| Literature DB >> 10498511 |
Abstract
Electron beam output (dose/MU) is generally specified at the depth of maximum dose (zmax). The location of this point depends on beam energy, field size and field shape. Useful relationships have been developed to estimate zmax as a function of field size and beam energy. The formalism uses a pencil beam theory applied to circular fields.Mesh:
Year: 1999 PMID: 10498511 DOI: 10.1088/0031-9155/44/6/401
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609