| Literature DB >> 10080286 |
A M Howseman1, D L Thomas, G S Pell, S R Williams, R J Ordidge.
Abstract
Magnetic resonance imaging methods that are sensitive to T2* are widely used in the study of blood oxygenation changes, most notably in functional studies of the brain. In these studies the signal intensity change in T2*-weighted imaging is related to the coupling of cerebral blood flow and metabolism. Rapid measurement of T2* itself would offer a valuable method to quantify blood oxygenation changes indirectly and monitor their time course. An interleaved echoplanar imaging (EPI) sequence is presented here that allows maps of T2* to be generated in a few seconds. The sequence benefits from reduced geometric distortion and an improved point spread function compared with single-shot EPI. A comparison among a set of T2*-weighted interleaved EPI images, single-shot EPI, and conventional gradient-echo and spin-echo methods is made using a compartmentalized doped water phantom. The interleaved sequence yields accurate T2* values when compared with reference measurements made using the slower gradient-echo technique. Data acquired from the rat brain at 2.35 T prior to and during an anoxic challenge show, with high temporal resolution, the reduction in T2* associated with increased levels of deoxyhemoglobin.Entities:
Mesh:
Substances:
Year: 1999 PMID: 10080286 DOI: 10.1002/(sici)1522-2594(199902)41:2<368::aid-mrm22>3.0.co;2-1
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668