An evaluation of the resistance to flow through the patient valves of twelve adult manual resuscitators.

D Hess1, M Simmons.   

Abstract

UNLABELLED: What is the inspiratory and expiratory resistance to flow through the patient valves of adult manual resuscitators? MATERIALS &
METHODS: We evaluated the resistance to flow through the patient valves of 12 adult resuscitators (Ambu, Code Blue, DMR, Hope 4, Hospitak, Hudson, Intertech, Laerdal, Mercury, Respironics, SPUR, Vitalograph). Expiratory resistance was evaluated by directing a flow of oxygen through the valve in the direction that the patient expires. Inspiratory resistance was evaluated by directing oxygen through the valve in the direction of flow when the bag is squeezed. Flow was controlled by a Timeter 0-75 flowmeter, and measured using a calibrated Timeter RT-200. Flows of 10, 20, 30, 40, 50, 60, 70, 80, and 90 L/min were used. Resistive back pressure of the resuscitator valves was measured using a calibrated Timeter RT-200. Resistance was calculated by dividing back pressure by flow. Five measurements were made at each flow setting for each resuscitator.
RESULTS: Significant differences in back pressures and resistances existed between the resuscitators for both expiratory and inspiratory flows (p less than 0.001 in each case). Significant interaction effects also existed between resuscitator brands and flows (p less than 0.001 in each case). At an expiratory flow of 50 L/min, all resuscitators except the Hospitak and Vitalograph produced a back pressure less than 5 cm H2O (the International Standards Organization standard). At an inspiratory flow of 50 L/min, all resuscitators but the Hospitak, Mercury, and Vitalograph produced a back pressure less than 5 cm H2O.
CONCLUSIONS: Significant differences existed in the back pressures produced due to the flow resistance through the patient valves of these resuscitators, and these might be considered excessive in some cases. Because this was a bench study, further work is needed to determine the clinical importance of these findings.

Entities:  

Mesh:

Year:  1992        PMID: 10145632

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  2 in total

1.  Apnea test during brain death assessment in mechanically ventilated and ECMO patients.

Authors:  Marco Giani; Vittorio Scaravilli; Sebastiano Maria Colombo; Andrea Confalonieri; Rosambra Leo; Elena Maggioni; Leonello Avalli; Alessia Vargiolu; Giuseppe Citerio
Journal:  Intensive Care Med       Date:  2015-11-10       Impact factor: 17.440

2.  Evaluation of manual resuscitators used in ICUs in Brazil.

Authors:  Tatiana de Arruda Ortiz; Germano Forti; Márcia Souza Volpe; Marcelo do Amaral Beraldo; Marcelo Britto Passos Amato; Carlos Roberto Ribeiro Carvalho; Mauro Roberto Tucci
Journal:  J Bras Pneumol       Date:  2013 Sep-Oct       Impact factor: 2.624

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.