Cardiovascular collapse caused by electrocardiographically silent 60-Hz intracardiac leakage current. Implications for electrical safety.

C D Swerdlow1, W H Olson, M E O'Connor, D M Gallik, R A Malkin, M Laks.   

Abstract

BACKGROUND: The national standard for safe 60-Hz intracardiac leakage current under a single-fault condition is 50 microA. This standard is intended to protect patients from alternating current (AC) at levels below the threshold for sensation, but the minimum unsafe level for AC in closed-chest humans is not known. To determine this value, we studied 40 patients at testing of implantable cardioverter-defibrillators using a programmable source of 60-Hz AC. METHODS AND
RESULTS: We applied AC for 5-second test periods in increasing strengths until ventricular fibrillation (VF) was induced or 1 mA was reached. Two current paths were tested: bipolar, between tip and ring electrodes of a right ventricular pacing catheter, and unipolar, from tip to a remote electrode. We observed a characteristic sequence of 3 responses as AC was increased: (1) intermittent ventricular capture with QRS morphology identical to pacing through the electrodes (minimum value, 20 microA); (2) continuous capture at cycle length 282+/-88 ms (minimum value, 32 microA); and (3) VF persisting after AC termination (minimum value, 49 microA). Continuous capture caused loss of pulsatile arterial pressure and cardiovascular collapse (mean arterial pressure, 32+/-8 mm Hg) for the duration of AC with no ECG evidence of AC stimulation. Thus, the clinical picture was that of hypotensive ventricular tachycardia (VT). The continuous-capture threshold was </=50 microA in 9 patients (22%) for bipolar AC and in 5 (12%) for unipolar AC. All patients showed continuous capture over a wide range for both bipolar AC (68+/-18 to 216+/-238 microA) and unipolar AC (84+/-27 to 278+/-226 microA).
CONCLUSIONS: Leakage current causes cardiovascular collapse at levels below the VF threshold. Stimulation by silent AC that is neither felt nor visible on the ECG presents as hypotensive VT. In patients with intracardiac electrodes, leakage current less than or equal to the present standard of 50 microA may cause VT or VF. The safety standard for leakage current lasting >/=5 seconds should be </=20 microA. This standard should be based on the continuous-capture threshold.

Entities:  

Mesh:

Year:  1999        PMID: 10330388     DOI: 10.1161/01.cir.99.19.2559

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  9 in total

1.  Excitation of a cardiac muscle fiber by extracellularly applied sinusoidal current.

Authors:  E J Vigmond; N A Trayanova; R A Malkin
Journal:  J Cardiovasc Electrophysiol       Date:  2001-10

2.  Conductively coupled flexible silicon electronic systems for chronic neural electrophysiology.

Authors:  Jinghua Li; Enming Song; Chia-Han Chiang; Ki Jun Yu; Jahyun Koo; Haina Du; Yishan Zhong; Mackenna Hill; Charles Wang; Jize Zhang; Yisong Chen; Limei Tian; Yiding Zhong; Guanhua Fang; Jonathan Viventi; John A Rogers
Journal:  Proc Natl Acad Sci U S A       Date:  2018-09-18       Impact factor: 11.205

3.  Capacitively Coupled Arrays of Multiplexed Flexible Silicon Transistors for Long-Term Cardiac Electrophysiology.

Authors:  Hui Fang; Ki Jun Yu; Christopher Gloschat; Zijian Yang; Chia-Han Chiang; Jianing Zhao; Sang Min Won; Siyi Xu; Michael Trumpis; Yiding Zhong; Enming Song; Seung Won Han; Yeguang Xue; Dong Xu; Gert Cauwenberghs; Matthew Kay; Yonggang Huang; Jonathan Viventi; Igor R Efimov; John A Rogers
Journal:  Nat Biomed Eng       Date:  2017-03-01       Impact factor: 25.671

4.  Ventricular fibrillation: are swine a sensitive species?

Authors:  Gregory P Walcott; Mark W Kroll; Raymond E Ideker
Journal:  J Interv Card Electrophysiol       Date:  2015-01-16       Impact factor: 1.900

5.  Frequency dependence of the cardiac threshold to alternating current between 10 Hz and 160 Hz.

Authors:  R A Malkin; A de Jongh Curry
Journal:  Med Biol Eng Comput       Date:  2003-11       Impact factor: 2.602

6.  Ventricular fibrillation caused by electrocoagulation in monopolar mode during laparoscopic subphrenic mass resection.

Authors:  Chun-Yan Yan; Xiu-Jun Cai; Yi-Fan Wang; Hong Yu
Journal:  Surg Endosc       Date:  2010-06-30       Impact factor: 4.584

7.  Invited Article: Emerging soft bioelectronics for cardiac health diagnosis and treatment.

Authors:  Faheem Ershad; Kyoseung Sim; Anish Thukral; Yu Shrike Zhang; Cunjiang Yu
Journal:  APL Mater       Date:  2019-03-01       Impact factor: 5.096

8.  A compact pulse-modulation cold air plasma jet for the inactivation of chronic wound bacteria: development and characterization.

Authors:  Phuthidhorn Thana; Apiwat Wijaikhum; Pipath Poramapijitwat; Chakkrapong Kuensaen; Jomkhwan Meerak; Athipong Ngamjarurojana; Sureeporn Sarapirom; Dheerawan Boonyawan
Journal:  Heliyon       Date:  2019-09-13

9.  In situ diagnosis and simultaneous treatment of cardiac diseases using a single-device platform.

Authors:  Jae Chul Hwang; Moohyun Kim; Sumin Kim; Hunkyu Seo; Soohwan An; Eui Hwa Jang; Seung Yeop Han; Mi Jung Kim; Nam Kyun Kim; Seung-Woo Cho; Sak Lee; Jang-Ung Park
Journal:  Sci Adv       Date:  2022-09-14       Impact factor: 14.957

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.