Dependency of dose response of five charge-coupled device-based digital intra-oral radiographic systems on tube voltage.

K Nishikawa1, H Shibuya, M Wakoh, K Kuroyanagi.   

Abstract

OBJECTIVE: To compare the dependency of dose response of five CCD-based digital intra-oral radiographic systems on tube voltage.
METHODS: Characteristic curves for Sens-A-Ray (old and new generations; Regam Medical Systems, Sundsvall, Sweden), Computed Dental Radiography (Schick Technologies, Long Island City, NY, USA), CompuRay (Yoshida, Tokyo, Japan), and VIXA (Dentsply/Gendex, Des Plaines, IL, USA) were generated between 60 and 90 kVp at intervals of 10 kVp and corrected by subtraction of dark current. Their sensitivities were compared.
RESULTS: The sensitivity of the old generation Sens-A-Ray decreased with increase in tube voltage whereas in the other systems it increased.
CONCLUSION: In the newer systems using a rare earth intensifying screen as a scintillator, the sensitivity increases with increase in tube voltage, thus reducing the radiation dose to the patient.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10578191     DOI: 10.1038/sj/dmfr/4600484

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  2 in total

1.  Do CEC guidelines under-utilise the full potential of increasing kVp as a dose-reducing tool?

Authors:  P Doherty; D O'Leary; P C Brennan
Journal:  Eur Radiol       Date:  2003-01-25       Impact factor: 5.315

2.  Diagnostic Value of Conventional and Digital Radiography for Detection of Cavitated and Non-Cavitated Proximal Caries.

Authors:  Mahdieh Dehghani; Rasool Barzegari; Hosein Tabatabai; Sahar Ghanea
Journal:  J Dent (Tehran)       Date:  2017-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.