The effect of a thin coating of polymethylmethacrylate on the torsional fatigue strength of the cement-metal interface.

J P Davies1, G Singer, W H Harris.   

Abstract

Recent studies have established that a mechanism of initiation of failure of fixation of cemented femoral components is debonding of the cement-metal interface. Other studies have shown that the torsional forces induced by stair climbing and rising from a chair are very high. Thus, the interface between the femoral prosthesis and the bone cement in total hip arthroplasty (THA) is required to transmit high torsional loads from the metal to the cement in a cyclic shear mode many times per year. These torsional loads likely contribute to the debonding. This study evaluated the efficacy of a thin layer of polymethylmethacrylate (PMMA) precoating in increasing the torsional fatigue strength of the cement-metal interface. Fatigue studies were performed on 15 specimens. Each specimen was tested with and without PMMA precoating. The PMMA precoat significantly and substantially increased the torsional fatigue strength of the cement-metal interface. Thus, PMMA precoating is likely to be a clinical advantage in maintaining the long-term integrity of the cement-prosthesis interface.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 10147704     DOI: 10.1002/jab.770030108

Source DB:  PubMed          Journal:  J Appl Biomater        ISSN: 1045-4861


  5 in total

1.  Multi-technique characterization of retrieved bone cement from revised total hip arthroplasties.

Authors:  T Eliades; J S Papadopulos; G Eliades; N Silikas; D C Watts
Journal:  J Mater Sci Mater Med       Date:  2003-11       Impact factor: 3.896

2.  Factors affecting the static shear strength of the prosthetic stem-bone cement interface.

Authors:  Jian-Sheng Wang; Mark Taylor; Gunnar Flivik; Lars Lidgren
Journal:  J Mater Sci Mater Med       Date:  2003-01       Impact factor: 3.896

Review 3.  Why have we left Charnley low friction arthroplasty?

Authors:  D D Goetz; W H Harris
Journal:  Iowa Orthop J       Date:  1993

4.  Comparative retrieval analysis of a novel anatomic tibial tray backside: alterations in tibial component design and surface coating can increase cement adhesions and surface roughness.

Authors:  Dominic T Mathis; Joshua Schmidli; Felix Amsler; Johann Henckel; Harry Hothi; Alister Hart; Michael T Hirschmann
Journal:  BMC Musculoskelet Disord       Date:  2022-05-19       Impact factor: 2.562

5.  Random damage and characteristics of debris particles are two important and yet ignored factors in the mechanical integrity of the stem-cement interface of a total hip replacement: influence of the surface finish of the metal stem.

Authors:  Gang Qi; Steven F Wayne; Kenneth A Mann; Bin Zhang; Gladius Lewis
Journal:  J Mater Sci Mater Med       Date:  2009-11-28       Impact factor: 3.896