13C mixed-triglyceride breath test: isotope selective non-dispersive infrared spectrometry in comparison with isotope ratio mass spectrometry in volunteers and patients with chronic pancreatitis.

C Boedeker1, O Goetze, B Pfaffenbach, A Luypaerts, B Geypens, R J Adamek.   

Abstract

BACKGROUND: The 13C mixed-triglyceride breath test (MTB) has been proposed for the non-invasive assessment of duodenal pancreatic lipase activity. Until now, stable isotope analysis of CO2 of the MTB has been carried out with isotope ratio mass spectrometry (IRMS). The aim of the present study was to compare MTB results by using the new non-dispersive infrared spectrometry (NDIRS) and the IRMS.
METHODS: Ten healthy volunteers and 10 patients with chronic pancreatitis and exocrine insufficiency were studied. After an overnight fast each subject received a test meal containing 250 mg 1,3 distearyl, 2[13C] octanoyl glycerol. Breath samples were taken at base line and at 30-min intervals over a period of 6 h postprandially. The 13C/12C ratio was determined in each breath sample by NDIRS and CF-IRMS as delta values. Results were expressed as delta over base line (DOB (per 1000)) and as cumulative percentage dose of 13C recovered (cPDR (%)). Correlations between IRMS and NDIRS were tested by linear regression analysis. For measuring agreement an Altman-Bland plot was performed.
RESULTS: A linear correlation was found (DOB: y = 0.645 +/- 0.040 x + 1.496 +/- 0.089, r = 0.70, P < 0.0001; cPDR: y = 1.269 +/- 0.031 x + 2.010 +/- 0.353, r = 0.93, P < 0.0001). For DOB the mean difference (d) was 1.0/1000, and the standard deviation (s) of the difference was 1.3/1000. The limits of agreement (d +/- 2 s) were -1.6/1000 and 3.6/1000.
CONCLUSION: The comparison of DOB and cPDR values by NDIRS and IRMS shows a moderate to good linear correlation. However, the distance of the limits of agreement is rather wide. Consequently, the validity of the MTB is diminished, which makes MTB by NDIRS less suitable for exact evaluation of non-invasive assessment of duodenal pancreatic lipase activity. Further studies are necessary to determine sensitivity and specificity of the MTB with NDIRS in larger study populations.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10582768     DOI: 10.1080/003655299750024977

Source DB:  PubMed          Journal:  Scand J Gastroenterol        ISSN: 0036-5521            Impact factor:   2.423


  4 in total

1.  (13)C mixed triglyceride breath testing using infrared spectrometry: comparison of two devices in early infancy.

Authors:  D S Kent; T Remer; C Blumenthal; K J Gaskin
Journal:  Eur J Clin Nutr       Date:  2016-01-13       Impact factor: 4.016

Review 2.  Diagnosis and treatment of pancreatic exocrine insufficiency.

Authors:  Björn Lindkvist
Journal:  World J Gastroenterol       Date:  2013-11-14       Impact factor: 5.742

3.  Sensitivity and specificity of an abbreviated (13)C-mixed triglyceride breath test for measurement of pancreatic exocrine function.

Authors:  Jutta Keller; Viola Meier; Kristina U Wolfram; Ulrich Rosien; Peter Layer
Journal:  United European Gastroenterol J       Date:  2014-08       Impact factor: 4.623

Review 4.  European guideline on indications, performance and clinical impact of 13 C-breath tests in adult and pediatric patients: An EAGEN, ESNM, and ESPGHAN consensus, supported by EPC.

Authors:  Jutta Keller; Heinz F Hammer; Paul R Afolabi; Marc Benninga; Osvaldo Borrelli; Enrique Dominguez-Munoz; Dan Dumitrascu; Oliver Goetze; Stephan L Haas; Bruno Hauser; Daniel Pohl; Silvia Salvatore; Marc Sonyi; Nikhil Thapar; Kristin Verbeke; Mark R Fox
Journal:  United European Gastroenterol J       Date:  2021-06-14       Impact factor: 4.623