Faheem Ershad1, Kyoseung Sim2, Anish Thukral3, Yu Shrike Zhang4, Cunjiang Yu. 1. Department of Biomedical Engineering, University of Houston, Houston, Texas 77204, USA. 2. Department of Mechanical Engineering, University of Houston, Houston, Texas 77204, USA. 3. Materials Science and Engineering Program, University of Houston, Houston, Texas 77204, USA. 4. Division of Engineering in Medicine, Department of Medicine, Brigham and Women's Hospital, Harvard Medical School, Cambridge, Massachusetts 02139, USA.
Abstract
Cardiovascular diseases are among the leading causes of death worldwide. Conventional technologies for diagnosing and treating lack the compliance and comfort necessary for those living with life-threatening conditions. Soft electronics presents a promising outlet for conformal, flexible, and stretchable devices that can overcome the mechanical mismatch that is often associated with conventional technologies. Here, we review the various methods in which electronics have been made flexible and stretchable, to better interface with the human body, both externally with the skin and internally with the outer surface of the heart. Then, we review soft, wearable, noninvasive heart monitors designed to be attached to the chest or other parts of the body for mechano-acoustic and electrophysiological sensing. A common method of treatment for various abnormal heart rhythms involves catheter ablation procedures and we review the current soft bioelectronics that can be placed on the balloon or head of the catheter. Cardiac mapping is integral to determine the state of the heart; we discuss the various parameters for sensing aside from electrophysiological sensing, such as temperature, pH, strain, and tactile sensing. Finally, we review the soft devices that harvest energy from the natural and spontaneous beating of the heart by converting its mechanical motion into electrical energy to power implants.
Cardiovascular diseases are among the leading causes of death worldwide. Conventional technologies for diagnosing and treating lack the compliance and comfort necessary for those living with life-threatening conditions. Soft electronics presents a promising outlet for conformal, flexible, and stretchable devices that can overcome the mechanical mismatch that is often associated with conventional technologies. Here, we review the various methods in which electronics have been made flexible and stretchable, to better interface with the human body, both externally with the skin and internally with the outer surface of the heart. Then, we review soft, wearable, noninvasive heart monitors designed to be attached to the chest or other parts of the body for mechano-acoustic and electrophysiological sensing. A common method of treatment for various abnormal heart rhythms involves catheter ablation procedures and we review the current soft bioelectronics that can be placed on the balloon or head of the catheter. Cardiac mapping is integral to determine the state of the heart; we discuss the various parameters for sensing aside from electrophysiological sensing, such as temperature, pH, strain, and tactile sensing. Finally, we review the soft devices that harvest energy from the natural and spontaneous beating of the heart by converting its mechanical motion into electrical energy to power implants.
The heart is the body’s mechanical pump, constantly pushing oxygenated and deoxygenated blood
throughout the body as it endlessly contracts and relaxes the atria and ventricles. As one of
the most vital organs of the body, the heart must be continuously monitored for those
afflicted by heart diseases. Nearly 41 × 106 individuals globally have experienced
heart failure. Greater than 5 ×
106 individuals are affected (in the developed world alone) by atrial
fibrillation (AF) specifically, which has been notorious for stroke in patients with cardiac
diseases. The number of patients with
complex arrhythmias is expected to rapidly increase in the near future. To better understand the origins of heart diseases and
clearly corroborate the need for soft electronics in cardiac sensing, the mechanism of blood
flow and the electrical conduction system of the heart are detailed.The cardiac cycle consists of two phases: systole, the contraction of the heart to pump
blood, and diastole, the relaxation of the heart after contraction. General blood flow in the
heart is as follows. Diastole involves the movement of blood lacking in oxygen flowing in from
the vena cava to the right atrium, in which the chordae tendineae open the tricuspid
valve. The right atrium contracts, causing
the deoxygenated blood to flow into the right ventricle (RV), which subsequently contracts
(now in systole) and pushes the blood through the pulmonary artery. This blood is transported
to the lungs. Blood returning from the lungs (oxygenated) through the pulmonary vein enters
the left atrium, which contracts and pushes blood through the bicuspid (or mitral) valve, into
the left ventricle. This ventricle contracts
(systole), directing blood through the aortic valve and then to the aorta, which is
responsible for distributing nutrient-rich blood to the rest of the body. The valves that
operate in the heart open/close in response to gradients in pressure. Ascertaining proper blood flow is critical in diseased patients
and numerous fatal conditions can arise, aside from myocardial infarction. Systolic heart
failure can occur if the pressure in the left ventricle is not enough to close the mitral
valve. If the mitral valve does not close
properly, then regurgitation, or backward flow of blood, from left ventricular contraction can
occur. Aortic stenosis can also lead to
improper blood flow as the heart must pump more to push blood through a narrowed aortic
valve. Ischemia, or inadequate supply of
blood to organs or parts of the body, can occur if blood flow is impeded. Thrombosis and/or embolization (blood clotting) can
significantly hinder blood flow as well. During beating, the heart expands to 145% of its
contraction volume and the inner chambers have been shown to place 20%-30% strain on
interfaced electronics.The mechanism of contraction and relaxation occurs in conjunction with a special conduction
system that maintains the persistent beating of the heart [Fig.
1(a)]. Cardiomyocytes are the excitable cells which compose the cardiac muscle (atria
and ventricles) while cardiac pacemaker cells initiate the electrical impulse. Between all
these cells are intercalated discs (junctions) that bridge the cells and allow for the
propagation of electrical impulses from one cell to another. These discs allow for the passage
of various ions that are responsible for the depolarization and repolarization events of the
heart, which are recordable and visualized in the electrocardiogram [Fig. 1(b)]. The sinoatrial (SA) node, located in the right atrium, contains
pacemaker cells; the heart rate directly depends on the rate of the action potential produced
by these cells. The pacemaker cells spontaneously depolarize, and this depolarization spreads
through the atria. Depolarization at the SA node and in the right atrium corresponds to the P
wave of the typical electrocardiography (ECG) waveform and occurs due to the flow of sodium
ions into the cells. After depolarization, sodium continues to enter the cells resulting in an
overshoot, which leads to the sodium channels’ closing and calcium channels’ opening, once the
membrane potential of the cell reaches a certain threshold. The impulse from the SA node
propagates to the atrioventricular (AV) node. From the AV node, the impulse travels through
the bundle of His and eventually reaches the Purkinje fibers, which contact the ventricular
cardiomyocytes. Following the depolarization of the atrium is the depolarization of the
ventricles and repolarization of the atria; this can be observed with the characteristic QRS
peak found in the ECG waveform. Finally, the T wave in the ECG represents the repolarization
of the ventricles. From the ECG, various diseases can be diagnosed, though a majority of these
diseases form a class of diseases known as arrythmias, or irregular beating of the heart. Some examples are P wave asystole (or no
ventricular activity/blood flow, only P wave in ECG), tachycardias (abnormally high heart
rate, only QRS is shown or lengthened), ventricular fibrillation (inconsistent and fast
beating, no identifiable wave can be observed), atrial fibrillation (inconsistent heartbeats,
no pattern between QRS complexes and no P wave), or other arrhythmias. The information that
can be provided from mechano-acoustic and electrophysiological signals is discussed below.
