Literature DB >> 10352593

Plain and covered stent-grafts: in vitro evaluation of characteristics at three-dimensional MR angiography.

P R Hilfiker1, H H Quick, J F Debatin.   

Abstract

PURPOSE: To evaluate the characteristics of various plain and covered stents as regards demonstration of the contained lumen with simulated contrast material-enhanced fast three-dimensional (3D) magnetic resonance (MR) angiography.
MATERIALS AND METHODS: Six stents (Easy Wallstent, Cragg, Palmaz, Cragg EndoPro System 1, Corvita, Passager) were implanted in plastic tubes and integrated into a closed-tubing circuit driven by a pulsatile roller flow pump. The circulating water was spiked with gadopentetate dimeglumine. Three-dimensional MR angiograms were obtained with an echo time of 1.4 or 2.1 msec and with partial (0.5 signal acquired) or full (one signal acquired) k-space sampling. The size of the stent-associated artifact was determined relative to the tube diameter.
RESULTS: The Easy Wallstent and Palmaz stent caused complete obliteration of the stent lumen with all sequences. The Cragg, Cragg EndoPro System 1, and Passager stents allowed good visualization of the stent lumen. The Corvita stent demonstrated major artifacts. The magnitude of the stent-associated artifact was related to the echo time (P < .01) but not to the type of k-space sampling (P = .35).
CONCLUSION: The luminal patency of selected plain and covered stents can be assessed with contrast-enhanced 3D MR angiography.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10352593     DOI: 10.1148/radiology.211.3.r99jn23693

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  3 in total

1.  Stent-assisted coil placement in a wide-necked persistent trigeminal artery aneurysm with jailing of the trigeminal artery: a case report.

Authors:  Mohammed I Mohammed; Johnny S Sandhu; Ajay K Wakhloo
Journal:  AJNR Am J Neuroradiol       Date:  2002-03       Impact factor: 3.825

2.  MR-velocity mapping in vascular stents to assess peak systolic velocity. In vitro comparison of various stent designs made of Stainless Steel and Nitinol.

Authors:  Jacqueline van Holten; Patrik Kunz; Paul G H Mulder; Peter M T Pattynama; Hildo J Lamb; Lukas C van Dijk
Journal:  MAGMA       Date:  2002-11       Impact factor: 2.310

3.  Evaluation of various image reconstruction parameters in lower extremity stents using multidetector-row CT angiography: initial findings.

Authors:  Martin Heuschmid; Benjamin Wiesinger; Gunnar Tepe; Oliver Luz; Andreas F Kopp; Claus D Claussen; Stephan H Duda
Journal:  Eur Radiol       Date:  2006-05-30       Impact factor: 5.315

  3 in total