| Literature DB >> 10486757 |
E Dabek-Zlotorzynska1, E P Lai.
Abstract
The applicability of capillary electrochromatography (CEC) with photodiode array detection for the analysis of carbonyl hydrazones is presented. The CEC separation of thirteen hydrazones was optimized by a systematic variation of conditions using a commercially available CE system and a 3-micron porous C18-bonded silica capillary column. The separation profile obtained under optimal isocratic conditions (60% actonitrile-4% tetrahydrofuran-5 mM Tris, pH 8) is similar to those reported for gradient HPLC, with significant improvements in efficiency (to 150,000-250,000 theoretical plates/m) and analysis time (by a factor of 4). The retention time reproducibility is better than 0.2% (RSD) from run to run and 1% from day to day. The limits of detection for individual carbonyl hydrazones range between 0.1 and 0.5 microgram/ml. Applications to ambient air and automobile exhaust are shown.Entities:
Mesh:
Substances:
Year: 1999 PMID: 10486757 DOI: 10.1016/s0021-9673(99)00473-2
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759