FIG. 1.
Mechanism of blood flow (a) with arrows indicating the direction of blood flow in the
heart. Light blue regions indicate areas through which deoxygenated blood flows, while
pink regions indicate where oxygenated blood flows. The origin of the characteristic electrocardiography waveform (b)
with the major depolarizations from which the sub-waves (P, QRS, T) arise.
Mechanism of blood flow (a) with arrows indicating the direction of blood flow in the
heart. Light blue regions indicate areas through which deoxygenated blood flows, while
pink regions indicate where oxygenated blood flows. The origin of the characteristic electrocardiography waveform (b)
with the major depolarizations from which the sub-waves (P, QRS, T) arise.Mechano-acoustic signals can disclose information about heart sounds, which are critical to
determine the proper heart function. Heart sounds indicate the closure of heart valves,
although they are ∼30 mm beneath the surface of the chest. Still, heart sounds can be used to
verify correct blood flow between the chambers of the heart, even from the skin. Methods of measuring these types of signals
include ballistocardiography (BCG, a measure of recoil forces from the body as the heart pumps
blood through blood vessels), phonocardiography (PCG, a measure of sounds the heart makes),
and seismocardiography (SCG, a measure of chest vibrations due to heartbeat). Photoplethysmography (PPG) is used to measure
the changes in blood volume in peripheral blood vessels, and it provides information about the
timing of cardiac cycles. PPG has been used in place of ECG in some cases although it has
limited signal accuracy. In
addition to these measurement techniques, ventricular assist devices have been developed to
act as support systems for those who have experienced heart failure; they are mechanical pumps
that help the heart to pump blood from the ventricles to the rest of the body.
Mechano-acoustic signals can supplement the information provided by ECG and can sometimes even
present information that cannot be directly found in ECG, specifically in the case of heart
sounds. Continuing to the
electrophysiological side of diagnosis, ECG is a critical tool for both noninvasively and
invasively diagnosing many disorders associated with the heart. By understanding the mechanism
behind the production of the ECG signal, it is possible to determine abnormal heart conditions
by observing the signals found in the ECG. Holter monitoring devices with multiple leads are
the current clinical standard for recording ECG. One common method of treatment for
abnormal heart rhythms is radiofrequency ablation. Cells that are thought to prevent the
appropriate propagation of the impulse through the conduction pathway in the heart are
destroyed by heating or freezing using catheters. In addition to characterizing the blood flow
and electrical activity of the heart, temperature, strain, and pH are other vital parameters
that should be monitored.Accelerometers have been used in bulky/rigid formats to detect mechano-acoustic signals. However, these have usually been strapped to
the user. In addition to discomfort from the mechanical coupling of the device with skin, the
conventional technology could have reduced sensitivity to more subtle movements that have
physiological consequences, exist in large form factors which prevent optimal placement of the
device, and only perform single-mode sensing. Typically, rigid ventricular assist devices
(VADs) are used to work in place of one or both ventricles of the heart for patients who are
waiting for heart transplants. However, since they create artificial surfaces in the body,
they put patients at a greater risk of stroke and thromboembolic events; blood-thinning
medications need to be offered to the patients requiring VADs. External devices have been shown to work
against the native cardiac mechanics and but they cannot always fully synchronize with cardiac
contraction. These sensors, as well as
ECG monitors, must both be considered.Current ECG monitors lack full compliance with the human body. Rigid, flat substrates were commonly used for
recording ECG in isolated cell culture studies and mapping impulse propagation. Although some recent
technologies exploit flexible substrates, the measurement interface is not always
optimal.
Skin-mounted, soft, and imperceptible devices have been shown to be more compliant while also
providing greater details about cardiac health. A survey from a recent study shows that patients overwhelmingly prefer
a skin-mounted and flexible sensor over the conventionally used Holter monitor for comfortable
daily usage. In addition, there are needs
for other types of sensors (such as pH sensors and tactile sensors) to be implemented in
flexible and stretchable formats.In order for bioelectronics to intimately interface with the beating heart or skin, they need
to be designed and constructed into flexible and stretchable formats. In other words, they
must comprise flexibility and/or stretchability. Various strategies exist to induce these two
characteristics and the underlying physics has been studied extensively. The requirement for mechanical flexibility usually
can be met by making the materials and devices thin. For example, consider the case of
plastics, such as polyimide (PI), a commonly utilized substrate for flexible electronics, the
bending stiffness is directly proportional to the cubic power of the thickness of the
material. Thus, reducing the
thickness of the substrate by a factor of 10 will lead to a bending stiffness reduced by a
factor of 1000. If a device is sufficiently bendable or flexible and minimizes the conformal
energy, then it can achieve conformal contact with the surface of the skin or heart.For the case of electronics that are compliant with the human skin, conformal contact can be
achieved if the device is less than 3.8-µm thick; this is due to sufficiently
low contact pressure on skin with average roughness. But again, this also depends on the roughness of the skin, bending
energy of the substrate, and interfacial adhesion energy. The rougher the skin is, the thinner the device
will need to be to account for the higher interfacial pressure. Some examples of conformal and
flexible electronics are shown in Figs. 2(a) and 2(b).
FIG. 2.
