Analytical error of home glucose monitors: a comparison of 18 systems.

R N Johnson1, J R Baker.   

Abstract

In a quality review of 18 home blood glucose monitors, we measured the imprecision and incidence of significant error of 12 colorimetric and six amperometric systems by using capillary blood specimens from patients attending diabetes clinics. Imprecision at the mean glucose concentration found in the respective blood specimens (about 9 mmol/L) gave coefficients of variation (CV) ranging from 5.2 to 22.8%. Eight monitors including five of amperometric design had a CV of less than 10%. The incidence of significant error (defined as the proportion of specimens differing in value by more than 15% from a reference hexokinase assay of glucose in capillary blood) varied from 6 to 76%. Among the eight monitors identified as being most precise, the majority produced results that differed markedly from the reference assay, underlining the need for a common approach to calibration of home glucose monitors.