Influence of high-energy photons on the spectrum of iodine-123 with low- and medium-energy collimators: consequences for imaging with 123I-labelled compounds in clinical practice.

A A Dobbeleir1, A S Hambÿe, P R Franken.   

Abstract

Using iodine-123 labelled radiotracers, the presence of 2.5% high-energy photons causes image deterioration due to increased scatter. To investigate the influence of these photons on image quality, we measured the spectrum of 123I with a medium-energy (ME), a low-energy all-purpose (LEAP) and a low-energy high-resolution (LEHR) collimator. Even in air, using low-energy collimators a high baseline activity was observed over the total energy detection range of the gamma camera. The 159-keV photopeak to scatter activity ratio fell from 5.9 for ME to 3.6 and 2.9 for LE collimators. Acquisition of images with LEHR collimators with energy windows set at 159 keV and 500 keV demonstrated that the 159-keV LEHR image is a combination of the ME image of the object and of the LEHR 500-keV image. Because of their important septal penetration and greater geometric detection efficiency compared with the 159-keV photons of 123I, the contribution of high-energy photons is dependent on the source-detector distance. For a small source placed in air, the scatter to photopeak activities varied from 17.4% at 80 cm to 37.8% at 5 cm distance from an LEHR collimator. Considering only the scatter problem, ME collimators are the best choice for 123I studies. When using LE collimators for high-resolution tomography with 123I-labelled compounds, scatter contribution from high-energy photons has to be corrected for quantitative analysis or when dual-isotope studies are performed, whether or not these studies are acquired simultaneously.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10369952     DOI: 10.1007/s002590050434

Source DB:  PubMed          Journal:  Eur J Nucl Med        ISSN: 0340-6997


  24 in total

1.  Collimator choice in cardiac SPECT with I-123-labeled tracers.

Authors:  Yusuke Inoue; Ichiro Shirouzu; Toru Machida; Yasunori Yoshizawa; Fumihide Akita; Manabu Minami; Kuni Ohtomo
Journal:  J Nucl Cardiol       Date:  2004 Jul-Aug       Impact factor: 5.952

2.  Influence of collimator choice and simulated clinical conditions on 123I-MIBG heart/mediastinum ratios: a phantom study.

Authors:  Hein J Verberne; Christel Feenstra; W Matthijs de Jong; G Aernout Somsen; Berthe L F van Eck-Smit; Ellinor Busemann Sokole
Journal:  Eur J Nucl Med Mol Imaging       Date:  2005-05-18       Impact factor: 9.236

3.  Optimized acquisition and processing protocols for I-123 cardiac SPECT imaging.

Authors:  Ji Chen; Ernest V Garcia; James R Galt; Russell D Folks; Ignasi Carrio
Journal:  J Nucl Cardiol       Date:  2006 Mar-Apr       Impact factor: 5.952

4.  Degradation of myocardial perfusion SPECT images caused by contaminants in thallous (201Tl) chloride.

Authors:  Steven G Staelens; Tim C de Wit; Ignace A Lemahieu; Freek J Beekman
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-01-25       Impact factor: 9.236

5.  How to use myocardial 123I-MIBG scintigraphy in chronic heart failure.

Authors:  Denis Agostini; Ignasi Carrio; Hein J Verberne
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-04       Impact factor: 9.236

6.  Calibration of gamma camera systems for a multicentre European ¹²³I-FP-CIT SPECT normal database.

Authors:  Livia Tossici-Bolt; John C Dickson; Terez Sera; Robin de Nijs; Maria Claudia Bagnara; Catherine Jonsson; Egon Scheepers; Felicia Zito; Anita Seese; Pierre Malick Koulibaly; Ozlem L Kapucu; Michel Koole; Maria Raith; Jean George; Markus Nowak Lonsdale; Wolfgang Münzing; Klaus Tatsch; Andrea Varrone
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-04-06       Impact factor: 9.236

7.  123I-mIBG and the phantom tollbooth.

Authors:  Albert Flotats; Ignasi Carrió
Journal:  J Nucl Cardiol       Date:  2017-03-13       Impact factor: 5.952

8.  ASNC imaging guidelines for SPECT nuclear cardiology procedures: Stress, protocols, and tracers.

Authors:  Milena J Henzlova; W Lane Duvall; Andrew J Einstein; Mark I Travin; Hein J Verberne
Journal:  J Nucl Cardiol       Date:  2016-06       Impact factor: 5.952

9.  EANM procedure guidelines for brain neurotransmission SPECT using (123)I-labelled dopamine transporter ligands, version 2.

Authors:  Jacques Darcourt; Jan Booij; Klaus Tatsch; Andrea Varrone; Thierry Vander Borght; Ozlem L Kapucu; Kjell Någren; Flavio Nobili; Zuzana Walker; Koen Van Laere
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-02       Impact factor: 9.236

10.  Effect of collimator choice on quantitative assessment of cardiac iodine 123 MIBG uptake.

Authors:  Yusuke Inoue; Akira Suzuki; Ichiro Shirouzu; Toru Machida; Yasunori Yoshizawa; Fumihide Akita; Satoshi Ohnishi; Kohki Yoshikawa; Kuni Ohtomo
Journal:  J Nucl Cardiol       Date:  2003 Nov-Dec       Impact factor: 5.952

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.