Examples of soft electronics as shown in (a) thin film mesh (thickness: 2.5
µm) reduces adhesion energy and allows for the device to wrap over a
glass beam, the inset shows the device in an unwrapped state, but it is too thick to
conform to the surface; (b)
ultra-flexible thin plastic electronics: the left panel shows the structure of the thin
film transistor (scale bar, 20 nm), and the right panel shows the device crumpled like
paper (scale bar, 1 cm); (c)
controlled wrinkling of stretchable Si ribbons on 50% pre-strained PDMS; (d) stretchable Si-based complementary
metal-oxide semiconductor (CMOS) electronics on PDMS, images show tensile strains on the
device in vertical (top) and horizontal (bottom) directions; (e) kirigami structures showing in-plane stretchability,
% stretch from bottom to top row: 43%, 62%, and 79%, and the inset (bottom right) shows
how kirigami structures rotate and stretch; (f) upper frames show the CMOS inverter and lower frames show
finite element modeling of stretched interconnects; and (g) intrinsically stretchable semiconductor
[poly(3-hexylthiophene-2,5-diyl), P3HT, in PDMS] based sensor at different strains in top
frames and the corresponding AFM phase mode image of P3HT fibers in bottom frames.
Examples of soft electronics as shown in (a) thin film mesh (thickness: 2.5
µm) reduces adhesion energy and allows for the device to wrap over a
glass beam, the inset shows the device in an unwrapped state, but it is too thick to
conform to the surface; (b)
ultra-flexible thin plastic electronics: the left panel shows the structure of the thin
film transistor (scale bar, 20 nm), and the right panel shows the device crumpled like
paper (scale bar, 1 cm); (c)
controlled wrinkling of stretchable Si ribbons on 50% pre-strained PDMS; (d) stretchable Si-based complementary
metal-oxide semiconductor (CMOS) electronics on PDMS, images show tensile strains on the
device in vertical (top) and horizontal (bottom) directions; (e) kirigami structures showing in-plane stretchability,
% stretch from bottom to top row: 43%, 62%, and 79%, and the inset (bottom right) shows
how kirigami structures rotate and stretch; (f) upper frames show the CMOS inverter and lower frames show
finite element modeling of stretched interconnects; and (g) intrinsically stretchable semiconductor
[poly(3-hexylthiophene-2,5-diyl), P3HT, in PDMS] based sensor at different strains in top
frames and the corresponding AFM phase mode image of P3HT fibers in bottom frames.To induce stretchability, a more complex characteristic to achieve than flexibility, multiple
strategies have been implemented. Most of the strategies for stretchability can be simplified
into two approaches. For conventionally non-stretchable materials, forms of structural
engineering are required, while the other approach involves utilizing intrinsically
stretchable materials. In making non-stretchable materials stretchable, one strategy involves
creating buckled/wavy structures and taking advantage of out-of-plane motion to account for
in-plane strain. By bonding thin
layers of materials to an elastomeric substrate, such as polydimethylsiloxane (PDMS), that
undergoes pre-strain and then is released, a wavy structure with distinct amplitudes and
wavelengths can be created.
Figures 2(c) and 2(d) show the success of this strategy to develop stretchable electronics from rigid
Si.Altering the material properties and pre-strain can tune the amplitude and wavelength of the
wavy structure. With greater pre-strain, larger amplitude waves with shorter wavelengths can
be formed. Like in an accordion, the wavy structures experience changes in their wavelength
and amplitude to account for stretching.
Instead of exploiting out-of-plane wavy structures, in-plane wavy or deformable structures
have also been employed to provide mechanical stretchability. For instance, stretchable serpentine
interconnects as bridges can be formed to connect strain-sensitive electronics as islands that
are bonded to the substrate. This is an
effective method to delocalize strain away from the passive/active sensing components of a
given device. Other strategies for in-plane wavy and deformable structures have been realized
as well [Figs. 2(e) and 2(f)]. These include
coiled springs with rotatable islands, leaf-arm springs, and serpentine-shaped
interconnects. Using fractal architecture is another strategy for attaining
in-plane stretchability. Advanced methods
can be used to allow for twisting, rotating, bending, and buckling both in-plane and
out-of-plane. Kirigami, which involves
strategically creating arrays of cuts on a material, is a concept that has been implemented to
enable reduced stress in buckling/folding applications, and also for providing stretchability
both in-plane and out-of-plane. For
interfacing with the heart, in-plane structures may be more beneficial [Fig. 2(e)]. Implementing intrinsically stretchable materials on
elastomeric substrates is another common approach for achieving stretchability as structural
engineering can be complicated and costly.
Although the development for fully intrinsically stretchable devices that combine
intrinsically stretchable conductors, dielectrics, and semiconductor materials is in progress,
individual components have been studied [Fig. 2(g)]. Intrinsically stretchable
interconnects, electrodes, and piezoresistive sensors are the most investigated components so
far.An emerging class of bioelectronics, namely, epidermal electronics, which can be
skin-mountable, compliant, and even mechanically imperceptible to the wearers, overcomes the
outstanding challenge of mechanical mismatches between electronics and the skin. Primarily, these sensors are used for
monitoring the electrical potentials generated during the heartbeat at the surface of the
skin. In some applications, these sensors are capable of acting as stethoscopes, a ubiquitous
tool for analyzing heart sounds found in all physician’s offices today. The studies here
present various devices and their applications for noninvasive heart monitoring.Liu et al. fabricated mechano-acoustic and electrophysiological sensors on a
single device platform using stretchable electronics that conform to skin [Fig. 3(a)]. The devices were 2 mm thick, have moduli of ∼31 kPa, and bending
stiffnesses of ∼1 µNm. The sensing platform had a core/shell structure in
which the core combined serpentinecopper interconnects, an accelerometer, hardware filters
(resistors and capacitors), and an amplifier at the neutral mechanical plane (NMP) between two
layers of PI encapsulation, all of which were embedded in a soft elastomer (Silbione).
Eco-flex acted as the shell. This type of structure both isolated the main components from
stress at the interface of the skin and restricted the movement of interconnects. Au
electrodes and an anisotropic conductive film (ACF) connector existed at openings of the
core/shell devices. An additional layer of Silbione underneath the devices served as the
adhesive layer. After confirming that their device was biocompatible through a cell viability
test, they accounted for mechanical impedance by minimizing the mass of the device, which
resulted in lower mechanical loading at the skin. They mounted the device onto the sternum of
a healthy humanpatient and a diseased patient, for simultaneous SCG and ECG measurements
[Figs. 3(b) and 3(c)]. For the healthy patient, it was observed that the mechano-acoustic signals were
similar in quality to those of the conventional stethoscope. For diseased patients who were
diagnosed with cardiac stenosis/regurgitation, the device captured abnormal heart sounds at
the pulmonary and tricuspid areas. Finally, the mechano-acoustic device was laminated onto a
left ventricular assist device (LVAD). LVADs are intended to be temporary solutions for those
waiting for heart transplants. In a simulated experiment, a blood clot was entered into the
air valve of the LVAD, and the mechano-acoustic device registered additional abnormal
frequencies on top of the frequencies associated with regular LVAD operation. In addition to
local heart monitoring, remote monitoring can be incredibly beneficial to patients.
