Determination of inferior vena cava diameter in the angiography suite: comparison of three common methods.

D B Brown1, M R Labuski, J F Cardella, H Singh, P N Waybill.   

Abstract

PURPOSE: Significant inferior vena cava (IVC) filter migration has been associated with deployment of some filter types in IVCs measuring more than 28 mm in diameter at inferior vena cavography. The purposes of this study were to (a) determine if significant differences exist between IVC measurements obtained using a gold standard technique and two other widely accepted methods, and (b) if differences exist, how often do these differences cause incorrect IVC sizing around a diameter of 28 mm, with its associated filter migration issues.
MATERIALS AND METHODS: One hundred thirteen consecutive inferior vena cavograms were retrospectively reviewed. The transverse diameter of the infrarenal IVC was determined by using a calibrated intravascular catheter (the gold standard), subtraction of 20% from the measured transverse IVC diameter on a cut-film radiograph, and a radiopaque ruler placed immediately posterior to the patient.
RESULTS: The concordance correlation of the 20% magnification method versus internal calibration was 0.94. Kappa analysis to determine agreement at a diameter of 28 mm yielded a Kappa coefficient of 0.490. The concordance correlation of an external ruler versus internal calibration was 0.43, with a Kappa coefficient of 0.
CONCLUSION: The poor Kappa correlations for both methods demonstrate that they are unreliable in identifying megacava. Inferior vena cavography prior to IVC filter placement should be performed with a calibrated intravascular catheter.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10082100     DOI: 10.1016/s1051-0443(99)70456-2

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  3 in total

1.  Influence of breathing movements and Valsalva maneuver on vena caval dynamics.

Authors:  Alicia Laborda; Sergio Sierre; Mauro Malvè; Ignacio De Blas; Ignatios Ioakeim; William T Kuo; Miguel Angel De Gregorio
Journal:  World J Radiol       Date:  2014-10-28

2.  Cavographic vs. cross-sectional measurement of the inferior vena cava diameter before filter placement: are we routinely oversizing?

Authors:  Yu-Dong Xiao; Zi-Shu Zhang; Cong Ma
Journal:  Eur Radiol       Date:  2018-11-09       Impact factor: 5.315

3.  Effect of calibration methods on the accuracy of angiographic measurements during transcatheter procedures in dogs.

Authors:  Lauren E Markovic; Brian A Scansen
Journal:  J Vet Intern Med       Date:  2018-03-12       Impact factor: 3.333

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.