Validation of the SpaceLabs 90207 ambulatory blood pressure device for hemodialysis patients.

A J Peixoto1, T A Gray, S T Crowley.   

Abstract

BACKGROUND: Background Ambulatory blood pressure monitoring (ABPM) has been used increasingly in assessing dialysis patients. However, devices have not been validated formally for this population. Formal device validation is important in order to guarantee adequate performance for populations with special characteristics, such as patients undergoing hemodialysis.
OBJECTIVE: To achieve formal validation of the SpaceLabs 90207 ambulatory blood pressure device (SLD) using a modified British Hypertension Society protocol.
METHODS: Eighty-five hemodialysis patients were studied, generating 255 readings. Readings were obtained with patients supine (in a dialysis chair) over 5-10 min during hemodialysis. Simultaneous, same-arm readings were obtained through the use of a T connector to a calibrated mercury manometer. The mean differences between readings obtained by observers and those obtained by the device were calculated. Limits of agreement between observers and SLD were determined and plotted according to the method of Bland and Altman. For grading of performance, we determined the number of readings for which the readings obtained by the device were within 5, 10, and 15mmHg of manometer readings. Final gradings were ascribed according to British Hypertension Society criteria.
RESULTS: The mean blood pressure (+/-SD) was 141/76+/-31/15mmHg (observers) and 141/77+/-27/15mmHg (SLD), and the mean (+/-SD) difference between observers and device (observer-device) was -0.5+/-7.5mmHg for systolic blood pressure (SBP) and -0.2+/-5.2mmHg for diastolic blood pressure (DBP). The device was less accurate in extreme ranges of SBP. In fact, there was a positive correlation between average [(observer+device)/2] and difference (observer-device) for SBP (r =0. 54, P <0.0001), showing that underestimation in higher ranges, and overestimation in lower ranges of blood pressure occurred for SBP. For SBP, 53% of readings were within 5mmHg of those obtained by the observers, 85% were within 10mmHg, and 97% within 15mmHg. For DBP, 78% were within 5mmHg, 96% within 10mmHg, and 98% within 15mmHg. These observations conferred on the device grade C for SBP and grade B for DBP. The type of vascular access and the presence of non-functioning arteriovenous grafts and fistulas in the ipsilateral arm did not alter these results significantly.
CONCLUSIONS: These data validate the use of this device for hemodialysis patients. However, caution should be exercised in the evaluation of upper and lower ranges of SBP.

Entities:  

Mesh:

Year:  1999        PMID: 10547641

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  11 in total

1.  Blood pressure measuring devices: recommendations of the European Society of Hypertension.

Authors:  E O'Brien; B Waeber; G Parati; J Staessen; M G Myers
Journal:  BMJ       Date:  2001-03-03

Review 2.  Managing hypertension using home blood pressure monitoring among haemodialysis patients--a call to action.

Authors:  Rajiv Agarwal
Journal:  Nephrol Dial Transplant       Date:  2010-03-28       Impact factor: 5.992

Review 3.  Blood pressure in hemodialysis: targets?

Authors:  Panagiotis I Georgianos; Rajiv Agarwal
Journal:  Curr Opin Nephrol Hypertens       Date:  2017-11       Impact factor: 2.894

4.  Arm circumference: its importance for dialysis patients in the obesity era.

Authors:  Tekin Akpolat; Coşkun Kaya; Cengiz Utaş; Turgay Arinsoy; Hülya Taşkapan; Emre Erdem; M Emin Yilmaz; Rezzan Ataman; Semra Bozfakioğlu; Çetin Özener; Ibrahim Karayaylali; Rümeyza Kazancioğlu; Taner Çamsari; Mahmut Yavuz; Fevzi Ersoy; Soner Duman; Kenan Ateş
Journal:  Int Urol Nephrol       Date:  2012-06-30       Impact factor: 2.370

5.  Peridialytic, intradialytic, and interdialytic blood pressure measurement in hemodialysis patients.

Authors:  Arjun D Sinha; Rajiv Agarwal
Journal:  Am J Kidney Dis       Date:  2009-11       Impact factor: 8.860

Review 6.  Assessment and management of hypertension in patients on dialysis.

Authors:  Rajiv Agarwal; Joseph Flynn; Velvie Pogue; Mahboob Rahman; Efrain Reisin; Matthew R Weir
Journal:  J Am Soc Nephrol       Date:  2014-04-03       Impact factor: 10.121

Review 7.  Blood Pressure Measurement: A KDOQI Perspective.

Authors:  Paul E Drawz; Srinivasan Beddhu; Holly J Kramer; Michael Rakotz; Michael V Rocco; Paul K Whelton
Journal:  Am J Kidney Dis       Date:  2019-12-18       Impact factor: 8.860

8.  Volume-associated ambulatory blood pressure patterns in hemodialysis patients.

Authors:  Rajiv Agarwal
Journal:  Hypertension       Date:  2009-06-15       Impact factor: 10.190

9.  Diagnosing Masked Hypertension Using Ambulatory Blood Pressure Monitoring, Home Blood Pressure Monitoring, or Both?

Authors:  D Edmund Anstey; Paul Muntner; Natalie A Bello; Daniel N Pugliese; Yuichiro Yano; Ian M Kronish; Kristi Reynolds; Joseph E Schwartz; Daichi Shimbo
Journal:  Hypertension       Date:  2018-11       Impact factor: 9.897

10.  QardioArm Upper Arm Blood Pressure Monitor Against Omron M3 Upper Arm Blood Pressure Monitor in Patients With Chronic Kidney Disease: A Validation Study According to the European Society of Hypertension International Protocol Revision 2010.

Authors:  Victoria Mazoteras-Pardo; Ricardo Becerro-De-Bengoa-Vallejo; Marta Elena Losa-Iglesias; Daniel López-López; David Rodríguez-Sanz; Israel Casado-Hernández; Cesar Calvo-Lobo; Patricia Palomo-López
Journal:  J Med Internet Res       Date:  2019-12-02       Impact factor: 5.428

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.