FIG. 3.
Mechano-acoustic sensor on (a) the sternum, lower frames show twisting and compression of
the device; (b) ECG and heart sounds
recorded from the sensor; (c) zoomed
in from (b) showing sub-waves of ECG; MC, mitral valve closed; AO, aortic valve opened;
RE, rapid ejection from ventricle; AC, aortic valve closed; MO, mitral valve opened; RF,
rapid filling of ventricles; time between S1 and S2 is the systolic
time of left ventricle; (d) schematic
of the WiSP device, right frames shows wireless data transmission (top) and twisting
(bottom); (e) survey data from a
hospital comparing comfortability of the WiSP and Holter monitors; (f) Ag embedded in adhesive PDMS electrodes on skin; (g) recordings from electrodes in
(f); and (h) zoomed in recording of
(g).
Mechano-acoustic sensor on (a) the sternum, lower frames show twisting and compression of
the device; (b) ECG and heart sounds
recorded from the sensor; (c) zoomed
in from (b) showing sub-waves of ECG; MC, mitral valve closed; AO, aortic valve opened;
RE, rapid ejection from ventricle; AC, aortic valve closed; MO, mitral valve opened; RF,
rapid filling of ventricles; time between S1 and S2 is the systolic
time of left ventricle; (d) schematic
of the WiSP device, right frames shows wireless data transmission (top) and twisting
(bottom); (e) survey data from a
hospital comparing comfortability of the WiSP and Holter monitors; (f) Ag embedded in adhesive PDMS electrodes on skin; (g) recordings from electrodes in
(f); and (h) zoomed in recording of
(g).Lee et al. created a bandage-like device (referred to as WiSP from here on)
that is flexible, wearable, and used for remote monitoring of the heart. The WiSP epidermal sensor logged heart rate, ECG, and
harvested energy for power through near-field communication, while it also had its own custom
battery. In addition, the sensor wirelessly streamed data to smartphones with near-field
communication (NFC) capabilities [Fig. 3(d)].
Polyurethane (PU) encapsulated a flexible PI substrate which held the flexible antenna,
battery, NFC chip, and circuitry. At the bottom of the device, a tape adhesive had exposed
areas for Au contact pads coated with conductive hydrogel. The electrodes were spaced
appropriately so that the entirety of the typical ECG waveform could be captured. The
thickness of the PI was optimized to 0.2 mm from 2 mm to allow for conformal contact with
skin. When a smartphone was brought within 3 cm of the WiSP, the device was activated. The
smartphone was used to control the functionality of the WiSP and securely streamed data from
the patient to the caregiver. For healthy subjects, the WiSP ECG waveforms were comparable in
quality to those of a commercial device from all standard 3-lead ECG measurement positions on
the body (lead I, II, and III). For remote monitoring of healthy subjects, they let the
patients resume daily life at home with the WiSP attached to their skin. They also attached a
commercial, wearable strap used for heart monitoring. The comparison of the data showed high
correlation between the WiSP and commercial monitor, suggesting that the WiSP performed well
during regular usage. The WiSP was then implemented in a remote monitoring study of subjects
with atrial fibrillation, in which both the WiSP and Holter monitors recorded heart rate data
for comparison. Atrial fibrillation episodes were evident and accurately detected from both
devices. A survey taken at a hospital showed that the WiSP is more comfortable in both daily
use and sleep in comparison to the Holter monitor [Fig. 3(e)]. In addition to flexible and stretchable devices, optically transparent
devices could be useful for noninvasive monitoring, both functionally and aesthetically.Park and co-workers developed stretchable electrodes with high adhesion and optical
transmittance for strain sensing and ECG recording from the skin [Fig. 3(f)]. The electrodes
were fabricated by embedding Ag nanowires (AgNWs) in adhesive PDMS (a-PDMS) by first spin
coating AgNWs on glass, then coating the AgNW with the a-PDMS, and finally releasing the film
from the glass. They tuned the adhesiveness, Young’s modulus, and failure strain of PDMS by
varying the weight percent of a nonionic surfactant, Triton X, when mixing with the PDMS
elastomer base. Varying the curing temperature and the addition of 4 wt. % Triton (referred to
as a4-PDMS) enhanced the mechanical properties of the PDMS. By differing the curing
temperature, they were able to determine that a4-PDMS cured at 40 °C demonstrated the best
Young’s modulus (40 kPa) and over 400% strain at failure. This is exceptional for applications
on human skin, which has a modulus between 500 kPa and 2 MPa, because a lower Young’s modulus
increases the interfacial contact area between the electrode and skin, leading to greater
conformal attachment and more comfortable user experience. From their peeling test, the
adhesion force of a4-PDMS was found to be 6-fold greater than that of unmodified PDMS. This
was a result of better spreading of polymer chains and improved wetting, leading to greater
surface contact. In the case of optical transparency, a4-PDMS showed greater than 80%
transmittance, although it was less transparent than non-modified PDMS. Considering all the
other benefits of the a4-PDMS, its optical transparency still permits it to be potentially
used in optogenetic studies involving pacing experiments should the device be capable of
adhering to the exterior surface of the heart. The stability of this adhesive PDMS was examined and after 2 months of
storage, Young’s modulus was about 51 kPa and the failure strain was ∼400%. After embedding
the AgNW (now referred to as a4-PDMS_NW), the stretchable conductor was found to have a sheet
resistance of 35 Ω/sq with a transmittance of 75%. The relative resistance change was found to
decrease after 2000 cycles of stretching at 15% strain; the AgNWs changed their alignment to
that of the stretching direction. To examine the performance of the a4-PDMS_AgNW conductor,
they compared its ECG recording capability against commercial gel electrodes. Using a
three-electrode setup on the chest and ribcage, they recorded the ECG. The a4-PDMS-AgNW showed
negligible noise, even less than that of the commercial electrodes. The ECG signal clearly
contained the expected P, QRS, and T waves [Figs. 3(g)
and 3(h)]. The clarity of the signal can be associated
with the low impedance and conformal attachment of the electrodes to the skin.Balloon catheters are powerful minimally invasive tools that can house multiple
