Ct and advanced computer-aided design techniques offer the means for designing customised femoral stems. Our aim was to determine the Hounsfield (HU) value of the bone at the corticocancellous interface, as part of the criteria for the design algorithm. We obtained transverse CT images from eight human cadaver femora. The proximal femoral canal was rasped until contact with dense cortical bone was achieved. The femora were cut into several sections corresponding to the slice positions of the CT images. After obtaining a computerised image of the anatomical sections using a scanner, the inner cortical contour was outlined and transferred to the corresponding CT image. The pixels beneath this contour represent the CT density of the bone remaining after surgical rasping. Contours were generated automatically at nine HU levels from 300 to 1100 and the mean distance between the transferred contour and each of the HU-generated contours was computed. The contour generated along the 600-HU pixels was closest to the inner cortical contour of the rasped femur and therefore 600 HU seem to be the CT density of the corticocancellous interface in the proximal part of cadaver femora. Generally, femoral bone with a CT density beyond 600 HU is not removable by conventional reamers. Thus, we recommend the 600 HU threshold as one of several criteria for the design of custom femoral implants from CT data.

Entities:  

Mesh:

Year:  1999        PMID: 10068022

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  24 in total

1.  The Hounsfield value for cortical bone geometry in the proximal humerus--an in vitro study.

Authors:  Daren Lim Fat; Jim Kennedy; Rose Galvin; Fergal O'Brien; Frank Mc Grath; Hannan Mullett
Journal:  Skeletal Radiol       Date:  2011-09-20       Impact factor: 2.199

2.  Usefullness of three-dimensional templating software to quantify the contact state between implant and femur in total hip arthroplasty.

Authors:  Daisuke Inoue; Tamon Kabata; Toru Maeda; Yoshitomo Kajino; Kenji Fujita; Kazuhiro Hasegawa; Takashi Yamamoto; Tomoharu Takagi; Takaaki Ohmori; Hiroyuki Tsuchiya
Journal:  Eur J Orthop Surg Traumatol       Date:  2015-11-05

Review 3.  Quantitative computer-assisted osteodensitometry in total hip arthroplasty.

Authors:  R P Pitto; L A Mueller; K Reilly; R Schmidt; J Munro
Journal:  Int Orthop       Date:  2006-10-17       Impact factor: 3.075

4.  CORR Insights®: Sex-specific differences of the infraacetabular corridor: a biomorphometric CT-based analysis on a database of 523 pelves.

Authors:  Joseph D Lipman
Journal:  Clin Orthop Relat Res       Date:  2014-10-22       Impact factor: 4.176

5.  Quantitative and qualitative bone analysis in the maxillary lateral region.

Authors:  Carmen Elena Georgescu; Mugurel Constantin Rusu; Mihai Sandulescu; Aurelia Magdalena Enache; Andreea Cristiana Didilescu
Journal:  Surg Radiol Anat       Date:  2012-03-22       Impact factor: 1.246

6.  Version Correction via Eccentric Reaming Compromises Remaining Bone Quality in B2 Glenoids: A Computational Study.

Authors:  Xiang Chen; Akhil S Reddy; Andreas Kontaxis; Daniel S Choi; Timothy Wright; David M Dines; Russell F Warren; Julien Berhouet; Lawrence V Gulotta
Journal:  Clin Orthop Relat Res       Date:  2017-09-25       Impact factor: 4.176

7.  Comprehensive quality assurance phantom for the small animal radiation research platform (SARRP).

Authors:  M Jermoumi; H Korideck; M Bhagwat; P Zygmanski; G M Makrigiogos; R I Berbeco; R C Cormack; W Ngwa
Journal:  Phys Med       Date:  2015-05-08       Impact factor: 2.685

8.  Three-dimensional analyses of proximal humeral fractures using computed tomography with multiplanar reconstruction: early stability of fixation after osteosynthesis in relation to preoperative bone quality.

Authors:  Koki Ueda; Satoshi Ikemura; Akihisa Yamashita; Takashi Harada; Tetsuya Watanabe; Kenzo Shirasawa
Journal:  Eur J Orthop Surg Traumatol       Date:  2013-09-17

9.  Uncemented custom femoral components in hip arthroplasty. A prospective clinical study of 191 hips followed for at least 7 years.

Authors:  Pål Benum; Arild Aamodt
Journal:  Acta Orthop       Date:  2010-08       Impact factor: 3.717

10.  Off-axis loads cause failure of the distal radius at lower magnitudes than axial loads: a finite element analysis.

Authors:  Karen L Troy; Mark D Grabiner
Journal:  J Biomech       Date:  2007-03-26       Impact factor: 2.712

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.