| Literature DB >> 10423738 |
G De Luca1, S Stampi, L Lezzi, F Zanetti.
Abstract
Two different decontamination systems, heat and acid, and two isolation media, GVPC and MWY agar were tested for the recovery of Legionella pneumophila from drinking water. The samples were concentrated by filtration through 0.2 micron polyamide filter and the membranes were resuspended in the original water samples. The suspension was divided into three parts: the first was placed in a 50 degrees C water bath, the second was acidified in HCl-KCl solution and the third did not undergo any treatment. The isolation was made by means of media containing charcoal, yeast extract and glycine with cycloeximide (GVPC) or vancomycin, polimixin B, anysomicin and dyes (MWY). Heating at 50 degrees C for 30 minutes was seen to be the best decontamination system above all when used with GVPC agar. Moreover, with this pretreatment higher counts were obtained both on MWY and GVPC agar. The MWY agar produced the highest isolatin percentages and the highest counts.Entities:
Mesh:
Substances:
Year: 1999 PMID: 10423738
Source DB: PubMed Journal: New Microbiol ISSN: 1121-7138 Impact factor: 2.479