functionalities. The shaft of the catheter can be inserted through small incisions in the
body. The configuration of the balloon can be controlled by the material choice. Catheters can
be inserted into multiple areas of the heart, but generally, those areas are in the epicardium
(inner layer of the pericardium, which is the outer layer covering the heart) or the
endocardium (inner walls of the chambers in the heart). In angioplasty, these catheters can
inflate coronary arteries to remove blockages, and in septostomy, the balloons provide greater
force in widening the area between the atria to allow for more efficient pumping in the right
side of the heart. Soft
catheter electronics have been fabricated by releasing ultrathin metal and/or inorganic
materials from their substrate onto the balloons in various ways, the main method being
transfer printing. PI or
epoxies bolster adhesion, allowing for many inflating/deflating cycles without loss of
performance from the sensors. Special layouts of the sensing materials and/or interconnects
need to be implemented to account for the 20%-30% strain created by contraction/relaxation of
the heart. Catheters can also be used for radiofrequency (RF) ablation therapy, which is a
technique that involves the electrical stimulation, cryoablation, or laser ablation of
individual, problematic cells that cause arrhythmias like AF. Once these cells are destroyed, the
normal conduction pathways can be restored, leading to a stabilization of the heart rate,
which can be observed in the ECG. During these procedures, temperatures greater than ∼40 °C
can cause tissue damage. The purpose of
the soft electronics on the balloon is to conformally contact the surface and chambers of the
heart, which are curvilinear. For the devices on the balloon, placement of the sensors and
actuators at the NMP limits the concentration of strain on the sensors. Current balloon
catheters cannot yield information about mechanical contact, local temperature, and the
electrical state of the tissue. The electrodes used for conventional catheters also lack the
spatial resolution needed to capture the rapid waves that compose the ECG, with electrode
densities around 0.1 electrodes/cm2. In addition to improving blood flow, catheters
can provide quantitative physiological information from the interfaced tissue, should they be
equipped with the appropriate sensors; these sensors are discussed here.Kim et al. deployed flexible and stretchable electronics on a
multifunctional balloon catheter that surmount the limitations of conventional catheters which
do not provide information regarding blood flow, contact pressure, lesion area, and
temperature [Fig. 4(a)]. The strain resistance of the devices can be attributed to the fact
that they were placed at the nodes of the mesh, which reduced the mechanical coupling to
strain as the balloon expanded and contracted and to the serpentine layouts and
interconnections. This catheter consisted of micro-tactile sensors, temperature sensors, RF
ablation electrodes, blood flow sensors, recording electrodes, and active semiconductor
devices. The micro-tactile sensors were used to quantify the mechanical forces on the heart
tissue and were in non-coplanar serpentine layouts with the sensors placed at nodes which
experience minimal strain (<1%), even as the substrate underwent large deformations. They
were sensitive enough to capture the normal forces without destroying the soft tissue. The
sensors were located on thick layers of SU8 (epoxy) and had an electrically conductive silicon
rubber (pressure-sensitive rubber, PSR) which overlaid PDMS. PI encapsulated this structure.
Lateral tensile strain was shown to modulate the resistance of PSR; this was the working
mechanism of the micro-tactile sensor. The temperature sensor, similar in design to the
micro-tactile sensor, was made of Pt and had a resistance change of 1.91 Ω/°C; the resistance
of the interconnects contributed negligibly to the measurement. Temperature sensing is
critical as heat is released from current flowing between the active and ground electrodes
during ablation. With this sensor, it is possible to detect the lesion size in the radial and
thickness directions. Such information can greatly inform a surgeon about the progress of the
surgery. The blood flow sensor detected changes in blood flow as resistance changes. With
greater flow rate, the resistance generally increased. The RF ablation electrodes,
electrophysiological recording electrodes, and LEDs were all also integrated into the same
platform. In their in vivo experiment, the device was inserted into the
epicardium of rat and rabbit hearts. ECG recordings showed elevated S-T intervals at the RV
and some regions of the left ventricle (LV). In evaluating the S-T interval, inflation-induced
injuries from the balloon were deduced. This could be imperative to understanding the contact
pressures of inflating the balloon in the endocardium as well. The micro-tactile sensors also
provided a method to monitor balloon inflation. In vivo ablation was
successfully performed as they were able to control the size of the lesion using the RF
ablation electrodes and temperature sensors. Finally, they modeled a disease state and
demonstrated that the electrodes functioned from the onset of the injury, leading to obvious
indicators of disease in the ECG. Follow-up recordings at other sites after this induced
injury also showed abnormalities. The sensors presented in this highly integrated and soft
platform can also be used on surgical gloves to probe more sensitive areas during cardiac
surgery, such as the posterior myocardium. In another study, Klinker et al.
developed a similar device with a blood flow sensor on a balloon catheter that also deployed
stretchable electrodes for recording and ablation [Fig. 4(b)]. A balloon catheter at
different inflation states is shown in Fig. 4(c).
Balloons catheters with higher-density electrodes could provide information that current
mapping electrodes for catheters do not have the resolution to do.
FIG. 4.
Balloon catheter with (a) stretchable electronics on the rabbit heart; (b) electrodes, a thermistor, and a
heater on the balloon and in the catheter shaft; (c) deflation states of a polyurethane balloon with ablation
and sensing electrodes; (d) an
impedance contact sensor on a balloon;
and (e) data from impedance sensor showing contact and no contact states.
