Validation of the force and frequency characteristics of the activator adjusting instrument: effectiveness as a mechanical impedance measurement tool.

T S Keller1, C J Colloca, A W Fuhr.   

Abstract

OBJECTIVE: To determine the dynamic force-time and force-frequency characteristics of the Activator Adjusting Instrument and to validate its effectiveness as a mechanical impedance measurement device; in addition, to refine or optimize the force-frequency characteristics of the Activator Adjusting Instrument to provide enhanced dynamic structural measurement reliability and accuracy.
METHODS: An idealized test structure consisting of a rectangular steel beam with a static stiffness similar to that of the human thoracolumbar spine was used for validation of a method to determine the dynamic mechanical response of the spine. The Activator Adjusting Instrument equipped with a load cell and accelerometer was used to measure forces and accelerations during mechanical excitation of the steel beam. Driving point and transfer mechanical impedance and resonant frequency of the beam were determined by use of a frequency spectrum analysis for different force settings, stylus masses, and stylus tips. Results were compared with beam theory and transfer impedance measurements obtained by use of a commercial electronic PCB impact hammer.
RESULTS: The Activator Adjusting Instrument imparted a very complex dynamic impact comprising an initial high force (116 to 140 N), short duration pulse (<0.1 ms) followed by several lower force (30 to 100 N), longer duration impulses (1 to 5 ms). The force profile was highly reproducible in terms of the peak impulse forces delivered to the beam structure (<8% variance). Spectrum analysis of the Activator Adjusting Instrument impulse indicated that the Activator Adjusting Instrument has a variable force spectrum and delivers its peak energy at a frequency of 20 Hz. Added masses and different durometer stylus tips had very little influence on the Activator Adjusting Instrument force spectrum. The resonant frequency of the beam was accurately predicted by both the Activator Adjusting Instrument and electronic PCB impact hammer, but variations in the magnitude of the driving point impedance at the resonant frequency were high (67%) compared with the transfer impedance measurements obtained with the electronic PCB impact hammer, which had a more uniform force spectrum and was more repeatable (<10% variation). The addition of a preload-control frame to the Activator Adjusting Instrument improved the characteristics of the force frequency spectrum and repeatability of the driving point impedance measurements.
CONCLUSION: These findings indicate that the Activator Adjusting Instrument combined with an integral load cell and accelerometer was able to obtain an accurate description of a steel beam with readily identifiable geometric and dynamic mechanical properties. These findings support the rationale for using the device to assess the dynamic mechanical behavior of the vertebral column. Such information would be useful for SMT and may ultimately be used to evaluate the [corrected] biomechanical effectiveness of various manipulative, surgical, and rehabilitative spinal procedures.

Entities:  

Mesh:

Year:  1999        PMID: 10073622     DOI: 10.1016/s0161-4754(99)70111-9

Source DB:  PubMed          Journal:  J Manipulative Physiol Ther        ISSN: 0161-4754            Impact factor:   1.437


  9 in total

1.  Characteristics of Paraspinal Muscle Spindle Response to Mechanically Assisted Spinal Manipulation: A Preliminary Report.

Authors:  William R Reed; Joel G Pickar; Randall S Sozio; Michael A K Liebschner; Joshua W Little; Maruti R Gudavalli
Journal:  J Manipulative Physiol Ther       Date:  2017-06-17       Impact factor: 1.437

2.  A review of the literature pertaining to the efficacy, safety, educational requirements, uses and usage of mechanical adjusting devices: Part 2 of 2.

Authors:  Shane H Taylor; Nicole D Arnold; Lesley Biggs; Christopher J Colloca; Dale R Mierau; Bruce P Symons; John J Triano
Journal:  J Can Chiropr Assoc       Date:  2004-06

3.  Effect of sampling rates on the quantification of forces, durations, and rates of loading of simulated side posture high-velocity, low-amplitude lumbar spine manipulation.

Authors:  Maruti Ram Gudavalli; James DeVocht; Ali Tayh; Ting Xia
Journal:  J Manipulative Physiol Ther       Date:  2013-06-18       Impact factor: 1.437

4.  An integrative treatment approach of a patient with cervical radiculitis: A case report.

Authors:  Leanne Apfelbeck
Journal:  J Chiropr Med       Date:  2005

5.  Increased multiaxial lumbar motion responses during multiple-impulse mechanical force manually assisted spinal manipulation.

Authors:  Tony S Keller; Christopher J Colloca; Robert J Moore; Robert Gunzburg; Deed E Harrison
Journal:  Chiropr Osteopat       Date:  2006-04-06

6.  The immediate effect of multiple mechanical impulses on electromyography and pressure pain threshold of lumbar latent trigger points: an experimental study.

Authors:  Bert Ameloot; Jeff Bagust
Journal:  Chiropr Man Therap       Date:  2016-07-04

7.  Glucose Metabolic Changes in the Brain and Muscles of Patients with Nonspecific Neck Pain Treated by Spinal Manipulation Therapy: A [18F]FDG PET Study.

Authors:  Akie Inami; Takeshi Ogura; Shoichi Watanuki; Md Mehedi Masud; Katsuhiko Shibuya; Masayasu Miyake; Rin Matsuda; Kotaro Hiraoka; Masatoshi Itoh; Arlan W Fuhr; Kazuhiko Yanai; Manabu Tashiro
Journal:  Evid Based Complement Alternat Med       Date:  2017-01-12       Impact factor: 2.629

8.  Collaborative Care for Older Adults with low back pain by family medicine physicians and doctors of chiropractic (COCOA): study protocol for a randomized controlled trial.

Authors:  Christine M Goertz; Stacie A Salsbury; Robert D Vining; Cynthia R Long; Andrew A Andresen; Mark E Jones; Kevin J Lyons; Maria A Hondras; Lisa Z Killinger; Fredric D Wolinsky; Robert B Wallace
Journal:  Trials       Date:  2013-01-16       Impact factor: 2.279

9.  Neural Response During a Mechanically Assisted Spinal Manipulation in an Animal Model: A Pilot Study.

Authors:  William R Reed; Michael A K Liebschner; Randall S Sozio; Joel G Pickar; Maruti R Gudavalli
Journal:  J Nov Physiother Phys Rehabil       Date:  2015-04-06
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.