Evaluation of a flexible fiberoptic catheter in confirming endotracheal tube placement in the intensive care unit.

M Suarez, A Chediak, P Ershowsky, B Krieger.   

Abstract

UNLABELLED: We evaluated the use of an inexpensive 2.2-mm fiberoptic catheter (FOC) to assess endotracheal tube (ETT) placement in 22 critically ill adult patients.
METHOD: The distance between the carina and the distal tip of the ETT was measured under direct visualization with the FOC and compared to the corresponding measurement obtained from concomitant chest radiographs.
RESULTS: There was a significant linear correlation between measurements made by the two methods (r = 0.97, P less than 0.0001), and all FOC measurements were within 1 cm of the corresponding radiograph measurements. No complications or significant changes occurred in peak airway pressure, tidal volume, respiratory rate, or heart rate. The procedure usually was performed in less than 1 minute. Cost analysis revealed the possibility of major savings when the FOC is used instead of portable chest radiography.
CONCLUSIONS: We conclude that documentation of endotracheal tube placement by means of a fiberoptic catheter is rapid, accurate, safe, and cost-effective.

Entities:  

Mesh:

Year:  1987        PMID: 10317847

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  3 in total

1.  Automatic Detection of Endotracheal Intubation During the Anesthesia Procedure.

Authors:  Ali Jalali; Mohamed Rehman; Arul Lingappan; C Nataraj
Journal:  J Dyn Syst Meas Control       Date:  2016-08-09       Impact factor: 1.372

2.  Novel device (AirWave) to assess endotracheal tube migration: a pilot study.

Authors:  Gustavo Cumbo Nacheli; Manish Sharma; Xiaofeng Wang; Amit Gupta; Jorge A Guzman; Adriano R Tonelli
Journal:  J Crit Care       Date:  2013-02-05       Impact factor: 3.425

Review 3.  Bronchoscopy in the intensive care unit.

Authors:  P Jolliet; J C Chevrolet
Journal:  Intensive Care Med       Date:  1992       Impact factor: 17.440

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.