor Safety and Product Quality Using Modern Apheresis Instruments.

Authors:  Sudipta Sekhar Das; Subrata Sen; R U Zaman; Rathindra Nath Biswas
Journal:  Indian J Hematol Blood Transfus       Date:  2020-08-16       Impact factor: 0.900