Literature DB >> 10456971

The open circuit nitrogen washout technique for measuring the lung volume in infants: methodological aspects.

M G Morris1.   

Abstract

BACKGROUND: Lung volume measurement by nitrogen washout is widely used in infants, though a lack of accuracy and changes of calibration over time have been reported. The potential sources of error were explored in order to increase the accuracy and reliability of the technique.
METHODS: A commercial system for nitrogen washout and a 0.5 litre calibrating syringe as a lung model were used to perform over 2000 in vitro washouts, including simulated rapid breathing, shallow breathing, periodic breathing, sighs, and brief apnoeas. A constant 10 l/min bias flow of oxygen and extended equipment warming times were employed. A collapsible breathing bag was incorporated into the washout circuit. Following a single two point calibration, known air volumes from 42 ml to 492 ml were measured by nitrogen washout over a 14 hour period. The flow waveform in the nitrogen mixing chamber during a washout in vitro, with and without the breathing bag in the circuit, was also studied.
RESULTS: The mean coefficient of variation of all volumes was 0.66%. The mean difference between measured and known volumes was 0.30 ml (95% confidence interval (CI) -0.18 to 0.79). This difference was not statistically significant (p = 0.22). The mean percentage error was -0.1% (range -0.47% to 0.46%). Nitrogen calibration remained stable for 14 hours. Without the breathing bag flow transients were frequent in the mixing chamber during in vitro washout.
CONCLUSIONS: This technique increases the accuracy in vitro and the precision in vivo of volume measurement by nitrogen washout. Sources of potential errors including baseline drifting and inadequate equipment warming times were identified. The breathing bag acted as a buffer reservoir, preventing large swings in flows within the nitrogen mixing chamber during washouts, and should be an integral component of the nitrogen washout circuit.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10456971      PMCID: PMC1745573          DOI: 10.1136/thx.54.9.790

Source DB:  PubMed          Journal:  Thorax        ISSN: 0040-6376            Impact factor:   9.139


  5 in total

1.  Pneumonia due to Legionella pneumophila and pneumococcal pneumonia: similarities and differences on presentation.

Authors:  A Granados; D Podzamczer; F Gudiol; F Manresa
Journal:  Eur Respir J       Date:  1989-02       Impact factor: 16.671

2.  An automated bedside method for measuring functional residual capacity by N2 washout in mechanically ventilated children.

Authors:  Y Sivan; T W Deakers; C J Newth
Journal:  Pediatr Res       Date:  1990-11       Impact factor: 3.756

3.  Lung volume measurements in wheezy infants: comparison of plethysmography and gas dilution.

Authors:  E Eber; B Steinbrugger; M Modl; E Weinhandl; M S Zach
Journal:  Eur Respir J       Date:  1994-11       Impact factor: 16.671

4.  Comparison of nitrogen washout and plethysmographic measurements of lung volume in healthy infants.

Authors:  M Gappa; M E Fletcher; C A Dezateux; J Stocks
Journal:  Am Rev Respir Dis       Date:  1993-12

5.  Functional residual capacity in normal neonates and children up to 5 years of age determined by a N2 washout method.

Authors:  T Gerhardt; L Reifenberg; D Hehre; R Feller; E Bancalari
Journal:  Pediatr Res       Date:  1986-07       Impact factor: 3.756

  5 in total
  4 in total

1.  A novel non-invasive technique for measuring the residual lung volume by nitrogen washout with rapid thoracoabdominal compression in infants.

Authors:  M G Morris
Journal:  Thorax       Date:  1999-10       Impact factor: 9.139

2.  Automated evaluation of functional residual capacity by oxygen washout.

Authors:  Wolfgang Eichler; Jan Schumacher; Angela Roth-Isigkeit; Jörg Braun; Hermann Kuppe; Karl-Friedrich Klotz
Journal:  J Clin Monit Comput       Date:  2002 Apr-May       Impact factor: 2.502

3.  A novel physiological investigation of the functional residual capacity by the bias flow nitrogen washout technique in infants.

Authors:  Mohy G Morris
Journal:  Pediatr Pulmonol       Date:  2009-07

4.  Comprehensive integrated spirometry using raised volume passive and forced expirations and multiple-breath nitrogen washout in infants.

Authors:  Mohy G Morris
Journal:  Respir Physiol Neurobiol       Date:  2009-11-06       Impact factor: 1.931

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.