Balloon catheter with (a) stretchable electronics on the rabbit heart; (b) electrodes, a thermistor, and a
heater on the balloon and in the catheter shaft; (c) deflation states of a polyurethaneballoon with ablation
and sensing electrodes; (d) an
impedance contact sensor on a balloon;
and (e) data from impedance sensor showing contact and no contact states.Kim et al. improved the ECG recording capability by implementing a
flexible/stretchable, high-density, multiplexed array with active electrodes and temperature,
pressure, and impedance sensors onto a balloon catheter. These high-density electrodes with a spatial density of 15
electrodes/cm2 are necessary, especially in the case of AF in which the
electrical events of the heart are not well understood or defined, leading to reduced
precision of the ablation treatment. With a greater number of electrodes, the need for more
connections and wires could add bulk to the catheter shaft. Instead of taking that route, the
group developed a distributed multiplexing circuit in which source followers (buffers) and
complementary metal-oxide semiconductor (CMOS) switches that controlled the output of the
active electrode. Using this strategy and controlling lines (rows/columns) of
electrodes/amplifiers, a 64-electrode device was implemented on a balloon catheter with a
reduced number (16) of inputs/outputs for high-density and high-resolution ECG recording. The
impedance sensor [Fig. 4(d)] shows the expected
performance, as shown in Fig. 4(e).Implantable soft cardiac devices are necessary for real-time analysis, and in
vitro/isolated cellular studies have provided insight into the function of the
heart. However, functional behavior at the organ level has not been extensively studied,
primarily due to the lack of competent tools, specifically for the in vivo
case. Soft, multifunctional cardiac electronics can provide multimodal information that can
assess fully remodeled disease states in order to augment the understanding of the heart’s
functionality. Optical mapping, pH
detection, temperature sensing, and strain sensing allow for a more comprehensive approach to
invasive monitoring; these modes of sensing are discussed here.Xu et al. printed heart models and fabricated
three-dimensional-multifunctional integumentary membranes (3D-MIMs) that provided the
capability to record electrophysiological signals, strain, optically map, and detect pH and
temperature from the exterior surface of a perfused rabbit heart [Fig. 5(a)].
Two-dimensional (2D) flexible sheets of semiconductor electronics were converted into 3D
elastic membranes that mimic the shape of the heart. Fabrication involved 3D printing a rat
heart model and development of the sensors on planar substrates. The sensors were transferred
to a cured layer of silicone elastomer that surrounded the 3D-printed heart. The 3D-MIM
consisted of μ-ILEDs (inorganic LEDs), strain gauges, sensing/stimulation electrodes, pH
sensors, and temperature sensors/heaters. The membranes experienced pressures of the native
pericardium. From their mechanical analysis, there was no indication of
induced ischemia. They performed spatiotemporal cardiac measurements on a perfused rabbit
heart. 68 Au electrodes (total surface area 1 mm2, 3.5-mm spacing) comprised part
of the multifunctional sensor array. The array was placed over the epicardium and was
transparent, allowing for optical mapping and verification of the ECG recording. The optical
mapping data and electrical measurements were shown to be highly correlated [Fig. 5(b)]. The temperature sensor consisted of 16 serpentine
traces made of Au. A linear response of the temperature sensor to the range of physiological
temperatures was observed, and they had average responses of about 1.23 Ω/°C. The function of
this sensor was tested by changing the temperature of the perfusion liquid and decreased
temperatures (and consequently, decreased heart rate) were recorded with lower temperature
perfusate [Fig. 5(c)]. This was confirmed with their ECG
recording, which further indicated normal heart rhythms resuming after the original perfusion
temperature was restored. 32 pH sensors were fabricated by depositing IrOx on Au.
Per unit of pH, a 68.9-mV potential response was recorded. pH mapping was demonstrated to
track the progression of induced ischemia; they stopped the perfusion (passage of fluid)
through the rabbit heart and observed that pH decreased from 7.4 (baseline) to 6.2, and then
increased up to below the baseline value with reperfusion. Ventricular tachycardia was induced
by this reperfusion, but eventually, the pH returned to baseline [Fig. 5(d)]. P-doped Si nanomembranes that exhibit piezoresistive behavior
constituted the strain sensors. Three strain sensors arranged in a rosette layout had
different purposes. Two sensors oriented in the ⟨110⟩ crystal direction characterized the
rhythms of the heart, while one sensor oriented in the ⟨100⟩ direction was used for
temperature calibration. In an experiment, they induced ventricular fibrillation with a drug
(Pinacidil) and confirmed a random pattern from the strain gauges and abnormal waveforms in
the ECG, which was expected for ventricular fibrillation [Fig. 5(e)]. The μ-ILEDs, based on AlInGaP, were used as light sources for optical mapping
of fluorescent and voltage-sensitive dyes. Using the LEDs, cardiac action potentials were
successfully recorded and a comparison with using an external light source verified this
finding. Although passive electrodes were used in this study, active electrodes could provide
clearer ECG signals.
FIG. 5.
Multiparametric mapping with (a) ECG electrodes, μ-inorganic LEDs, pH sensors,
temperature sensors, and strain sensors on the heart; (b) ECG and optical mapping comparison; (c) temperature mapping, inset shows the heat map during
ablation; (d) pH detection during
induced ischemia by stopping and resuming perfusion; (e) strain sensing with normal sinus rhythm, pacing, and
ventricular fibrillation (VF); (f)
array of capacitive sensors on ex vivo rabbit heart; (g) soft robotic sleeve with twisting actuators on
porcine heart; (h) ECG control scheme
for circumferential actuators, red dot indicates R peak trigger for actuation, gray
actuators are not activated, and blue indicates actuation; and (i) aortic flow rate as the trigger for both radial and
circumferential actuators, red dot indicates the start of increasing aortic flow.
Multiparametric mapping with (a) ECG electrodes, μ-inorganic LEDs, pH sensors,
temperature sensors, and strain sensors on the heart; (b) ECG and optical mapping comparison; (c) temperature mapping, inset shows the heat map during
ablation; (d) pH detection during
induced ischemia by stopping and resuming perfusion; (e) strain sensing with normal sinus rhythm, pacing, and
ventricular fibrillation (VF); (f)
array of capacitive sensors on ex vivo rabbit heart; (g) soft robotic sleeve with twisting actuators on
porcine heart; (h) ECG control scheme
for circumferential actuators, red dot indicates R peak trigger for actuation, gray
actuators are not activated, and blue indicates actuation; and (i) aortic flow rate as the trigger for both radial and
circumferential actuators, red dot indicates the start of increasing aortic flow.Fang et al. advanced cardiac electrophysiological mapping technology through
the development of ultrathin capacitive, flexible, and multiplexed electronics that were
durable in bio-fluids and conformally attached to the surface of the in vitro
perfused rabbit heart [Fig. 5(f)]. The advancements in this technology included leakage-free
encapsulation that provided long-term stability in physiological conditions and
high-resolution mapping of a dynamic surface using thin active electronics. The device had 396
multiplexed capacitive sensors that were 2500 µm2 in size, where
each sensor had two Si nanomembrane (Si NM) transistors, SiO2 that acted as both a
dielectric and a barrier from the physiological environment, and PI as the encapsulation. The
combination of the tissue interface with the SiO2 and Au contact pad of the
transistor gate comprised the capacitor; it is important to note that this coupling of the
semiconductor channel distinguished this device from conventional passive sensors. The sampling rate of this system
was 1136 Hz per sensing node and could be improved by adjusting the multiplexing rate. Noise
reached a maximum of ∼55 µV and the signal-to-noise ratio (SNR) of the device
was found to be 42 dB, which is excellent for heart monitoring. Mechanical analysis showed that the strain induced in the
Si and SiO2 is much lower than their fracture limits. The encapsulation of this
device was incredibly robust; the leakage levels were found to be 3 orders less than the
safety limit standard for implanted medical devices. Polyvinyl chloride (PVC) improved the
coupling of the device to the rabbit heart ventricles. The characteristic features of the ECG
were discernable, and mapping of the voltage showed the phases of the cardiac action
potential. The ECG was verified using optical methods, as the array was partially transparent.
