Perturbation of radiotherapy beams by radiographic film: measurements and Monte Carlo simulations.

N Suchowerska1, P Hoban, A Davison, P Metcalfe.   

Abstract

Radiographic film is an established practical tool used in the measurement of the dose distribution for radiotherapy purposes. The accuracy and reproducibility of film optical density as an indicator of dose has been associated with several factors including photon energy, processing conditions and film plane orientation. Few studies have investigated the factors causing variability in film dosimetry, due to the difficulty of separating the individual contributions. The effect that a sheet of radiographic film in a water phantom has on its response to a 6 MV photon and a cobalt-60 teletherapy beam, when orientated perpendicular and parallel to the beam central axis, is reported. Monte Carlo generated spectra were used to calculate collision kerma (Kcoll) for water and film elements. Measured and calculated results indicate a potential over-response at 25 cm depth of the order of 14 +/- 2.4% and 18 +/- 6.0% respectively for 6 MV photons and 15 +/- 3.4% and 32 +/- 4.5% respectively for a cobalt beam. For film exposed parallel as compared to perpendicular to the central axis of the beam, the calculated results suggest an explanation in terms of the predominantly forward directed secondary electrons for the measured difference in film response at depth. It is proposed that the difference in response of the parallel as compared to perpendicular exposed film be due to the predominantly 'upstream' photon interactions giving rise to energy deposition in film. The simulations indicate that the variation with depth of relative energy imparted in film and water elements correlates with the observed variation in film response with depth.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10442711     DOI: 10.1088/0031-9155/44/7/314

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Pilot study for compact microbeam radiation therapy using a carbon nanotube field emission micro-CT scanner.

Authors:  Mike Hadsell; Guohua Cao; Jian Zhang; Laurel Burk; Torsten Schreiber; Eric Schreiber; Sha Chang; Jianping Lu; Otto Zhou
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

2.  Analysis of variation in calibration curves for Kodak XV radiographic film using model-based parameters.

Authors:  Shu-Hui Hsu; Ravi Kulasekere; Peter L Roberson
Journal:  J Appl Clin Med Phys       Date:  2010-08-05       Impact factor: 2.102

3.  Monte Carlo simulations of EBT3 film dose deposition for percentage depth dose (PDD) curve evaluation.

Authors:  Spencer M Robinson; Nolan Esplen; Derek Wells; Magdalena Bazalova-Carter
Journal:  J Appl Clin Med Phys       Date:  2020-11-06       Impact factor: 2.102