Hemodynamic and physical performance during maximal exercise in patients with an aortic bioprosthetic valve: comparison of stentless versus stented bioprostheses.

P Pibarot1, J G Dumesnil, J Jobin, P Cartier, G Honos, L G Durand.   

Abstract

OBJECTIVES: The objective of this study was to compare stentless bioprostheses with stented bioprostheses with regard to their hemodynamic behavior during exercise.
BACKGROUND: Stentless aortic bioprostheses have better hemodynamic performances at rest than stented bioprostheses, but very few comparisons were performed during exercise.
METHODS: Thirty-eight patients with normally functioning stentless (n = 19) or stented (n = 19) bioprostheses were submitted to a maximal ramp upright bicycle exercise test. Valve effective orifice area and mean transvalvular pressure gradient at rest and during peak exercise were successfully measured using Doppler echocardiography in 30 of the 38 patients.
RESULTS: At peak exercise, the mean gradient increased significantly less in stentless than in stented bioprostheses (+5 +/- 3 vs. +12 +/- 8 mm Hg; p = 0.002) despite similar increases in mean flow rates (+137 +/- 58 vs. +125 +/- 65 ml/s; p = 0.58); valve area also increased but with no significant difference between groups. Despite this hemodynamic difference, exercise capacity was not significantly different, but left ventricular (LV) mass and function were closer to normal in stentless bioprostheses. Overall, there was a strong inverse relation between the mean gradient during peak exercise and the indexed valve area at rest (r = 0.90).
CONCLUSIONS: Hemodynamics during exercise are better in stentless than stented bioprostheses due to the larger resting indexed valve area of stentless bioprostheses. This is associated with beneficial effects with regard to LV mass and function. The relation found between the resting indexed valve area and the gradient during exercise can be used to project the hemodynamic behavior of these bioprostheses at the time of operation. It should thus be useful to select the optimal prosthesis given the patient's body surface area and level of physical activity.

Entities:  

Mesh:

Year:  1999        PMID: 10551713     DOI: 10.1016/s0735-1097(99)00360-5

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  11 in total

Review 1.  Exercise hemodynamics in valvular heart disease.

Authors:  Anna M Booher; David S Bach
Journal:  Curr Cardiol Rep       Date:  2011-06       Impact factor: 2.931

2.  Complications of prosthetic heart valves.

Authors:  Jeanne M Vesey; Catherine M Otto
Journal:  Curr Cardiol Rep       Date:  2004-03       Impact factor: 2.931

Review 3.  Prosthesis-Patient Mismatch After Aortic Valve Replacement.

Authors:  Abdellaziz Dahou; Haïfa Mahjoub; Philippe Pibarot
Journal:  Curr Treat Options Cardiovasc Med       Date:  2016-11

4.  Transesophageal Echocardiography in Healthy Young Adult Male Baboons (Papio hamadryas anubis): Normal Cardiac Anatomy and Function in Subhuman Primates Compared to Humans.

Authors:  Arthur A Bert; William B Drake; Rachael W Quinn; Kathleen M Brasky; James E O'Brien; Gary K Lofland; Richard A Hopkins
Journal:  Prog Pediatr Cardiol       Date:  2013-08-01

5.  Full-root aortic valve replacement with stentless xenograft achieves superior regression of left ventricular hypertrophy compared to pericardial stented aortic valves.

Authors:  Reza Tavakoli; Christoph Auf der Maur; Xavier Mueller; Reinhard Schläpfer; Peiman Jamshidi; François Daubeuf; Nelly Frossard
Journal:  J Cardiothorac Surg       Date:  2015-02-03       Impact factor: 1.637

6.  Complementary Role of the Computed Biomodelling through Finite Element Analysis and Computed Tomography for Diagnosis of Transcatheter Heart Valve Thrombosis.

Authors:  Francesco Nappi; Laura Mazzocchi; Sanjeet Singh Avtaar Singh; Simone Morganti; Jean-Louis Sablayrolles; Christophe Acar; Ferdinando Auricchio
Journal:  Biomed Res Int       Date:  2018-10-22       Impact factor: 3.411

7.  Hospital Results of a Single Center Database for Stentless Xenograft Use in a Full Root Technique in Over 970 Patients.

Authors:  Jerry Easo; Alexander Weymann; Philipp Hölzl; Michael Horst; Harald Eichstaedt; Ahmed Mashhour; Konstantin Zhigalov; Marcin Szczechowicz; Rohit Philip Thomas; Anton Sabashnikov; Otto E Dapunt
Journal:  Sci Rep       Date:  2019-03-13       Impact factor: 4.379

8.  Early Hemodynamics after Aortic Valve Replacement.

Authors:  Serik Aitaliyev; Egle Rumbinaitė; Karolina Mėlinytė-Ankudavičė; Rokas Nekrošius; Vytenis Keturakis; Rimantas Benetis
Journal:  Medicina (Kaunas)       Date:  2020-12-07       Impact factor: 2.430

9.  Impact of Prosthesis-Patient Mismatch on Hemodynamics During Exercise in Patients With Aortic Stenosis After Transcatheter Aortic Valve Implantation With a Balloon-Expandable Valve.

Authors:  Haruka Kameshima; Masaki Izumo; Tomomi Suzuki; Hiroshi Ohara; Yukio Sato; Mika Watanabe; Shingo Kuwata; Kazuaki Okuyama; Ryo Kamijima; Manabu Takai; Seisyou Kou; Yasuhiro Tanabe; Tomoo Harada; Yoshihiro J Akashi
Journal:  Front Cardiovasc Med       Date:  2022-01-31

Review 10.  Need for a standardized protocol for stress echocardiography in provoking subaortic and valvular gradient in various cardiac conditions.

Authors:  Pawel Petkow Dimitrow; Carlos Cotrim; Tsung O Cheng
Journal:  Cardiovasc Ultrasound       Date:  2014-07-14       Impact factor: 2.062

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.