To demonstrate the functionality of the sensing array even in disease states, they showed that
their phase maps could be used to identify ventricular fibrillation. The identification of the
singularity in the phase map is a method used for ablation procedures. This technology offers a possible solution to map
the Q-T interval (or repolarization of the heart) in QT syndrome studies that currently lack
the appropriate tools. Wireless transmission of data from these types of sensors could allow
for usage outside of hospitals.Zhang et al. realized a flexible and stretchable magnetic strain sensor for
wireless monitoring of cardiac strain on the pericardium of the heart. Laser-micromachined polymeric magnets were encapsulated by
PDMS, which was further embedded into the softer Ecoflex. After fabrication, magnetization was
performed using an induced magnetic field in a Halbach array, which strengthens the magnetic
field on one side of the magnet while zeroing the field on the side. In their mechanical
characterization, they found that although the bonding between the PDMS and magnetic stripe
was not ideal, the PDMS still fixed the magnet in place. The Ecoflex served as a stretchable
interconnect between the PDMS islands. The sensor was calibrated to successfully map
physiologically relevant strains of the cardiac cycle, specifically 40%-60% with a range of
60-100 beats per minute (bpm). They mimicked cardiac cycle conditions and captured relevant
signals using a smartphone application. Although this device can be used for passive
monitoring, active devices, one of which is described here, could assist heart function.Soft robotics is a burgeoning field in which mechanically soft materials are replacing the
conventionally rigid materials used in the construction of robots. Walsh and co-workers
created a soft robotic sleeve that can replace the conventionally bulky and rigid VADs used to
temporarily sustain failing hearts [Fig. 5(g)]. The tethered sleeve closely emulated the
beating of the heart by replicating the helical and circumferential arrangement of cardiac
muscle. These
muscles both twist and compress during contraction. The fabricated sleeves employed a
multilayered design that simulated the two outer muscle layers of the myocardium. The soft
pneumatic artificial muscles (PAMs) were implemented in two designs, both of which conformally
rested on the surface of the heart. Two designs were made. Design 1 was fabricated with two
layers of silicon (Ecoflex) to make the individual actuators and the layers were combined; the
final thickness was 16 mm. Design 2 was fabricated by selectively bonding two layers of
silicone sheets in a predefined pattern and then combining them with layers of thermoplastic
urethane (TPU) actuators, resulting in an overall thickness of 550 µm.
Computed tomography imaging confirmed the device’s conformability on the heart. A custom setup
with ECG electrodes was developed to both record electrophysiological data along with
confirming that the sleeve contracts in conjunction with the ex vivo heart. With various
control schemes for the circumferential and twisting actuators, the sleeve synchronized with
native heart motion; triggering of the actuation occurred with the ECG or hemodynamic
parameters of the cardiac cycle [Figs. 5(h) and 5(i)]. Physiological volume displacement was a feature of
this device that demonstrated relevance to the native cardiac environment. In a porcine model
with acute heart failure, the robotic sleeve recovered the cardiac output from ∼45% of
baseline up to 97% of baseline when activated. Further development of this device will require
making it untethered, more miniaturized, and portable. In addition, this type of sleeve could
deliver drugs to treat any patient’s heart condition and must also be further investigated to
verify long-term stability and reduce the immune response.Powering implantable devices has also come with various complications. Traditionally, surgery
has been a requirement to replace any batteries used in implanted devices, bringing
unnecessary risk and cost to the patient and caregiver. Although technology for batteries is
constantly advancing, their relatively short operational lifetimes inhibit their usage as
power supplies for devices that usually perform long-term operation, such as cardiac
pacemakers. With a spontaneously functioning mechanical pump in the body, harvesting
mechanical energy with implantable devices has transformed from an opportunity into a reality.
Many strategies have been implemented to attempt to harness power from various processes in
the human body. Yet, many of these devices fail to be mechanically compliant
with the surface of the heart. The studies described below demonstrate functional energy
harvesting and storage devices that conformally interface with the heart, through all its
contractions and relaxations.Dagdeviren et al. demonstrated the in vivo operation of a
piezoelectric energy-harvesting device that was capable of concurrently generating and storing
power from the contractions of the heart. The working mechanism of a piezoelectric material is as follows:
external forces on the material induce electric fields within it. Where positive strain (stretch) occurs, a positive
electrical potential can be detected, and where a negative strain (compression) occurs, a
negative electrical potential can be detected. The main sensing element was a capacitor-like
structure that had a layer of lead zirconate titanate (PZT) sandwiched between Ti/Pt and Cr/Au
top and bottom electrodes, respectively [Fig. 6(a)]. The
flexible mechanical energy harvester (MEH) consisted of 12 groups of this sensing element,
each group containing 10 connected in parallel; this connectivity increased the output
voltage. PI encapsulation isolated the device from leakage currents that can occur due to
contact with bodily fluid and the device was shown to be stable in phosphate-buffered saline
(PBS). The MEH was shown to be biocompatible in a cell viability test. When it experienced
compression, the PZT and electrodes bent out of plane with the buckled substrate, and in an
experimental simulation with different loads, higher frequency of bending led to more frequent
current and voltage output. Even after 20 × 106 bend and release cycles, the device
did not show degradation in function. A chip-scale rechargeable battery and Schottky bridge
rectifier stored the energy generated by the MEH. The device was sutured to 3 points of a
bovine heart, specifically, the RV, the LV, and a free wall. From these points, the voltage
output at the RV (∼4 V) was higher than that at the LV (∼3 V). This was attributed to the
enhanced wall motion during contraction of the RV which allowed for greater bending of the MEH
in comparison to the twisting motion of the LV during contraction. Exhale and inhale cycles
were also recorded [Figs. 6(b) and 6(c)]. Numerous factors, such as the orientation of the MEH on the heart
(angled placement), size of the heart, force of contraction, and heart rate were all
considered when attempting to maximize output voltage. To ensure real-world applicability, the
chest was closed, and the device was found to continue normal operation. Finally, by stacking
multiple MEHs together and spin-casting silicone layers between them, they were able to
generate a power density of up to 1.2 µW/cm2, enough to power a
cardiac pacemaker. In addition to taking advantage of the piezoelectric effect, the
triboelectric effect has been shown to be effective for energy harvesting.
