In vivo assessment of stent expansion and recoil in normal porcine coronary arteries: differential outcome by stent design.

J P Carrozza1, S E Hosley, D J Cohen, D S Baim.   

Abstract

BACKGROUND: Despite the routine use of high pressure, coronary stents generally fail to achieve a cross-sectional area (CSA) >60% to 80% of the nominal CSA of their dilating balloon. The extent to which incomplete balloon expansion and postdeflation stent recoil contribute to this failure has not been fully evaluated. METHODS AND
RESULTS: Thirty-two stents (8 Gianturco-Roubin II, 8 Palmaz-Schatz, 8 MultiLink, and 8 NIR) were deployed in nondiseased coronary arteries of 8 Yorkshire pigs. All stents were then expanded according to 1 of 3 balloon strategies: appropriately sized compliant balloons, oversized compliant balloons, or oversized noncompliant balloons. Continuous ultrasound imaging was performed during stepwise balloon inflation and deflation, with an 0.018-in imaging core positioned within the guidewire lumen of the balloon. In these normal arteries, balloon underexpansion relative to the nominal size was not observed. After balloon deflation, however, all stents showed significant recoil from their maximum inflated CSA. Recoil was significantly greater for the coil Gianturco-Roubin II stent (30% CSA) than for the 3 slotted-tube stent designs (15% to 17% CSA).
CONCLUSIONS: In normal, compliant coronary arteries, stent recoil is the predominant mechanism by which stents fail to achieve the nominal CSA of their dilating balloon. The magnitude of this in vivo stent recoil is significantly greater than reported from bench testing and varies with stent design (coil versus slotted tube). Postdilatation strategies that result in controlled stent overexpansion are needed to overcome this recoil phenomenon and result in a final stent CSA that approximates the reference artery CSA.

Entities:  

Mesh:

Year:  1999        PMID: 10449699     DOI: 10.1161/01.cir.100.7.756

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  4 in total

1.  Coronary artery overexpansion increases neointimal hyperplasia after stent placement in a porcine model.

Authors:  Robert J Russo; Patricia D Silva; Mark Yeager
Journal:  Heart       Date:  2007-07-16       Impact factor: 5.994

2.  Coronary stenting: why size matters.

Authors:  Edoardo De Benedetti; Philip Urban
Journal:  Heart       Date:  2007-12       Impact factor: 5.994

3.  Coronary artery stretch versus deep injury in the development of in-stent neointima.

Authors:  J Gunn; N Arnold; K H Chan; L Shepherd; D C Cumberland; D C Crossman
Journal:  Heart       Date:  2002-10       Impact factor: 5.994

4.  Over-expansion of drug-eluting stents in patients with left main coronary artery disease: An in vivo study.

Authors:  Yicong Ye; Hao Qian; Ming Yang; Xueqing Zhu; Tianpeng Gan; Shuyang Zhang; Yong Zeng
Journal:  J Int Med Res       Date:  2017-06-12       Impact factor: 1.671

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.