Literature DB >> 10202702

Worst-Fechner biconcave minus power phakic iris-claw lens.

P U Fechner1, I Haubitz, W Wichmann, K Wulff.   

Abstract

OBJECTIVE: Evaluation of long-term results of correction of high myopia by implantation of a biconcave Worst-Fechner minus power iris-claw lens into phakic eyes.
METHODS: One hundred and twenty-seven eyes of 70 patients were implanted with the phakic IOL between November 1986 and November 1991. The mean preoperative spherical equivalent refraction was -14.29 +/- 5.77 D (range -5.00 to -31.75 D). The total population decreased from 127 to 68 eyes (53.5%) at the 8-year examination. By this time, 26 of the 70 patients (37%) had been lost to follow-up for unknown reasons. Corneal endothelial density was estimated by comparing the image gained by non-contact specular microscopy with a graduated reticule (Karickhoff method).
RESULTS: At the 6-month examination, mean deviation of achieved from calculated refractive correction was +0.52 +/- 1.46 D (range -2.25 to +6.50 D). Seventy-seven eyes (62.1%) deviated 1.00 D or less, and 15 eyes (12.1%) deviated by more than 2.00 D from the calculated correction. A refractive outcome of +/- 1.00 D was attempted in 68 eyes (54% of the 124 eyes) and was achieved in 51 eyes (75%). Mean spectacle-corrected visual acuity preoperatively was 0.54 +/- 0.27 (range 1.20 to 0.05). At the 6-month examination it was 0.73 +/- 0.3, and by the 8-year examination had decreased to 0.65 +/- 0.26. Statistical analysis of corneal endothelial cell density revealed a significant correlation of endothelial cell loss with age (> or = 45 years), anterior chamber depth (< or = 3.4 mm), and IOL power (> or = -11.00 D). Seventeen eyes (13.4%) had a decrease in endothelial cell density and projection to 8-year follow-up resulted in a decrease in 27% of eyes. Four eyes needed a penetrating keratoplasty.
CONCLUSION: Implantation of a biconcave Worst-Fechner lens into a phakic eye effectively corrected high myopia with a stable refractive outcome, but 13.4% of implanted eyes suffered significant progressive reduction of endothelial cell density. This style lens is no longer in clinical use.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10202702     DOI: 10.3928/1081-597X-19990301-04

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  8 in total

Review 1.  [Intraocular lenses for the correction of refraction errors. Part II. Phakic posterior chamber lenses and refractive lens exchange with posterior chamber lens implantation].

Authors:  T Kohnen; T Kasper; E Terzi
Journal:  Ophthalmologe       Date:  2005-11       Impact factor: 1.059

Review 2.  [Intraocular lenses for the correction of refraction errors. Part 1: phakic anterior chamber lenses].

Authors:  T Kohnen; M Baumeister; M Cichocki
Journal:  Ophthalmologe       Date:  2005-10       Impact factor: 1.059

3.  Correction of high myopia with different phakic anterior chamber intraocular lenses: ICARE angle-supported lens and Verisyse iris-claw lens.

Authors:  Stanislawa Gierek-Ciaciura; Ariadna Gierek-Lapinska; Krzysztof Ochalik; Ewa Mrukwa-Kominek
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2006-07-01       Impact factor: 3.117

Review 4.  [Scheimpflug photography for the examination of phakic intraocular lenses].

Authors:  M Baumeister
Journal:  Ophthalmologe       Date:  2014-10       Impact factor: 1.059

5.  Long-term endothelial cell loss with the iris-claw intraocular phakic lenses (Artisan®).

Authors:  Virgilio Galvis; John F Villamil; María Fernanda Acuña; Paul A Camacho; Jesús Merayo-Lloves; Alejandro Tello; Sandra Lizeth Zambrano; Juan José Rey; Juan Vicente Espinoza; Angélica María Prada
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-10-28       Impact factor: 3.117

6.  [Influence of incision parameters on astigmatism during implantation of phakic-6-mm-iris-claw intraocular lenses].

Authors:  W Sekundo; M Schneider; A Tietjen
Journal:  Ophthalmologe       Date:  2004-03       Impact factor: 1.059

7.  Long-Term Comparison of Posterior Chamber Phakic Intraocular Lens With and Without a Central Hole (Hole ICL and Conventional ICL) Implantation for Moderate to High Myopia and Myopic Astigmatism: Consort-Compliant Article.

Authors:  Kimiya Shimizu; Kazutaka Kamiya; Akihito Igarashi; Hidenaga Kobashi
Journal:  Medicine (Baltimore)       Date:  2016-04       Impact factor: 1.889

8.  Myopic Correction with Iris-Fixated Phakic Intraocular Lenses: Twelve-Year Results.

Authors:  Iveta Nemcova; Jiri Pasta; Katerina Hladikova; Martin Komarc; Darina Pospisilova; Pavel Nemec; Jan Tesar; Vladimir Kratky; Martin Sin
Journal:  J Ophthalmol       Date:  2021-10-07       Impact factor: 1.909