Effects of scatter substraction on detection and quantitation in hepatic SPECT.

D J de Vries1, M A King, E J Soares, B M Tsui, C E Metz.   

Abstract

UNLABELLED: The purpose of this investigation was to examine the effects of subtractive scatter compensation methods on lesion detection and quantitation.
METHODS: Receiver operating characteristic (ROC) methodology was used to measure human observer detection accuracy for tumors in the liver using synthetic images. Furthermore, ROC results were compared with mathematical models for detection and activity quantitation to examine (a) the potential for predicting human performance and (b) the relationship between the detection and quantitation tasks. Images with both low and high amounts of scatter were compared with the ideal case of images of primary photons only (i.e., perfect scatter rejection) and with images corrected by subtracting a scatter image estimated by the dual photopeak window method.
RESULTS: With low contrast tumors in a low count background, the results showed that scatter subtraction improved quantitation but did not produce statistically significant increases in detection accuracy. However, primary images did produce some statistically significant improvements in detection accuracy when compared with uncorrected images, particularly for high levels of scatter.
CONCLUSION: Although scatter subtraction methods may provide improved activity quantitation, they may not significantly improve detection for liver SPECT. The results imply that significant improvement in detection accuracy for the conditions tested may depend on the development of gamma cameras with better scatter rejection.

Entities:  

Mesh:

Year:  1999        PMID: 10452320

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  2 in total

Review 1.  Scatter modelling and compensation in emission tomography.

Authors:  Habib Zaidi; Kenneth F Koral
Journal:  Eur J Nucl Med Mol Imaging       Date:  2004-03-31       Impact factor: 9.236

2.  Effects of motion, attenuation, and scatter corrections on gated cardiac SPECT reconstruction.

Authors:  Xiaofeng Niu; Yongyi Yang; Mingwu Jin; Miles N Wernick; Michael A King
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.