Risks of spontaneous injury and extraction of an active fixation pacemaker lead: report of the Accufix Multicenter Clinical Study and Worldwide Registry.

G N Kay1, J A Brinker, D T Kawanishi, C J Love, M A Lloyd, R C Reeves, G Pioger, J A Fee, M K Overland, L G Ensign, G L Grunkemeier.   

Abstract

BACKGROUND: The Telectronics Accufix pacing leads were recalled in November 1994 after 2 deaths and 2 nonfatal injuries were reported. This multicenter clinical study (MCS) of patients with Accufix leads was designed to determine the rate of spontaneous injury related to the J retention wire and results of lead extraction. METHODS AND
RESULTS: The MCS included 2589 patients with Accufix atrial pacing leads that were implanted at or who were followed up at 12 medical centers. Patients underwent cinefluoroscopic imaging of their lead every 6 months. The risk of J retention wire fracture was approximately 5.6%/y at 5 years and 4.7%/y at 10 years after implantation. The annual risk of protrusion was 1.5%. A total of 40 spontaneous injuries were reported to a worldwide registry (WWR) that included data from 34 672 patients (34 892 Accufix leads), including pericardial tamponade (n=19), pericardial effusion (n=5), atrial perforation (n=3), J retention wire embolization (n=4), and death (n=6). The risk of injury was 0.02%/y (95% CI, 0.0025 to 0. 072) in the MCS and 0.048%/y (95% CI, 0.035 to 0.067) in the WWR. A total of 5299 leads (13%) have been extracted worldwide. After recall in the WWR, fatal extraction complications occurred in 0.4% of intravascular procedures (16 of 4023), with life-threatening complications in 0.5% (n=21). Extraction complications increased with implant duration, female sex, and J retention wire protrusion.
CONCLUSIONS: Accufix pacing leads pose a low, ongoing risk of injury. Extraction is associated with substantially higher risks, and a conservative management approach is indicated for most patients.

Entities:  

Mesh:

Year:  1999        PMID: 10587339     DOI: 10.1161/01.cir.100.23.2344

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  13 in total

1.  The effect of device advisories on implantable cardioverter-defibrillator therapy.

Authors:  Mitesh S Amin; Kenneth A Ellenbogen
Journal:  Curr Cardiol Rep       Date:  2010-09       Impact factor: 2.931

2.  Focus on management of pacemaker and ICD advisories, recalls, and alerts.

Authors:  Mitesh S Amin; Kenneth A Ellenbogen
Journal:  Curr Treat Options Cardiovasc Med       Date:  2006-09

3.  Implantable device recalls: are we throwing away the baby with the bath water?

Authors:  T A Simmers; L V A Boersma
Journal:  Neth Heart J       Date:  2006-12       Impact factor: 2.380

4.  Cost of a recall of a single-center experience managing the Riata defibrillator lead.

Authors:  Sarah Hussain; Liza Moorman; J Randall Moorman; John P DiMarco; Rohit Malhotra; Andrew Darby; Kenneth Bilchick; J Michael Mangrum; John D Ferguson; Pamela K Mason
Journal:  Am J Cardiol       Date:  2014-10-30       Impact factor: 2.778

Review 5.  Subcutaneous and Transvenous ICDs: an Update on Contemporary Questions and Controversies.

Authors:  Daniel J Friedman; Albert S Tully; Emily P Zeitler
Journal:  Curr Cardiol Rep       Date:  2022-05-31       Impact factor: 3.955

6.  Increasing lead burden correlates with externalized cables during systematic fluoroscopic screening of Riata leads.

Authors:  Liza P Moorman; J Randall Moorman; John P DiMarco; Rohit Malhotra; Andrew Darby; Kenneth Bilchick; John D Ferguson; J Michael Mangrum; Sandeep Kamath; Pamela K Mason
Journal:  J Interv Card Electrophysiol       Date:  2012-12-20       Impact factor: 1.900

Review 7.  Pacemaker and ICD leads: strategies for long-term management.

Authors:  P Peter Borek; Bruce L Wilkoff
Journal:  J Interv Card Electrophysiol       Date:  2008-04-08       Impact factor: 1.900

8.  Cell and gene therapy for arrhythmias: Repair of cardiac conduction damage.

Authors:  Yong-Fu Xiao
Journal:  J Geriatr Cardiol       Date:  2011-09       Impact factor: 3.327

9.  Managing patients with advisory defibrillator leads: what can we learn from published data?

Authors:  F A Bracke; B M van Gelder
Journal:  Neth Heart J       Date:  2015-04       Impact factor: 2.380

10.  Critical appraisal of cardiac implantable electronic devices: complications and management.

Authors:  Luigi Padeletti; Giosuè Mascioli; Alessandro Paoletti Perini; Gino Grifoni; Laura Perrotta; Procolo Marchese; Luca Bontempi; Antonio Curnis
Journal:  Med Devices (Auckl)       Date:  2011-09-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.