Residual bacterial contamination on reusable pulse oximetry sensors.

M C Wilkins1.   

Abstract

BACKGROUND: An early study demonstrated that standard cleaning techniques did not adequately disinfect reusable pulse oximetry sensors that had been intentionally contaminated with high titers of pathogenic microorganisms. This current study evaluated patient-ready sensors being used in hospitals, by testing them for bacterial contamination.
METHODS: Reusable pulse oximetry sensors from 15 hospitals throughout the United States were tested. Each sensor was deemed by the hospital to be ready for patient use, and it had been prepared for use according to hospital procedures. Patient-contact areas of each sensor were swabbed, and the swabs were analyzed for bacterial contamination using standard testing procedures. This study had two stages, and different sensors were tested in each stage. In the "bacterial-growth stage," swabs were evaluated for bacterial contamination but organisms were not identified; in the "identification stage," bacterial species were identified.
RESULTS: Forty-four sensors were evaluated, 16 in the bacterial-growth stage and 28 in the identification stage. Bacteria were cultured from 29 of the 44 sensors (66%), including 20 that had been cleaned with alcohol or an antibacterial/antiviral agent. Among the isolated organisms were Staphylococcus aureus, Staphylococcus haemolyticus, Enterococcus faecalis, and Klebsiella oxytoca. Bacterial contamination was found on sensors from 12 of the 15 participating hospitals.
CONCLUSIONS: These data demonstrate the need to define effective cleaning methods for reusable sensors, and we are currently conducting such studies. The data also suggest that disposable patient-dedicated sensors may be the most appropriate choice when infection control is of particular concern.

Entities:  

Mesh:

Year:  1993        PMID: 10145923

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  3 in total

1.  Port central venous catheters-associated bloodstream infection during outpatient-based chemotherapy.

Authors:  Davide Mauri; Sofia Roumbkou; Stella Michalopoulou; Lamprini Tsali; Anastasia Spiliopoulou; Charalampos Panou; Antonis Valachis; Angelos Panagopoulos; Nikolaos P Polyzos
Journal:  Med Oncol       Date:  2009-12-05       Impact factor: 3.064

2.  Non-contact measurement of oxygen saturation with an RGB camera.

Authors:  Alessandro R Guazzi; Mauricio Villarroel; João Jorge; Jonathan Daly; Matthew C Frise; Peter A Robbins; Lionel Tarassenko
Journal:  Biomed Opt Express       Date:  2015-08-11       Impact factor: 3.732

3.  Microbial contamination of manually reprocessed, ready to use ECG lead wire in intensive care units.

Authors:  Trisasi Lestari; Sylvia Ryll; Axel Kramer
Journal:  GMS Hyg Infect Control       Date:  2013-04-29
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.