In vitro effectiveness study of three hydrodynamic thrombectomy devices.

S Müller-Hülsbeck1, C Bangard, H Schwarzenberg, C C Glüer, M Heller.   

Abstract

PURPOSE: To determine the in vitro efficacy of three hydrodynamic thrombectomy devices.
MATERIALS AND METHODS: Thrombectomy of clots was performed with three thrombectomy devices (Angiojet [AJ], Possis Medical, Minneapolis, Minn; Hydrolyser [HL] Cordis Europe, Roden, the Netherlands; and the Shredding Embolectomy Thrombectomy [SET] catheter, HP-Medica, Augsburg, Germany) in a flow model.
RESULTS: Mean thrombectomy time ranged from 10.22 seconds (HL) to 37.73 seconds (AJ with guide wire). For the AJ and HL, the use of guide wires prolonged thrombectomy time (P < .01). The AJ with and without a guide wire and the HL with a 0.018-inch guide wire worked isovolumetrically, whereas the mean ratio of applied saline and aspirated fluid for the other devices was different from 1, ranging from 0.54 to 0.72. Mean embolus weight with the AJ alone (56.44 mg) was significantly higher than that with the SET catheter alone (3.15 mg) and with a guide wire (1.31 mg, P < .01 for both) and the HL alone (3.9 mg, P < .05), as was the embolus weight with the HL with a 0.018-inch guide wire (66.5 mg) compared with the SET catheter with and without a guide wire (P < .01), AJ with a guide wire (22.33 mg, P < .05), the HL alone (P < .01), and the HL with a 0.025-inch guide wire (24.86 mg, P < .05).
CONCLUSION: The devices showed performance differences. The SET catheter alone and with a guide wire and the HL may bear an increased risk of procedure-related anemia. In clinical applications, hydrodynamic thrombectomy might substantially reduce the need for thrombolytic therapy.

Entities:  

Mesh:

Year:  1999        PMID: 10228525     DOI: 10.1148/radiology.211.2.r99ma06433

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  5 in total

1.  A reconstituted in vitro clot model for evaluating laser thrombolysis.

Authors:  Abram D Janis; Lisa A Buckley; Abby N Nyara; Scott A Prahl; Kenton Gregory
Journal:  J Thromb Thrombolysis       Date:  2002-06       Impact factor: 2.300

2.  In vitro and in vivo high-intensity focused ultrasound thrombolysis.

Authors:  Cameron Wright; Kullervo Hynynen; David Goertz
Journal:  Invest Radiol       Date:  2012-04       Impact factor: 6.016

Review 3.  New insights on thrombolytic treatment of acute ischemic stroke.

Authors:  J F Meschia; T G Brott
Journal:  Curr Neurol Neurosci Rep       Date:  2001-01       Impact factor: 5.081

4.  Solitaire FR device for treatment of dural sinus thrombosis.

Authors:  Bryan Anthony Pukenas; Monisha Kumar; Michael Stiefel; Michelle Smith; Robert Hurst
Journal:  BMJ Case Rep       Date:  2012-12-19

5.  Experimental Evaluation of the Effectiveness of Aspiration-Based Techniques to Treat Different Types of Acute Thromboembolic Occlusions in the Femoropopliteal Vascular System Using an In Vitro Flow Model.

Authors:  Schekeb Aludin; Philipp Jost Schäfer; Christoph Borzikowsky; Olav Jansen; Julian Pfarr; Rouven Berndt; René Rusch; Jens Trentmann
Journal:  Cardiovasc Intervent Radiol       Date:  2021-12-17       Impact factor: 2.740

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.