Interactions between electronic article surveillance systems and implantable cardioverter-defibrillators.

W J Groh1, S A Boschee, E D Engelstein, W M Miles, M E Burton, P R Foster, B J Crevey, D P Zipes.   

Abstract

BACKGROUND: In patients with implantable cardioverter-defibrillators (ICDs). inappropriate shocks have been reported with exposure to electronic article surveillance systems. The risk to patients with ICDs of walking through or lingering near surveillance systems requires further investigation. METHODS AND
RESULTS: We evaluated the response in ICD function in 170 subjects during a 10- to 15-second midgate walk-through of and during extreme (2 minutes within 6 in of the gate) exposure to 3 common article surveillance systems. Complete testing was done in 169 subjects. During a 10- to 15-second (very slow) walk-through of the 3 surveillance systems, no interactions were observed that would negatively affect ICD function. During extreme exposure (169 subjects) and during extreme exposure and pacing via the ICD (126 subjects), interactions between the ICD and the article surveillance systems were observed in 19 subjects. In 7 subjects, this interaction was clinically relevant and would have likely (3 subjects) and possibly (4 subjects) resulted in ICD shocks. In 12 subjects, the interaction was minor.
CONCLUSIONS: It is safe for a patient with an ICD to walk through electronic article surveillance systems. Lingering in a surveillance system may result in an inappropriate ICD shock.

Entities:  

Mesh:

Year:  1999        PMID: 10421599     DOI: 10.1161/01.cir.100.4.387

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  5 in total

1.  Effects of mobile telephones on the function of implantable cardioverter defibrillators.

Authors:  Izzet Tandogan; Bulent Ozin; Huseyin Bozbas; Sibel Turhan; Ramazan Ozdemir; Ertan Yetkin; Ergun Topal
Journal:  Ann Noninvasive Electrocardiol       Date:  2005-10       Impact factor: 1.468

2.  Electromagnetic interference in patients with implanted cardioverter-defibrillators and implantable loop recorders.

Authors:  Marcos de Sousa; Gunnar Klein; Thomas Korte; Michael Niehaus
Journal:  Indian Pacing Electrophysiol J       Date:  2002-07-01

Review 3.  [Electromagnetic interference : Pacemakers, cardiac resynchronization therapy devices, implantable cardioverter-defibrillator].

Authors:  Bernd Nowak; Oliver Przibille; Andreas Napp
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2022-07-04

4.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

Authors:  Roy Beinart; Saman Nazarian
Journal:  Circulation       Date:  2013-12-24       Impact factor: 29.690

5.  Digital music players cause interference with interrogation telemetry for pacemakers and implantable cardioverter-defibrillators without affecting device function.

Authors:  Gregory Webster; Ligia Jordao; Maria Martuscello; Tarun Mahajan; Mark E Alexander; Frank Cecchin; John K Triedman; Edward P Walsh; Charles I Berul
Journal:  Heart Rhythm       Date:  2008-03-04       Impact factor: 6.343

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.