Analysis of titanium induced CT artifacts in the development of biomechanical finite element models.

C Zannoni1, M Viceconti, L Pierotti, A Cappello.   

Abstract

X-ray computerized tomography (CT) is capable of providing detailed information about the geometry and mineral density of skeletal structures. Such accurate data are of great interest in studying the effects of orthopaedic implants on bone adaptive behaviour in vivo. Metallic implants, however, generate artifacts, typically seen as starburst streaking. These artifacts can degrade the capabilities of CT images to provide accurate information about the geometry and mineral density of bone structures. The aim of this work was to investigate the possibility of developing finite element models (FEM) of the human femur after hip joint arthroplasty using CT images acquired directly after surgery. The capability of modern CT scanners to accurately reconstruct the cross-section geometry of titanium alloy hip joint prosthetic stems was primarily investigated. A new measuring procedure dealing with the geometry of real stems was developed and its accuracy assessed. Secondly, the artifacts generated by a prosthetic stem on the radiological density of the bone were analysed, and their effects on the assessment of FEM material properties were evaluated. Results showed that CT images provide accurate information on metal stem geometry. An average error of 0.45 mm was estimated in the reconstruction of stem cross-section geometry. Concerning bone density estimation around the implant, it was observed that the effect of metal artifacts on tissue density becomes zero at a distance of 2 mm from the implant.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 10098609     DOI: 10.1016/s1350-4533(98)00076-9

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  5 in total

1.  Proximal tibial bone density is preserved after unicompartmental knee arthroplasty.

Authors:  Bradley I Richmond; Simon V Hadlow; Tim G Lynskey; Cameron G Walker; Jacob T Munro
Journal:  Clin Orthop Relat Res       Date:  2013-01-17       Impact factor: 4.176

2.  BPrevention of Cutaneous Tissue Contracture During Removal of Craniofacial Implant Superstructures for CT and MRI Studies.

Authors:  Maureen Sullivan; Rachael Rossitto; David Casey
Journal:  J Oral Maxillofac Res       Date:  2010-07-01

3.  [Progress of change in bone mineral density after knee arthroplasty].

Authors:  Wenxing Wei; Yuangang Wu; Yi Zeng; Bin Shen
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2021-01-15

4.  Quantitative Computerized Assessment of the Degree of Acetabular Bone Deficiency: Total radial Acetabular Bone Loss (TrABL).

Authors:  Frederik Gelaude; Tim Clijmans; Hendrik Delport
Journal:  Adv Orthop       Date:  2011-10-17

5.  Quantification of dental prostheses on cone-beam CT images by the Taguchi method.

Authors:  Rong-Fu Kuo; Kwang-Ming Fang; Wong Ty; Chia Yu Hu
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.