Effect of small amounts of 131I on 99mTc detection by the scintillation camera.

P T Makler, M J Eymontt, P Bloch, K Koch.   

Abstract

A patient who had recently received a therapeutic dose of Na131I showed no 99mTc activity on the scintillation camera when a pertechnetate brain scan was attempted. Subsequent experiments showed that, when even small amounts of 131I are present in the field of view of a scintillation camera equipped with a collimator designed for low-energy photons, counting rates from 99mTc are markedly affected.

Entities:  

Mesh:

Substances:

Year:  1976        PMID: 993844

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  1 in total

1.  Performance characteristics of a gamma camera over a wide range of energies.

Authors:  F R Hudson; J B Davis; A G Whittingham