FIG. 6.
Piezoelectric energy harvester (a) schematic structure and device; (b) mechanical energy harvester (MEH) on a bovine heart
during exhaling and inhaling; (c)
data recorded showing exhale/inhale cycles; (d) schematic structure of a triboelectric nanogenerator
(TENG); (e) custom wireless
transmission system for cardiac monitoring; (f) wirelessly transmitted signals at different heart rates (top)
and wirelessly transmitted signals at different charging times.
Piezoelectric energy harvester (a) schematic structure and device; (b) mechanical energy harvester (MEH) on a bovine heart
during exhaling and inhaling; (c)
data recorded showing exhale/inhale cycles; (d) schematic structure of a triboelectric nanogenerator
(TENG); (e) custom wireless
transmission system for cardiac monitoring; (f) wirelessly transmitted signals at different heart rates (top)
and wirelessly transmitted signals at different charging times.Wang and co-workers implanted a triboelectric nanogenerator (TENG) for harvesting energy from
the heartbeat of an in vivo porcine model and subsequently used it for
cardiac monitoring. The working
principle of TENG devices is as follows: by contacting two materials together, charge transfer
occurs between them in a contact electrification process, leading to a deficit of electrons in
one material (donor), while the other gains electrons (acceptor). Depending on the electrode setup used, alternating current
will flow between the electrodes (to which each of the donor and acceptor is attached) or it
will flow to ground (if a single electrode setup is used). This is also known as vertical
contact separation mode. Their TENG device had a triboelectric layer of nanostructured
polytetrafluoroethylene (n-PTFE), an ultrathin layer of Au deposited on a flexible substrate
(Kapton film) which served as one electrode, and Al foil, which was the other electrode [Fig.
6(d)]. A strip of Ti was implemented to act as the
adhesion support for the contact/separation of the TENG when the device was utilized
in vivo. PDMS and parylene C encapsulated the entirety of the structure.
Mouse fibroblasts were successfully cultured on the encapsulation material, proving that the
device was biocompatible. In vitro, the TENG outputted an open circuit
voltage of 45 V and a short circuit current of 7.5 µA. After attaching a load
resistance, the power density was found to be 107 mW/cm2. A capacitor was connected
to the TENG, along with a rectifier, and it could be charged from 1.3 to 4.2 V in 150 s. In
addition to measuring the output in vivo (with a porcine heart), the arterial
pressure and ECG were recorded. The device was placed on the interior wall of the LV and an
output of 14 V and 5 µA was recorded. They were able to time synchronize the
heart rate with this output; a capacitor was charged from 0.2 to 2.8 V with a heart rate of 80
bpm in 200 s. Like in the aforementioned paper, factors such as the location of placement,
respiratory movement (diaphragm causing deformation of the pericardium), and cardiac
contraction all impacted the output capability of the TENG. They created a self-powered
wireless transmission system (SWTS) to monitor motion frequency of the heart [Fig. 6(e)]. An implantable wireless transmitter sent
electromagnetic waves to an external receiving coil, which charged a capacitor. By analyzing
the wirelessly transmitted power, they interpreted the frequency of beating and thus
quantified the heart rate [Fig. 6(f)]. Again, the ECG
verified the consistency of this observation. At 60 bpm, 3 s of charging was enough to send
wireless data. Their long-term service evaluation showed that the device successfully operates
for 3 days without hindering cardiac motion or causing any inflammatory reaction.Soft cardiac bioelectronics have the potential to be fully integrated into the lifestyles of
patients, as they will eventually allow the transmission of encrypted cardiac data from the
patients’ homes to physicians. They will be used to identify the causes of pathophysiological
conditions associated with the heart. Integrating these conformal devices with ubiquitous
smartphones will boost their usability. This may lead to larger screening capabilities of
communities and provide greater healthcare accessibility to those living in geographically
isolated or economically limited areas. The omnipresence of machine learning ensures that these sensors
could intelligently and continuously classify various cardiac events and immediately notify
the user and caregiver of any critical conditions.Soft bioelectronics for cardiac sensing and treatment is a growing area that needs future
development. Some of the existing challenges involve the strict requirements of materials that
(1) are resistive to corrosion, degradation, and electrical current leakage that may even
induce disease states at the interfaced tissue and (2) are biocompatible to prevent immune
responses when interfaced with the tissue. In addition, multifunctionality integrated in one device with
the same mechanical properties is challenging yet critical for a more comprehensive
understanding of the heart condition, since cardiac monitoring is not only limited to heart
rate and ECG recordings.Future directions for soft and wearable heart monitors will include multi-modal recording
capabilities, integration of wireless and secure data transmission to enable remote and
ambulatory monitoring, and incorporation of state-of-the-art machine learning technologies
such as neural networks for providing enhanced feedback to users and their caregivers. The
advances in engineering and science of materials and micromanufacturing techniques render the
future scalability of these devices to make them more applicable to full human heart mapping
and reducing the inflammatory response of the body to these devices. Replacing sensing
elements with capacitive and soft sensors can circumvent the issues arising from metal
electrodes. Integrated active electronics such as transistors may be utilized for higher
density and greater resolution when recording electrophysiological data. Finally, power
generators/supplies are indispensable components to be developed to power chronic
implants.
Authors: Omer T Inan; Pierre-Francois Migeotte; Kwang-Suk Park; Mozziyar Etemadi; Kouhyar Tavakolian; Ramon Casanella; John Zanetti; Jens Tank; Irina Funtova; G Kim Prisk; Marco Di Rienzo Journal: IEEE J Biomed Health Inform Date: 2014-10-07 Impact factor: 5.772