Gated TI-201 myocardial perfusion SPECT: application of energy window optimization.

S S Wadhwa1, R Mansberg, D Wilkinson, D Abbati.   

Abstract

PURPOSE: TI-201 scintigraphy is plagued with poor image quality because of the low-energy photons of TI-201 decay. Traditionally, a narrow 20% window centered on 71-72 keV has been used to improve sensitivity. Recent studies indicate that better imaging may be possible by optimizing the energy window to 34% centered on 77 keV. In this study, energy window optimization (EWO) was applied to gated TI-201 myocardial perfusion SPECT, and myocardial functional parameters were compared for gated TI-201 SPECT and gated Tc-99m sestamibi (Tc-99m MIBI) SPECT.
METHODS: Count statistics for standard and optimal TI-201 myocardial scintigraphy were noted in 25 patients by assessing the total counts in a mid-ventricular slice of a rest-gated TI-201 myocardial SPECT study. The feasibility of performing functional studies with the application of EWO to TI-201 was assessed using the count statistics of a mid-ventricular slice of an optimized gated TI-201 SPECT study and a gated Tc-99m MIBI SPECT study. The functional parameters (ejection fraction, wall motion, and thickening) of TI-201 with EWO and Tc-99m MIBI were compared in 60 patients who underwent rest-gated TI-201 SPECT followed by poststress gated Tc-99m MIBI SPECT. The left ventricular ejection fraction was calculated using commercially available software, whereas wall thickness and motion were assessed by the consensus of two readers.
RESULTS: The application of EWO increased available counts by more than 25%. It also resulted in sufficient counts being available to perform gated TI-201 SPECT without increasing acquisition times or the dose of TI-201. The average ejection fraction was 60.4% for gated TI-201 SPECT and 59.6% for gated Tc-99m MIBI SPECT (not significantly different). Overall, the image quality was rated excellent in 12% for TI-201 and Tc-99m MIBI and good in 50% and 62%, respectively, and poor in 38% and 26%, respectively.
CONCLUSION: The application of EWO to TI-201 SPECT allows myocardial functional parameters to be assessed without having to increase the acquisition times or the administered dose of TI-201.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10401998     DOI: 10.1097/00003072-199907000-00001

Source DB:  PubMed          Journal:  Clin Nucl Med        ISSN: 0363-9762            Impact factor:   7.794


  1 in total

1.  Assessment of ejection fraction with Tl-201 gated SPECT in myocardial infarction: Precision in a rest-redistribution study and accuracy versus planar angiography.

Authors:  E Itti; J Rosso; P Damien; M Auffret; J P Thirion; M Meignan
Journal:  J Nucl Cardiol       Date:  2001 Jan-Feb       Impact factor: 5.952

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.