A comparison of bronchodilator therapy delivered by nebulization and metered-dose inhaler in mechanically ventilated patients.

P Marik1, J Hogan, J Krikorian.   

Abstract

BACKGROUND: The optimal method of delivering bronchodilators in mechanically ventilated patients is unclear. The purpose of this study was to compare the pulmonary bioavailability of albuterol delivered by the nebulizer, the metered-dose inhaler (MDI) and spacer, and the right-angle MDI adaptor in ventilated patients using urinary analysis of drug levels.
METHODS: Mechanically ventilated patients who had not received a bronchodilator in the previous 48 h and who had normal renal function were randomized to receive the following: (1) five puffs (450 microg) of albuterol delivered by the MDI with a small volume spacer; (2) five puffs of albuterol delivered by the MDI port on a right-angle adaptor; or (3) 2.5 mg albuterol delivered by a nebulizer. Urine was collected 6 h after the administration of the drug, and the amounts of albuterol and its sulfate conjugate were determined in the urine by a chromatographic assay.
RESULTS: Thirty patients were studied, 10 in each group: their mean age and serum creatinine level were 62 years and 1.3 mg/dL, respectively. With the MDI and spacer, (mean +/- SD) 169+/-129 microg albuterol (38%) was recovered in the urine; with the nebulizer, 409+/-515 microg albuterol (16%) was recovered in the urine; and with the MDI port on the right-angle adaptor, 41+/-61 microg albuterol (9%) was recovered in the urine (p = 0.02 between groups). The level of albuterol in the urine was below the level of detection in four patients in whom the drug was delivered using the right-angle MDI adaptor.
CONCLUSION: The three delivery systems varied markedly in their efficiency of drug delivery to the lung. As previous studies have confirmed, this study has demonstrated that using an MDI and spacer is an efficient method for delivering inhaled bronchodilators to the lung. The pulmonary bioavailability was poor with the right-angle MDI port. This port should not be used to deliver bronchodilators in mechanically ventilated patients.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10378564     DOI: 10.1378/chest.115.6.1653

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  9 in total

Review 1.  Aerosol delivery via invasive ventilation: a narrative review.

Authors:  Hui-Ling Lin; James B Fink; Huiqing Ge
Journal:  Ann Transl Med       Date:  2021-04

Review 2.  Pharmacologic treatment of the adult hospitalized asthma patient.

Authors:  M L Kreutzer; S Louie
Journal:  Clin Rev Allergy Immunol       Date:  2001-06       Impact factor: 8.667

3.  In vitro evaluation of aerosol bronchodilator delivery during mechanical ventilation: pressure-control vs. volume control ventilation.

Authors:  Dean R Hess; Christine Dillman; Robert M Kacmarek
Journal:  Intensive Care Med       Date:  2003-05-15       Impact factor: 17.440

Review 4.  Metered dose inhalers versus nebulizers for aerosol bronchodilator delivery for adult patients receiving mechanical ventilation in critical care units.

Authors:  Agi Holland; Fiona Smith; Kay Penny; Gill McCrossan; Linda Veitch; Caroline Nicholson
Journal:  Cochrane Database Syst Rev       Date:  2013-06-06

Review 5.  Bronchodilator delivery with metered-dose inhaler during mechanical ventilation.

Authors:  D Georgopoulos; E Mouloudi; E Kondili; M Klimathianaki
Journal:  Crit Care       Date:  2000-07-11       Impact factor: 9.097

6.  The Clinical Practice and Best Aerosol Delivery Location in Intubated and Mechanically Ventilated Patients: A Randomized Clinical Trial.

Authors:  Chuanlin Zhang; Jie Mi; Zeju Zhang; Xueqin Wang; Yunxiao Zhu; Xinyi Luo; Ruiying Gan; Xiaoya Chen; Yujun Zou
Journal:  Biomed Res Int       Date:  2021-04-03       Impact factor: 3.411

7.  The Bioavailability of Salbutamol in Urine via Volumatic and Nonvolumatic Valved Holding Chambers.

Authors:  Fanak Fahimi; Farzad Kobarfard; Jamshid Salamzadeh; Atefeh Fakharian; Pegah Abdolahi; Azita Hajhossein Talasaz; Hamid Mahboobi Pour; Shadi Baniasadi; Mohammadreza Masjedi
Journal:  World Allergy Organ J       Date:  2011-11-18       Impact factor: 4.084

Review 8.  Inhalation therapy in mechanical ventilation.

Authors:  Juçara Gasparetto Maccari; Cassiano Teixeira; Marcelo Basso Gazzana; Augusto Savi; Felippe Leopoldo Dexheimer-Neto; Marli Maria Knorst
Journal:  J Bras Pneumol       Date:  2015 Sep-Oct       Impact factor: 2.624

Review 9.  Aerosol delivery during invasive mechanical ventilation: a systematic review.

Authors:  Jonathan Dugernier; Stephan Ehrmann; Thierry Sottiaux; Jean Roeseler; Xavier Wittebole; Thierry Dugernier; François Jamar; Pierre-François Laterre; Gregory Reychler
Journal:  Crit Care       Date:  2017-10-21       Impact factor: 9.097

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.