Literature DB >> 10050949

The design of guide surfaces for fixed-bearing and mobile-bearing knee replacements.

P S Walker1, S Sathasivam.   

Abstract

In the natural knee, the femoral tibial contacts move posteriorly as the knee is flexed, guided primarily by the cruciate ligaments. This kinematic behaviour is important regarding muscle lever arms and in achieving a high flexion range. Most contemporary total knee designs use either posterior cruciate preservation or a cam system to produce posterior displacement with flexion, but there is no specific provision for anterior displacement. In this study, a method for the design of cams is described where the cams would guide the motion in both posterior and anterior directions, without requiring cruciate ligaments. The cams consist of a femoral Guide Surface interacting with a tibial Guide Surface while the main lateral and medial bearing surfaces carry the forces across the knee. It is shown that Guide Surfaces can be designed which provide the required motion, but with some laxity at different flexion ranges. It is then demonstrated that the Guide Surfaces can be applied to a range of possible knee designs including mobile-bearing types, rotating-platform types, and fixed-bearing types. The relative advantages of the different possibilities are discussed.

Mesh:

Year:  1999        PMID: 10050949     DOI: 10.1016/s0021-9290(98)00140-7

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  3 in total

1.  In vitro investigation of the influence of tibial slope on quadriceps extension force after total knee arthroplasty.

Authors:  Sven Ostermeier; Christof Hurschler; Henning Windhagen; Christina Stukenborg-Colsman
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2006-04-21       Impact factor: 4.342

2.  Dynamic in-vivo tibio-femoral and bearing motions in mobile bearing knee arthroplasty.

Authors:  S Fantozzi; A Leardini; S A Banks; M Marcacci; S Giannini; F Catani
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2003-07-09       Impact factor: 4.342

3.  Custom-designed orthopedic implants evaluated using finite element analysis of patient-specific computed tomography data: femoral-component case study.

Authors:  Ola L A Harrysson; Yasser A Hosni; Jamal F Nayfeh
Journal:  BMC Musculoskelet Disord       Date:  2007-09-13       Impact factor: 2.362