Commissioning, clinical implementation and quality assurance of Siemen's Virtual Wedge.

S Rathee1, C B Kwok, C MacGillivray, M Mirzaei.   

Abstract

This report presents the results of commissioning, clinical implementation and quality assurance of Siemens Virtual Wedge. Our measurements show that: (1) wedge factors are within 2% of unity, (2) percentage depth doses are within 1% of open beam data, and (3) wedged beam profiles can be modeled similarly to a physical wedge and follow a well defined equation to facilitate modeling of an arbitrary wedge angle. The gantry angle dependence of wedge profiles is similar to open beam profiles. The output of wedged fields is linear with delivered monitor units within 1%. Quality assurance results indicate the wedge profiles are very stable over time. Day to day variations of two points measured along the wedge gradient direction are within 1.5%.

Mesh:

Year:  1999        PMID: 10379513     DOI: 10.1016/s0958-3947(99)00003-5

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  3 in total

1.  Comparison of dosimetric characteristics of physical and enhanced dynamic wedges.

Authors:  Sathiyan Saminathan; Ravikumar Manickam; Sanjay S Supe
Journal:  Rep Pract Oncol Radiother       Date:  2011-08-15

2.  The accuracy of dynamic wedge dose computation in the ADAC Pinnacle RTP system.

Authors:  H Shao; X Wu; C Luo; A Crooks; A Bernstein; A Markoe
Journal:  J Appl Clin Med Phys       Date:  2004-10-01       Impact factor: 2.102

3.  Behavior of the Siemens Virtual Wedge following an interruption to beam delivery.

Authors:  N D Richmond; C P Walker
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.