Literature DB >> 10321660

Effects of pressure ramp slope values on the work of breathing during pressure support ventilation in restrictive patients.

G Bonmarchand1, V Chevron, J F Ménard, C Girault, F Moritz-Berthelot, P Pasquis, J Leroy.   

Abstract

OBJECTIVE: To investigate, in restrictive patients, the influence of pressure ramp slope values on the efficacy of pressure support ventilation.
DESIGN: Prospective study.
SETTING: A university hospital medical intensive care unit. PATIENTS: Twelve intubated restrictive patients.
INTERVENTIONS: Patients were randomly assigned to four sequences in which the values of the slope of the pressure ramp increase were modulated so that the plateau pressure was reached within a predetermined time: 0.1, 0.50, 1, or 1.50 secs. The more rapidly the pressure plateau was achieved, the higher was the initial flow rate. For convenience, these four different ventilatory settings were termed T 0.1, T 0.5, T 1, and T 1.5.
MEASUREMENTS AND MAIN RESULTS: We measured the following parameters 10 mins after application of each pressure ramp slope: inspiratory work of breathing, breathing pattern, and intrinsic PEEP (PEEPi). Work of breathing was evaluated using Campbell's diagram, and expressed as a percentage of the values observed under spontaneous ventilation. A marked interindividual variation of the values for work of breathing was observed under spontaneous ventilation; the mean value for work of breathing was 1.97 +/- 0.82 joule/L, with a range of 1.22 to 4.10 joule/L. Comparison between the means for each sequence and each variable measured was performed by two-way analysis of variance with internal comparisons between sequences by Duncan's test. Between the first (T 0.1) and the last (T 1.5) sequence, the reduction of values of the pressure ramp slope induced a progressive increase in the values for work of breathing, regardless of the mode of expression (in joule, joule/L, or joule/min). The values for work of breathing (joule/ L), expressed as a percentage of the values observed under spontaneous ventilation, increased from 44.2 +/- 14.4% to 78.3 +/- 17.8% (p < .001). In contrast, the reduction of the pressure ramp slope values and initial flow rate did not induce any significant change in tidal volume, respiratory frequency, and PEEPi.
CONCLUSION: Among the four tested slope values, the steepest was that which induced the lowest possible work of breathing in restrictive patients ventilated by pressure support ventilation. In this type of patient, we therefore suggest that the programmed pressure value should be reached by using a steep pressure ramp slope.

Entities:  

Mesh:

Year:  1999        PMID: 10321660     DOI: 10.1097/00003246-199904000-00023

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  12 in total

1.  Bench studies evaluating devices for non-invasive ventilation: critical analysis and future perspectives.

Authors:  Carlo Olivieri; Roberta Costa; Giorgio Conti; Paolo Navalesi
Journal:  Intensive Care Med       Date:  2011-11-29       Impact factor: 17.440

Review 2.  Hypothesis: Fever control, a niche for alpha-2 agonists in the setting of septic shock and severe acute respiratory distress syndrome?

Authors:  F Petitjeans; S Leroy; C Pichot; A Geloen; M Ghignone; L Quintin
Journal:  Temperature (Austin)       Date:  2018-05-22

3.  A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators.

Authors:  Arnaud W Thille; Aissam Lyazidi; Jean-Christophe M Richard; Fabrice Galia; Laurent Brochard
Journal:  Intensive Care Med       Date:  2009-04-08       Impact factor: 17.440

4.  Is proportional-assist ventilation with load-adjustable gain factors a user-friendly mode?

Authors:  Nectaria Xirouchaki; Eumorfia Kondili; Maria Klimathianaki; Dimitris Georgopoulos
Journal:  Intensive Care Med       Date:  2009-06-16       Impact factor: 17.440

Review 5.  Building on the Shoulders of Giants: Is the use of Early Spontaneous Ventilation in the Setting of Severe Diffuse Acute Respiratory Distress Syndrome Actually Heretical?

Authors:  Fabrice Petitjeans; Cyrille Pichot; Marco Ghignone; Luc Quintin
Journal:  Turk J Anaesthesiol Reanim       Date:  2018-09-01

6.  Weaning from mechanical ventilation with pressure support in patients failing a T-tube trial of spontaneous breathing.

Authors:  Eric Ezingeard; Eric Diconne; Stéphane Guyomarc'h; Christophe Venet; Dominique Page; Pierre Gery; Régine Vermesch; Monique Bertrand; Juliette Pingat; Bernard Tardy; Jean-Claude Bertrand; Fabrice Zeni
Journal:  Intensive Care Med       Date:  2005-11-10       Impact factor: 17.440

7.  A diaphragmatic electrical activity-based optimization strategy during pressure support ventilation improves synchronization but does not impact work of breathing.

Authors:  Francois Beloncle; Lise Piquilloud; Nuttapol Rittayamai; Christer Sinderby; Hadrien Rozé; Laurent Brochard
Journal:  Crit Care       Date:  2017-01-31       Impact factor: 9.097

8.  New setting of neurally adjusted ventilatory assist for noninvasive ventilation by facial mask: a physiologic study.

Authors:  Federico Longhini; Chun Pan; Jianfeng Xie; Gianmaria Cammarota; Andrea Bruni; Eugenio Garofalo; Yi Yang; Paolo Navalesi; Haibo Qiu
Journal:  Crit Care       Date:  2017-07-07       Impact factor: 9.097

Review 9.  Setting up home noninvasive ventilation.

Authors:  Jean-Michel Arnal; Charles-Philippe Thevenin; Benoit Couzinou; Joelle Texereau; Aude Garnero
Journal:  Chron Respir Dis       Date:  2019 Jan-Dec       Impact factor: 2.444

Review 10.  Patient-Ventilator Dyssynchrony.

Authors:  Elvira-Markela Antonogiannaki; Dimitris Georgopoulos; Evangelia Akoumianaki
Journal:  Korean J Crit Care Med       Date:  2017-11-30
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.