Literature DB >> 10089956

Cochlear reimplantation.

T J Balkany1, A V Hodges, O Gómez-Marín, P A Bird, S Dolan-Ash, S Butts, F F Telischi, D Lee.   

Abstract

OBJECTIVE: A small number of multichannel cochlear implant (CI) recipients require reimplantation. This study describes the causes of failure, surgical technique, and hearing outcomes in a consecutive series of 16 patients undergoing reimplantation of multichannel devices. We hypothesize that reimplantation is safe and that hearing results are at least as good as those measured following primary implantation. STUDY
DESIGN: Retrospective analysis of consecutive clinical series.
METHODS: Chart analysis of 191 consecutive CI operations performed at the University of Miami Ear Institute between 1990 and 1997 revealed 16 patients who received a second multichannel device. All but one had a minimum follow-up of 1 year after reimplantation. Ten of these patients had initial implantation performed by us, and six were initially operated on elsewhere. Main outcomes of the initial procedure were compared with those of the reimplantation, including electrode insertion length, number of channels programmed, and audiometric results. In addition, cause of failure and relevant surgical findings at the second procedure are described.
RESULTS: There were no surgical complications after reimplantation surgery. Device failure was the most frequent cause for reimplantation. Time between initial implantation and failure ranged from 0 to 46 months (mean, 22.4 mo; median, 23 mo). Common intraoperative findings include mastoid fibrosis, bone growth at the cochleostomy, and skin flap breakdown. Following reimplantation, mean length of insertion, number of channels actively programmed, and speech recognition scores were at least as good as findings before initial implant failure.
CONCLUSION: CI reimplantation is safe and effective.

Entities:  

Mesh:

Year:  1999        PMID: 10089956     DOI: 10.1097/00005537-199903000-00002

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


  6 in total

1.  Residual hearing in cochlear implant patients.

Authors:  Walter Di Nardo; Italo Cantore; Pietro Melillo; Francesca Cianfrone; Alessandro Scorpecci; Gaetano Paludetti
Journal:  Eur Arch Otorhinolaryngol       Date:  2007-02-27       Impact factor: 2.503

2.  Revision surgeries in cochlear implant patients: a review of 45 cases.

Authors:  Lela Migirov; Riki Taitelbaum-Swead; Minka Hildesheimer; Jona Kronenberg
Journal:  Eur Arch Otorhinolaryngol       Date:  2006-09-29       Impact factor: 2.503

3.  Audiological Outcomes and Associated Factors after Pediatric Cochlear Reimplantation.

Authors:  Fabian Blanc; Catherine Blanchet; Marielle Sicard; Fanny Merklen; Frederic Venail; Michel Mondain
Journal:  J Clin Med       Date:  2022-06-01       Impact factor: 4.964

4.  Characteristics of malfunctioning channels in pediatric cochlear implants.

Authors:  Jerry W Lin; Avni Mody; Ross Tonini; Claudia Emery; Jody Haymond; Jeffrey T Vrabec; John S Oghalai
Journal:  Laryngoscope       Date:  2010-02       Impact factor: 3.325

5.  Clinical outcomes of cochlear reimplantation due to device failure.

Authors:  Chong-Sun Kim; Dong-Kyu Kim; Myung-Whan Suh; Seung Ha Oh; Sun O Chang
Journal:  Clin Exp Otorhinolaryngol       Date:  2008-03-20       Impact factor: 3.372

6.  Cochlear implant revision surgeries in children.

Authors:  Maria Stella Arantes do Amaral; Ana Cláudia Mirândola B Reis; Eduardo T Massuda; Miguel Angelo Hyppolito
Journal:  Braz J Otorhinolaryngol       Date:  2018-02-16
  6 in total