Changes in the size and location of kidneys from the supine to standing positions and the implications for block placement during total body irradiation.

J E Reiff1, M Werner-Wasik, R K Valicenti, M S Huq.   

Abstract

PURPOSE: The use of total body irradiation (TBI) as a conditioning regimen for bone marrow transplantation often calls for partial transmission kidney blocks. These blocks are frequently designed based on the location of the kidneys during the abdominal computerized tomography (CT) scan. At our institution, TBI patients are treated in the standing position. As the kidneys can shift with different patient positions, a study was undertaken to evaluate the magnitude of the changes in the size and location of the kidneys from the supine CT position to the upright treatment position. METHODS AND MATERIALS: Intravenous contrast was administered to 15 patients. The patients were initially positioned supine on a simulator table and then positioned upright immediately in front of the image intensifier. PA radiographs were obtained with the patients in both positions. Changes in the size of the kidneys and their location relative to the vertebral bodies were noted.
RESULTS: In going from the supine to upright position, all the kidneys shifted inferiorly between 0.5 cm and 7.5 cm with an average of 3.6 cm. Most of the kidneys also shifted in the transverse dimension and incurred a change in width. The range of the transverse shift was from 0.9 cm in the lateral direction to 4.9 cm medially. The maximum width broadening was 1.2 cm and the maximum decrease in width was 1.8 cm.
CONCLUSIONS: When compared to the supine position, patients in the upright position show a dramatic inferior shift of the kidneys with other obvious, but less predictable, changes. For TBI treatments delivered in the upright position, kidney blocks should not be designed on the basis of supine abdominal CT scans.

Entities:  

Mesh:

Year:  1999        PMID: 10487569     DOI: 10.1016/s0360-3016(99)00208-4

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  4 in total

1.  The relative risk to the femoral nerve as a function of patient positioning: potential implications for trigger point dry needling of the iliacus muscle.

Authors:  Andrew M Ball; Michelle Finnegan; Shane Koppenhaver; Will Freres; Jan Dommerholt; Orlando Mayoral Del Moral; Carel Bron; Randy Moore; Erin E Ball; Emily E Gaffney
Journal:  J Man Manip Ther       Date:  2019-02-20

2.  Reappraisal of the classical abdominal anatomical landmarks using in vivo computerized tomography imaging.

Authors:  Ruchi Goyal; Anjali Aggarwal; Tulika Gupta; Ajay Gulati; Shallu Jaggi; S Ali Mirjalili; Daisy Sahni
Journal:  Surg Radiol Anat       Date:  2019-09-11       Impact factor: 1.246

3.  Renal shielding and dosimetry for patients with severe systemic sclerosis receiving immunoablation with total body irradiation in the scleroderma: cyclophosphamide or transplantation trial.

Authors:  Oana I Craciunescu; Beverly A Steffey; Chris R Kelsey; Nicole A Larrier; Cathy J Paarz-Largay; Robert G Prosnitz; Nelson Chao; John Chute; Cristina Gasparetto; Mitchell Horwitz; Gwynn Long; David Rizzieri; Keith M Sullivan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-08-26       Impact factor: 7.038

4.  A translational couch technique for total body irradiation.

Authors:  M Sarfaraz; C Yu; D J Chen; L Der
Journal:  J Appl Clin Med Phys       Date:  2001       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.