Literature DB >> 10524820

Biomechanical test comparing the load to failure of the biodegradable meniscus arrow versus meniscal suture.

E K Song1, K B Lee.   

Abstract

The biodegradable Meniscus Arrow (Bionx, Blue Bell, PA) is a newly introduced method for repair of meniscal injury. However, because of a lack of studies, there is still little understanding of its biomechanical behavior. Biomechanical test was carried out to evaluate the load to failure and failure modes of the Meniscus Arrow and to compare the results with those of three traditionally used methods. Thirty-five lateral menisci obtained from Yorkshire pigs were incised longitudinally, simulating a peripheral longitudinal tear, and were repaired by means of Meniscus Arrows (one-point and two-point fixation), as well as with the techniques of knot-end, horizontal, and vertical suture using No. I PDS II monofilament suture. Tensile strength testing was performed according to each repair method with the Instron tensometer (Model No. 5569; Instron, Canton, MA) and its results were analyzed using a computerized statistical program. The average maximal tensile strengths were 113.9 +/- 14.6 N in vertical suture, 75.1 +/- 18.4 N in horizontal suture, 53.9 +/- 6.4 N in knot-end suture, 38.3 +/- 4.3 N in one-point fixation Meniscus Arrow repair, and 56.5 +/- 3.5 N in two-point fixation Meniscus Arrow repair. The initial failure strength for two-point fixation Meniscus Arrow repair is comparable to that of a knot-end suture. Following the results of this study, meticulous caution should be taken when meniscal tears are repaired with Meniscus Arrows.

Entities:  

Mesh:

Year:  1999        PMID: 10524820     DOI: 10.1016/s0749-8063(99)70004-6

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  11 in total

1.  A new rigid biodegradable anchor for meniscus refixation: biomechanical evaluation.

Authors:  Thore Zantop; Anne Kathleen Eggers; Volker Musahl; Andre Weimann; Joachim Hassenpflug; Wolf Petersen
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2003-12-19       Impact factor: 4.342

2.  Evaluation of the second-generation meniscus arrow in the fixation of bucket-handle tears in the vascular area of the meniscus. A prospective study of 20 patients with a mean follow-up of 26 months.

Authors:  Janne Sarimo; Jussi Rantanen; Tapio Tarvainen; Markku Härkönen; Sakari Orava
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2005-05-05       Impact factor: 4.342

3.  Pullout strength of meniscal repair after cyclic loading: comparison of vertical, horizontal, and oblique suture techniques.

Authors:  Yavuz Kocabey; Omer Taser; John Nyland; Mahmut Nedim Doral; Mehmet Demirhan; David N M Caborn; Sezgin Sarban
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2006-05-04       Impact factor: 4.342

4.  Modified cruciate suture technique for arthroscopic meniscal repair: a technical note.

Authors:  Ashraf Abdelkafy
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2007-02-13       Impact factor: 4.342

Review 5.  Meta-analysis on biomechanical properties of meniscus repairs: are devices better than sutures?

Authors:  Daniel M Buckland; Patrick Sadoghi; Matthias D Wimmer; Patrick Vavken; Geert I Pagenstert; Victor Valderrabano; Claudio Rosso
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-05-20       Impact factor: 4.342

6.  Short- to mid-term results of arthroscopic meniscal repair of long vertical longitudinal tears using combined cruciate and horizontal suture techniques: a retrospective study.

Authors:  Ashraf Abdelkafy
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-05-28

Review 7.  [Complications with all-inside devices used in reconstructive meniscal surgery].

Authors:  P Wilmes; O Lorbach; P Brogard; R Seil
Journal:  Orthopade       Date:  2008-11       Impact factor: 1.087

8.  Meniscal sutures: biomechanical study of "mulberry" and horizontal loop techniques.

Authors:  Fabiano Fantasia; Gabriele Potalivo; Giacomo Placella; Luigi Fantasia; Giuliano Cerulli
Journal:  J Orthop Traumatol       Date:  2011-12-22

9.  Biomechanical testing of transcapsular meniscal repair.

Authors:  Ryo Iuchi; Tatsuo Mae; Konsei Shino; Tomohiko Matsuo; Hideki Yoshikawa; Ken Nakata
Journal:  J Exp Orthop       Date:  2017-01-25

10.  FUNCTIONAL EVOLUTION OF MENISCAL REPAIR USING ABSORBABLE IMPLANTS.

Authors:  Waldo Lino Júnior
Journal:  Rev Bras Ortop       Date:  2015-11-16
View more