Distal migration of stent-grafts after endovascular repair of abdominal aortic aneurysms.

T Resch1, K Ivancev, J Brunkwall, U Nyman, M Malina, B Lindblad.   

Abstract

PURPOSE: To analyze patients after endovascular repair of abdominal aortic aneurysm (AAA) with respect to distal migration of stent-grafts and its underlying causes.
MATERIALS AND METHODS: Sixty-five patients underwent endovascular repair between January 1994 and February 1997. There were seven women and 58 men, with a mean age of 71 years (range, 51-84 years). Three patients died in the perioperative period (one of myocardial infarction and two of multiorgan failure) and two patients died within 4 months of the procedure of non-procedure-related causes. In addition, two patients were followed at another hospital. The remaining 58 patients were followed up with spiral computed tomography scans at 1, 3, and 6 months, and biannually thereafter. Angiography was performed at 1 month and 1 year after the procedure and additionally when deemed clinically necessary. Mean follow-up was 29 months (range, 1-49). Migration more than 5 mm was considered significant.
RESULTS: Twenty-six patients (45%) showed distal migration of stent-grafts during follow-up. Mean follow-up time at detection of migration was 13 months (range, 1-36 months). Thirteen cases of migration were ascribed to dilatation of the proximal aneurysmal neck during follow-up. Ten cases of migration were ascribed to causes other than neck dilatation or poor patient selection. In three cases, no obvious cause for the migration was found. The migration was complete in eight cases, leading to late conversion to open surgical repair. On two of these occasions, complete migration lead to aneurysm rupture. In addition, four patients received additional stent-grafts as proximal extensions.
CONCLUSIONS: Distal migration of stent-grafts after endovascular AAA repair occurred frequently in this series. Dilatation of the proximal aneurysmal neck is a major cause of distal migration of stent-grafts. Improved proximal fixation is needed to secure long-term durability.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10102188     DOI: 10.1016/s1051-0443(99)70027-8

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  7 in total

1.  Evaluation of the proximal aortic neck enlargement following endovascular repair of abdominal aortic aneurysm: 3-years experience.

Authors:  Vinicio Napoli; Savino G Sardella; Irene Bargellini; Pasquale Petruzzi; Roberto Cioni; Claudio Vignali; Mauro Ferrari; Carlo Bartolozzi
Journal:  Eur Radiol       Date:  2003-04-12       Impact factor: 5.315

2.  Preliminary 3D computational analysis of the relationship between aortic displacement force and direction of endograft movement.

Authors:  C Alberto Figueroa; Charles A Taylor; Victoria Yeh; Allen J Chiou; Madhu L Gorrepati; Christopher K Zarins
Journal:  J Vasc Surg       Date:  2010-06       Impact factor: 4.268

Review 3.  Endoluminal abdominal aortic aneurysm repair: the latest advances in prevention of distal endograft migration and type 1 endoleak.

Authors:  Maaz Ghouri; Zvonimir Krajcer
Journal:  Tex Heart Inst J       Date:  2010

Review 4.  Diagnostic imaging methods applied in long-term surveillance after EVAR. Will computed tomography angiography be replaced by other methods?

Authors:  Karolina Stefaniak; Michal Stanisic; Marcin Gabriel; Grzegorz Oszkinis
Journal:  Postepy Kardiol Interwencyjnej       Date:  2016-02-11       Impact factor: 1.426

5.  An in Vitro Twist Fatigue Test of Fabric Stent-Grafts Supported by Z-Stents vs. Ringed Stents.

Authors:  Jing Lin; Robert Guidoin; Jia Du; Lu Wang; Graeham Douglas; Danjie Zhu; Mark Nutley; Lygia Perron; Ze Zhang; Yvan Douville
Journal:  Materials (Basel)       Date:  2016-02-16       Impact factor: 3.623

6.  Early and Late Endograft Limb Proximal Migration with Resulting Type 1b Endoleak following an EVAR for Ruptured AAA.

Authors:  Patrick T Jasinski; Demetri Adrahtas; Spyridon Monastiriotis; Apostolos K Tassiopoulos
Journal:  Case Rep Vasc Med       Date:  2017-01-31

7.  Fast and Accurate Computation of the Displacement Force of Stent Grafts after Endovascular Aneurysm Repair.

Authors:  Ming Qing; Zhan Liu; Tinghui Zheng
Journal:  Bioengineering (Basel)       Date:  2022-09-06
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.