Atrial defibrillation with a transvenous lead: a randomized comparison of active can shocking pathways.

M Cooklin1, M R Olsovsky, R G Brockman, S R Shorofsky, M R Gold.   

Abstract

OBJECTIVES: The purpose of this study was to compare transvenous atrial defibrillation thresholds with lead configurations consisting of an active left pectoral electrode and either single or dual transvenous coils.
BACKGROUND: Low atrial defibrillation thresholds are achieved using complex lead systems including coils in the coronary sinus. However, the efficacy of more simple ventricular defibrillation leads with active pectoral pulse generators to defibrillate atrial fibrillation (AF) is unknown.
METHODS: This study was a prospective, randomized assessment of shock configuration on atrial defibrillation thresholds in 32 patients. The lead system was a dual coil Endotak DSP lead with a left pectoral pulse generator emulator. Shocks were delivered either between the right ventricular coil and an active can in common with the proximal atrial coil (triad) or between the atrial coil and active can (transatrial).
RESULTS: Delivered energy at defibrillation threshold was 7.1 +/- 6.0 J in the transatrial configuration and 4.0 +/- 4.2 J in the triad configuration (p < 0.005). Moreover, a low threshold (< or = 3 J) was observed in 69% of subjects in the triad configuration but only 47% in the transatrial configuration. Peak voltage and shock impedance were also lowered significantly in the triad configuration. Left atrial size was the only clinical predictor of the defibrillation threshold (r = 0.57, p < 0.002).
CONCLUSIONS: These results indicate that low atrial defibrillation thresholds can be achieved using a single-pass transvenous ventricular defibrillation lead with a conventional ventricular defibrillation pathway. These data support the development of the combined atrial and ventricular defibrillator system.

Entities:  

Mesh:

Year:  1999        PMID: 10440146     DOI: 10.1016/s0735-1097(99)00214-4

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  3 in total

Review 1.  Dual- versus single-coil implantable defibrillator leads: review of the literature.

Authors:  Jörg Neuzner; Jörg Carlsson
Journal:  Clin Res Cardiol       Date:  2012-01-10       Impact factor: 5.460

Review 2.  Is An Atrial Defibrillator Still An Option In Treating Patients With Atrial Fibrillation?

Authors:  Ziad El Khoury; Deepak Bhakta
Journal:  J Atr Fibrillation       Date:  2013-02-12

3.  Implantable dual-chamber defibrillator for the selective treatment of spontaneous atrial and ventricular arrhythmias: arrhythmia incidence and device performance.

Authors:  Andreas Schuchert; Giuseppe Boriani; Christian Wollmann; Mauro Biffi; Martin Kühl; Johannes Sperzel; Sascha Stiller; Gianni Gasparini; Dirk Böcker
Journal:  J Interv Card Electrophysiol       Date:  2005-03       Impact factor: 1.900

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.